Class II Cabinets

II 级生物安全柜

CLS · Biosafety Cabinets · 5 models

By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification)

Standards referenced

  • EN 12469 (biosafety cabinets)
  • NSF/ANSI 49 (biosafety cabinetry)
  • ISO 14644-1 (cleanroom classification)
  • EU GMP Annex 1 (sterile manufacturing)
  • GB 19489 (biosafety laboratory requirements)
  • IEC 61010-1

What are Class II Cabinets?

Class II biosafety cabinets use HEPA-filtered supply and exhaust air with a controlled inflow to protect the operator, the environment and the work. They are the standard containment device in microbiology, cell culture and sterile-handling laboratories.

5 models of Class II Cabinets

The 5 models below are current configurations; specifications are confirmed on quotation.

Class II Type A2 Biosafety Cabinet

II 级 A2 型生物安全柜

  • CLS-0001
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet type:
Class II Type A2
Inflow velocity:
0.38–0.51 m/s
Downflow velocity:
0.25–0.35 m/s
Recirculation:
≈ 70 %
HEPA grade:
H14 supply and exhaust
Width:
900–1800 mm

A Class II Type A2 biosafety cabinet recirculates about 70 % of HEPA-filtered air and exhausts 30 % to the room. It is the most common cabinet for cell culture and microbiology without volatile toxic chemicals.

Request a quote
View details

Class II Type B2 Biosafety Cabinet

II 级 B2 型生物安全柜

  • CLS-0002
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet type:
Class II Type B2
Inflow velocity:
0.38–0.51 m/s
Downflow velocity:
0.25–0.35 m/s
Recirculation:
0 %, total exhaust
Total exhaust:
700–1600 m3/h
HEPA grade:
H14 supply and exhaust

A Class II Type B2 biosafety cabinet exhausts 100 % of its air to the atmosphere through a dedicated duct, with no recirculation. It suits work with volatile toxic chemicals, radionuclides or hazardous vapours.

Request a quote
View details

Class II Type B1 Biosafety Cabinet

II 级 B1 型生物安全柜

  • CLS-0003
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet type:
Class II Type B1
Inflow velocity:
0.38–0.51 m/s
Downflow velocity:
0.25–0.35 m/s
Recirculation:
≈ 30 %
Exhaust fraction:
≈ 70 %
HEPA grade:
H14 supply and exhaust

A Class II Type B1 biosafety cabinet exhausts about 70 % of its air through a dedicated duct and recirculates the rest. It suits work needing both product protection and exhaust of small amounts of volatile chemicals.

Request a quote
View details

Class II Type A1 Biosafety Cabinet

II 级 A1 型生物安全柜

  • CLS-0004
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet type:
Class II Type A1
Inflow velocity:
0.38–0.51 m/s
Downflow velocity:
0.25–0.35 m/s
Recirculation:
≈ 70 %
Contaminated plenum:
Positive or atmospheric
HEPA grade:
H14 supply and exhaust

A Class II Type A1 biosafety cabinet recirculates about 70 % of its HEPA-filtered air and protects the work. It is limited to non-volatile, non-toxic work because its contaminated plenum is under positive pressure.

Request a quote
View details

Class II Type C1 Biosafety Cabinet

II 级 C1 型生物安全柜

  • CLS-0005
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet type:
Class II Type C1 (convertible)
Inflow velocity:
0.38–0.51 m/s
Downflow velocity:
0.25–0.35 m/s
Modes:
Recirculating and exhaust
Exhaust range:
30–100 % of airflow
HEPA grade:
H14 supply and exhaust

A Class II Type C1 biosafety cabinet is convertible, switching between recirculating and exhaust modes as the workload changes. It gives flexibility where both chemical and biological work occur in one cabinet.

Request a quote
View details

Compare Class II Cabinets models

ModelCabinet typeInflow velocityDownflow velocityRecirculationHEPA grade
Class II Type A2 Biosafety CabinetCLS-0001Class II Type A20.38–0.51 m/s0.25–0.35 m/s≈ 70 %H14 supply and exhaust
Class II Type B2 Biosafety CabinetCLS-0002Class II Type B20.38–0.51 m/s0.25–0.35 m/s0 %, total exhaustH14 supply and exhaust
Class II Type B1 Biosafety CabinetCLS-0003Class II Type B10.38–0.51 m/s0.25–0.35 m/s≈ 30 %H14 supply and exhaust
Class II Type A1 Biosafety CabinetCLS-0004Class II Type A10.38–0.51 m/s0.25–0.35 m/s≈ 70 %H14 supply and exhaust
Class II Type C1 Biosafety CabinetCLS-0005Class II Type C1 (convertible)0.38–0.51 m/s0.25–0.35 m/s—H14 supply and exhaust

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

How do Class II Cabinets work?

A Class II cabinet maintains three protective airflows at once. Room air is drawn through the front grille at an inflow velocity of about 0.38–0.51 m/s, forming an air curtain that keeps aerosols inside. A second stream of HEPA-filtered air flows downward over the work zone at about 0.25–0.35 m/s to protect the sample from contamination. Exhaust air passes a second HEPA filter before recirculation or discharge. The proportion recirculated defines the type: Type A2 returns about 70 % and exhausts 30 %, Type B1 exhausts about 70 %, and Type B2 exhausts 100 % without recirculation. A pressure or airflow sensor and alarm warn of loss of flow, and a negative-pressure plenum may surround the contaminated ductwork. Filter integrity and airflow are verified by aerosol testing at certification. The work surface is stainless steel, and a sash limits the operator's access to the work zone.

Which Class II Cabinets variants are available?

  • Class II Type A2 Biosafety Cabinet — A Class II Type A2 biosafety cabinet recirculates about 70 % of HEPA-filtered air and exhausts 30 % to the room.
  • Class II Type B2 Biosafety Cabinet — A Class II Type B2 biosafety cabinet exhausts 100 % of its air to the atmosphere through a dedicated duct, with no recirculation.
  • Class II Type B1 Biosafety Cabinet — A Class II Type B1 biosafety cabinet exhausts about 70 % of its air through a dedicated duct and recirculates the rest.
  • Class II Type A1 Biosafety Cabinet — A Class II Type A1 biosafety cabinet recirculates about 70 % of its HEPA-filtered air and protects the work.
  • Class II Type C1 Biosafety Cabinet — A Class II Type C1 biosafety cabinet is convertible, switching between recirculating and exhaust modes as the workload changes.

How do I choose the right variant?

Start from the biological risk assessment and the chemicals used. Type A2 suits most microbiology, cell culture and sterile work where no volatile toxic chemicals or radionuclides are involved. Choose Type B2 when volatile toxic chemicals, radionuclides or large amounts of vapour must be exhausted to atmosphere without recirculation. Type B1 is used when small amounts of volatile chemicals are handled and a dedicated exhaust exists. Type A1 is limited to non-toxic, non-volatile work because its plenum is under positive pressure. Type C1 is convertible and can switch between recirculating and exhaust modes as the workload changes. Confirm internal width and height for equipment, decide on recirculating or ducted exhaust, and verify certification to EN 12469 or NSF/ANSI 49. Plan annual recertification and filter replacement.

Where are Class II Cabinets used?

Class II Cabinets are specified in cell culture and aseptic technique, clinical microbiology and diagnostics, pharmaceutical sterility testing, virology and molecular biology and vaccine and biologics production.

Which standards and certifications apply?

Work with Class II Cabinets is commonly governed by EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification), EU GMP Annex 1 (sterile manufacturing), GB 19489 (biosafety laboratory requirements) and IEC 61010-1. Instrument certifications such as CE marking are confirmed per configuration and never assumed.

What are the key considerations before purchase?

Place the cabinet at least 30 cm from walls, doors and air diffusers, since cross-drafts above 0.15 m/s degrade the air curtain. Move arms slowly and avoid rapid movements that break the inflowing air. Do not place items over the front grille, because this blocks inflow and reduces protection, and keep large equipment towards the rear. Recertify at least annually and after filter change, fan repair or relocation, using aerosol leak, airflow and inflow velocity tests. For B2 and ducted types, confirm the exhaust fan is dedicated and interlocked, and verify make-up air. Clean surfaces with compatible disinfectants, use UV only as a supplement, and never work with toxic chemicals in an A2 cabinet without a vapour assessment.

Frequently asked questions about Class II Cabinets

It protects the operator, the environment and the work simultaneously, using inflow to contain aerosols, HEPA-filtered downflow to protect the sample, and HEPA-filtered exhaust. This three-way protection distinguishes it from fume hoods and clean benches.