What are Class I Cabinets?

Class I biosafety cabinets draw room air across the work surface and HEPA filter it before exhaust, protecting the operator and environment but not the sample. Research, veterinary and clinical laboratories use them for low-risk work and for housing equipment.

5 models of Class I Cabinets

The 5 models below are current configurations; specifications are confirmed on quotation.

Benchtop Class I Biosafety Cabinet

台式 I 级生物安全柜

  • CIC-0001
  • Class I Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet class:
Class I
Inflow velocity:
0.7–1.0 m/s
HEPA grade:
H14, ≥ 99.97 % at 0.3 µm
Width:
900–1500 mm
Internal height:
600–750 mm
Work surface:
Stainless steel 304

A benchtop Class I biosafety cabinet sits on a laboratory bench and provides single-pass HEPA exhaust for low-risk work. It protects the operator and environment from aerosols where sample protection is not required.

Request a quote
View details

Floor-Standing Class I Biosafety Cabinet

落地式 I 级生物安全柜

  • CIC-0002
  • Class I Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet class:
Class I
Inflow velocity:
0.7–1.0 m/s
Width:
1500–2400 mm
Internal height:
900–1200 mm
Working height:
850–950 mm
HEPA grade:
H14, ≥ 99.97 % at 0.3 µm

A floor-standing Class I biosafety cabinet provides a large work area at standing height for bulky equipment or multiple operators. It is used where benchtop units cannot accommodate the process or equipment size.

Request a quote
View details

Ducted Class I Biosafety Cabinet

管道式 I 级生物安全柜

  • CIC-0003
  • Class I Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet class:
Class I
Inflow velocity:
0.7–1.0 m/s, monitored
HEPA grade:
H14, ≥ 99.97 % at 0.3 µm
Duct connection:
Thimble or direct collar
Duct diameter:
200–300 mm
Airflow:
300–800 m3/h

A ducted Class I biosafety cabinet exhausts all air to the atmosphere after HEPA filtration. It is chosen when aerosols, odours or chemicals must leave the room completely rather than recirculate into it.

Request a quote
View details

Class I Animal Transfer Cabinet

I 级动物转移柜

  • CIC-0004
  • Class I Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet class:
Class I
Inflow velocity:
0.7–1.0 m/s
Width:
1200–1800 mm
Work surface:
Stainless steel 304
HEPA grade:
H14, ≥ 99.97 % at 0.3 µm
Exhaust:
Ducted to atmosphere

A Class I animal transfer cabinet provides a HEPA-exhausted work zone for cage changing and animal handling. It protects operators and the room from allergens and aerosols during routine animal facility work.

Request a quote
View details

Mobile Class I Biosafety Cabinet

移动式 I 级生物安全柜

  • CIC-0005
  • Class I Cabinets
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Cabinet class:
Class I
Inflow velocity:
0.7–1.0 m/s
Exhaust:
Recirculating, optional duct
HEPA grade:
H14, ≥ 99.97 % at 0.3 µm
Width:
900–1200 mm
Mobility:
Lockable castors, 4 positions

A mobile Class I biosafety cabinet is mounted on a wheeled base for use at different locations or in shared facilities. It gives single-pass HEPA protection where a fixed ducted installation is not practical.

Request a quote
View details

Compare Class I Cabinets models

ModelCabinet classInflow velocityHEPA gradeWidthInternal height
Benchtop Class I Biosafety CabinetCIC-0001Class I0.7–1.0 m/sH14, ≥ 99.97 % at 0.3 µm900–1500 mm600–750 mm
Floor-Standing Class I Biosafety CabinetCIC-0002Class I0.7–1.0 m/sH14, ≥ 99.97 % at 0.3 µm1500–2400 mm900–1200 mm
Ducted Class I Biosafety CabinetCIC-0003Class I0.7–1.0 m/s, monitoredH14, ≥ 99.97 % at 0.3 µm900–1800 mm—
Class I Animal Transfer CabinetCIC-0004Class I0.7–1.0 m/sH14, ≥ 99.97 % at 0.3 µm1200–1800 mm—
Mobile Class I Biosafety CabinetCIC-0005Class I0.7–1.0 m/sH14, ≥ 99.97 % at 0.3 µm900–1200 mm—

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

How do Class I Cabinets work?

A Class I cabinet works like a ventilated enclosure with a single HEPA-filtered exhaust path. Room air is pulled through the front opening at an inflow velocity of about 0.7–1.0 m/s, sweeps across the work surface, and carries aerosols and particles away from the operator. The air then passes a HEPA filter rated at least 99.97 % efficient at 0.3 µm before being discharged, either back into the room through a recirculating arrangement or, more commonly, to the atmosphere through a duct. Because there is no filtered supply air over the work zone, the cabinet does not protect the sample from contamination. Cabinets have a stainless steel work surface, a glass or polycarbonate sash, interior lighting and usually a UV lamp for surface decontamination between uses. Airflow alarms and a front grille complete the design; performance is confirmed by filter leak and airflow testing.

Which Class I Cabinets variants are available?

  • Benchtop Class I Biosafety Cabinet — A benchtop Class I biosafety cabinet sits on a laboratory bench and provides single-pass HEPA exhaust for low-risk work.
  • Floor-Standing Class I Biosafety Cabinet — A floor-standing Class I biosafety cabinet provides a large work area at standing height for bulky equipment or multiple operators.
  • Ducted Class I Biosafety Cabinet — A ducted Class I biosafety cabinet exhausts all air to the atmosphere after HEPA filtration.
  • Class I Animal Transfer Cabinet — A Class I animal transfer cabinet provides a HEPA-exhausted work zone for cage changing and animal handling.
  • Mobile Class I Biosafety Cabinet — A mobile Class I biosafety cabinet is mounted on a wheeled base for use at different locations or in shared facilities.

How do I choose the right variant?

Choose a Class I cabinet when operator and environmental protection are needed but the sample does not require protection, for example for low-risk microbiology, homogenising, or housing a centrifuge or shaking device. Confirm the inflow velocity and HEPA exhaust arrangement, and decide whether exhaust goes to the atmosphere by duct or is filtered and recirculated, since recirculation is not acceptable for many agents. Match the internal height to the equipment housed and the external footprint to the bench or floor space. Check whether the cabinet must be connected to an exhaust system and whether a thimble or direct connection is used. Confirm certification to EN 12469 or NSF/ANSI 49 where applicable, and plan annual recertification. Class I cabinets are not suitable when sterile product protection is required.

Where are Class I Cabinets used?

Class I Cabinets are specified in low-risk microbiology, centrifuge and equipment enclosures, veterinary and agricultural work, homogenising and sample disruption and teaching laboratories.

Which standards and certifications apply?

Work with Class I Cabinets is commonly governed by EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), ISO 14644-1 (cleanroom classification), GB 19489 (biosafety laboratory requirements) and IEC 61010-1 (safety requirements for electrical equipment). Instrument certifications such as CE marking are confirmed per configuration and never assumed.

What are the key considerations before purchase?

Install the cabinet away from doors, vents and high-traffic areas, with clearance for service and filter replacement. If ducted, verify the exhaust fan and duct are dedicated and that airflow cannot fall below the alarm threshold; an undersized or blocked duct can reverse the internal flow. Confirm filter integrity and inflow velocity at installation and at least annually, and record results. Reduce clutter inside the cabinet, since bulky items disturb airflow and reduce protection. Clean the work surface before and after use, use UV lamps only as a supplement to chemical disinfection, and never rely on UV for penetrating surfaces. Confirm that the exhaust discharge point meets local requirements.

Frequently asked questions about Class I Cabinets

No. It protects the operator and environment only, because unfiltered room air flows over the work surface. Use a Class II cabinet when the sample must also be protected from contamination.

Need help specifying Class I Cabinets?