

Class II Type A2 Biosafety Cabinet
II 级 A2 型生物安全柜
- SKU CLS-0001
- Class II Cabinets
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Cabinet type | Class II Type A2 |
|---|---|
| Inflow velocity | 0.38–0.51 m/s |
| Downflow velocity | 0.25–0.35 m/s |
| Recirculation | ≈ 70 % |
| HEPA grade | H14 supply and exhaust |
| Width | 900–1800 mm |
By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification)
What is the Class II Type A2 Biosafety Cabinet?
A Class II Type A2 biosafety cabinet recirculates about 70 % of HEPA-filtered air and exhausts 30 % to the room. It is the most common cabinet for cell culture and microbiology without volatile toxic chemicals.
What is the Class II Type A2 Biosafety Cabinet?
A Class II Type A2 biosafety cabinet recirculates about 70 % of HEPA-filtered air and exhausts 30 % to the room. It is the most common cabinet for cell culture and microbiology without volatile toxic chemicals.
Overview
The A2 cabinet draws room air through the front grille at 0.38–0.51 m/s, passes part of the air through a HEPA supply filter for downflow over the work zone, and sends the rest plus the exhaust through another HEPA filter. About 30 % of the total air is exhausted to the room in the standard configuration, and 70 % recirculates. Both contaminated plenums are under negative pressure, which improves safety if a leak develops. Internal widths are commonly 900, 1200, 1500 and 1800 mm, with working height of 600–750 mm. Airflow alarms, a sash position sensor and a pressure gauge are standard. Because a small amount of air is exhausted to the room, the unit suits microbial and cell work but not volatile toxic chemicals or radionuclides, and can be connected to a canopy duct to remove the exhaust.
How does it work?
A Class II cabinet maintains three protective airflows at once. Room air is drawn through the front grille at an inflow velocity of about 0.38–0.51 m/s, forming an air curtain that keeps aerosols inside. A second stream of HEPA-filtered air flows downward over the work zone at about 0.25–0.35 m/s to protect the sample from contamination. Exhaust air passes a second HEPA filter before recirculation or discharge. The proportion recirculated defines the type: Type A2 returns about 70 % and exhausts 30 %, Type B1 exhausts about 70 %, and Type B2 exhausts 100 % without recirculation. A pressure or airflow sensor and alarm warn of loss of flow, and a negative-pressure plenum may surround the contaminated ductwork. Filter integrity and airflow are verified by aerosol testing at certification. The work surface is stainless steel, and a sash limits the operator's access to the work zone.
What are the key features?
- 70 % recirculation and 30 % exhaust
- Inflow 0.38–0.51 m/s and downflow 0.25–0.35 m/s
- Both contaminated plenums at negative pressure
- Widths 900–1800 mm, height 600–750 mm
- Airflow alarm and sash position sensor
- Optional canopy connection for exhaust
What are the technical specifications?
A typical class ii type a2 biosafety cabinet configuration offers cabinet type of Class II Type A2, inflow velocity of 0.38–0.51 m/s, downflow velocity of 0.25–0.35 m/s, recirculation of ≈ 70 %, HEPA grade of H14 supply and exhaust and width of 900–1800 mm. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Class II Type A2 Biosafety Cabinet is used in cell culture, clinical microbiology, sterility testing and molecular biology. Across the wider Class II Cabinets class, laboratories also deploy it for cell culture and aseptic technique, clinical microbiology and diagnostics and pharmaceutical sterility testing.
Which standards and certifications apply?
Relevant standards and methods include EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification), EU GMP Annex 1 (sterile manufacturing), GB 19489 (biosafety laboratory requirements) and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the cabinet at least 30 cm from walls, doors and air diffusers, since cross-drafts above 0.15 m/s degrade the air curtain. Move arms slowly and avoid rapid movements that break the inflowing air. Do not place items over the front grille, because this blocks inflow and reduces protection, and keep large equipment towards the rear. Recertify at least annually and after filter change, fan repair or relocation, using aerosol leak, airflow and inflow velocity tests. For B2 and ducted types, confirm the exhaust fan is dedicated and interlocked, and verify make-up air. Clean surfaces with compatible disinfectants, use UV only as a supplement, and never work with toxic chemicals in an A2 cabinet without a vapour assessment.
How do I select the right Class II Type A2 Biosafety Cabinet?
Choose Type A2 for most cell culture, microbiology and aseptic work without volatile toxic chemicals. Confirm width and work height for equipment, and decide whether a canopy duct is needed. RFQ: width, exhaust option, alarm package and certification schedule.
How does it compare with other Class II Cabinets variants?
Start from the biological risk assessment and the chemicals used. Type A2 suits most microbiology, cell culture and sterile work where no volatile toxic chemicals or radionuclides are involved. Choose Type B2 when volatile toxic chemicals, radionuclides or large amounts of vapour must be exhausted to atmosphere without recirculation. Type B1 is used when small amounts of volatile chemicals are handled and a dedicated exhaust exists. Type A1 is limited to non-toxic, non-volatile work because its plenum is under positive pressure. Type C1 is convertible and can switch between recirculating and exhaust modes as the workload changes. Confirm internal width and height for equipment, decide on recirculating or ducted exhaust, and verify certification to EN 12469 or NSF/ANSI 49. Plan annual recertification and filter replacement.
Full specifications
| Cabinet type | Class II Type A2 |
|---|---|
| Inflow velocity | 0.38–0.51 m/s |
| Downflow velocity | 0.25–0.35 m/s |
| Recirculation | ≈ 70 % |
| HEPA grade | H14 supply and exhaust |
| Width | 900–1800 mm |
| Work height | 600–750 mm |
| Work surface | Stainless steel 304 |
| Exhaust | 30 % to room or canopy duct |
| Noise level | ≤ 62 dB(A) |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Cell culture
- Clinical microbiology
- Sterility testing
- Molecular biology
Key features
- 70 % recirculation and 30 % exhaust
- Inflow 0.38–0.51 m/s and downflow 0.25–0.35 m/s
- Both contaminated plenums at negative pressure
- Widths 900–1800 mm, height 600–750 mm
- Airflow alarm and sash position sensor
- Optional canopy connection for exhaust
Frequently asked questions about Class II Type A2 Biosafety Cabinet
Negative pressure around the contaminated ducts means that if a seal or joint leaks, air flows inward rather than releasing aerosols into the room. This makes the A2 safer than older positive-pressure designs.
Yes. A canopy or thimble duct can capture the 30 % exhaust and carry it outdoors, which is useful for odours. It must not affect cabinet balance and should be verified at certification.
Small benchtop items such as pipettes, racks and a small centrifuge that fit under the sash. Large or tall equipment blocks airflow, so measure it and keep the rear and front grilles clear.
It protects the operator, the environment and the work simultaneously, using inflow to contain aerosols, HEPA-filtered downflow to protect the sample, and HEPA-filtered exhaust. This three-way protection distinguishes it from fume hoods and clean benches.
Type A2 is usual for cell culture without volatile toxic chemicals, because it protects the product and recirculates most air. Choose Type B2 if volatile toxic chemicals or radionuclides are used.
Not for volatile toxic or flammable chemicals, because recirculated vapour passes the HEPA filter and returns to the room. Use a Type B2 cabinet with 100 % exhaust for such work, after a risk assessment.
Related models






