Class II Type B1 Biosafety Cabinet

II 级 B1 型生物安全柜

  • SKU CLS-0003
  • Class II Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Cabinet typeClass II Type B1
Inflow velocity0.38–0.51 m/s
Downflow velocity0.25–0.35 m/s
Recirculation≈ 30 %
Exhaust fraction≈ 70 %
HEPA gradeH14 supply and exhaust
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification)

What is the Class II Type B1 Biosafety Cabinet?

A Class II Type B1 biosafety cabinet exhausts about 70 % of its air through a dedicated duct and recirculates the rest. It suits work needing both product protection and exhaust of small amounts of volatile chemicals.

What is the Class II Type B1 Biosafety Cabinet?

A Class II Type B1 biosafety cabinet exhausts about 70 % of its air through a dedicated duct and recirculates the rest. It suits work needing both product protection and exhaust of small amounts of volatile chemicals.

Overview

In a B1 cabinet, about 70 % of the air is exhausted and 30 % recirculated, the reverse of the A2. Air is arranged so that the most contaminated portion, taken from the work zone and lower rear plenum, goes directly to exhaust, while cleaner air from the upper portion is recirculated. This gives product protection with greater chemical removal than A2 but without the total external air demand of B2. Inflow is 0.38–0.51 m/s and downflow 0.25–0.35 m/s. A dedicated duct and exhaust fan with interlock are required, and volatile chemical use is limited to small quantities. Widths are commonly 900–1800 mm. Because some air is recirculated, B1 is not suitable for radionuclides or large volumes of volatile toxic chemicals, for which B2 is preferred.

How does it work?

A Class II cabinet maintains three protective airflows at once. Room air is drawn through the front grille at an inflow velocity of about 0.38–0.51 m/s, forming an air curtain that keeps aerosols inside. A second stream of HEPA-filtered air flows downward over the work zone at about 0.25–0.35 m/s to protect the sample from contamination. Exhaust air passes a second HEPA filter before recirculation or discharge. The proportion recirculated defines the type: Type A2 returns about 70 % and exhausts 30 %, Type B1 exhausts about 70 %, and Type B2 exhausts 100 % without recirculation. A pressure or airflow sensor and alarm warn of loss of flow, and a negative-pressure plenum may surround the contaminated ductwork. Filter integrity and airflow are verified by aerosol testing at certification. The work surface is stainless steel, and a sash limits the operator's access to the work zone.

What are the key features?

  • About 70 % exhaust and 30 % recirculation
  • Most contaminated air routed to exhaust
  • Dedicated duct and interlocked fan
  • Inflow 0.38–0.51 m/s, downflow 0.25–0.35 m/s
  • Product protection with chemical removal
  • Widths 900–1800 mm

What are the technical specifications?

A typical class ii type b1 biosafety cabinet configuration offers cabinet type of Class II Type B1, inflow velocity of 0.38–0.51 m/s, downflow velocity of 0.25–0.35 m/s, recirculation of ≈ 30 %, exhaust fraction of ≈ 70 % and HEPA grade of H14 supply and exhaust. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Class II Type B1 Biosafety Cabinet is used in cell culture with trace solvents, microbiology with small chemical volumes, clinical and reference diagnostics and aseptic handling in ducted rooms. Across the wider Class II Cabinets class, laboratories also deploy it for cell culture and aseptic technique, clinical microbiology and diagnostics and pharmaceutical sterility testing.

Which standards and certifications apply?

Relevant standards and methods include EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification), EU GMP Annex 1 (sterile manufacturing), GB 19489 (biosafety laboratory requirements) and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Place the cabinet at least 30 cm from walls, doors and air diffusers, since cross-drafts above 0.15 m/s degrade the air curtain. Move arms slowly and avoid rapid movements that break the inflowing air. Do not place items over the front grille, because this blocks inflow and reduces protection, and keep large equipment towards the rear. Recertify at least annually and after filter change, fan repair or relocation, using aerosol leak, airflow and inflow velocity tests. For B2 and ducted types, confirm the exhaust fan is dedicated and interlocked, and verify make-up air. Clean surfaces with compatible disinfectants, use UV only as a supplement, and never work with toxic chemicals in an A2 cabinet without a vapour assessment.

How do I select the right Class II Type B1 Biosafety Cabinet?

Choose Type B1 when product protection is needed and small volumes of volatile chemicals must be exhausted, but total external air supply is not available. Confirm the duct and interlock. RFQ: width, chemical type and volume, duct route, fan capacity and interlock.

How does it compare with other Class II Cabinets variants?

Start from the biological risk assessment and the chemicals used. Type A2 suits most microbiology, cell culture and sterile work where no volatile toxic chemicals or radionuclides are involved. Choose Type B2 when volatile toxic chemicals, radionuclides or large amounts of vapour must be exhausted to atmosphere without recirculation. Type B1 is used when small amounts of volatile chemicals are handled and a dedicated exhaust exists. Type A1 is limited to non-toxic, non-volatile work because its plenum is under positive pressure. Type C1 is convertible and can switch between recirculating and exhaust modes as the workload changes. Confirm internal width and height for equipment, decide on recirculating or ducted exhaust, and verify certification to EN 12469 or NSF/ANSI 49. Plan annual recertification and filter replacement.

Full specifications

Full specifications
Cabinet typeClass II Type B1
Inflow velocity0.38–0.51 m/s
Downflow velocity0.25–0.35 m/s
Recirculation≈ 30 %
Exhaust fraction≈ 70 %
HEPA gradeH14 supply and exhaust
Duct connectionDedicated, hard ducted
InterlockFan to cabinet, standard
Width900–1800 mm
Work surfaceStainless steel 304

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Cell culture with trace solvents
  • Microbiology with small chemical volumes
  • Clinical and reference diagnostics
  • Aseptic handling in ducted rooms

Key features

  • About 70 % exhaust and 30 % recirculation
  • Most contaminated air routed to exhaust
  • Dedicated duct and interlocked fan
  • Inflow 0.38–0.51 m/s, downflow 0.25–0.35 m/s
  • Product protection with chemical removal
  • Widths 900–1800 mm

Frequently asked questions about Class II Type B1 Biosafety Cabinet

B1 exhausts about 70 % and recirculates 30 %, A2 does the reverse, and B2 exhausts everything. B1 sits between them, giving more chemical removal than A2 without B2's full external air demand.

Request a quotation for the Class II Type B1 Biosafety Cabinet

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote