What are TMA?

A thermomechanical analyzer measures the change in a sample's dimensions under a controlled load as temperature is programmed. It determines thermal expansion, shrinkage, softening, sintering and glass transition temperatures, and is used for metals, ceramics, glasses, polymers, composites and electronic materials.

5 models of TMA

The 5 models below are current configurations; specifications are confirmed on quotation.

Benchtop Thermomechanical Analyzer

台式热机械分析仪

  • TMA-0001
  • TMA
  • CE (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Temperature range:
−150–1000 °C (cooling dependent)
Displacement resolution:
1 nm
Displacement range:
±5 mm
Load range:
0.001–5 N programmable
Heating rate:
0.1–50 °C/min
Temperature accuracy:
±1 °C

A benchtop thermomechanical analyzer uses a vertical push-rod and LVDT displacement sensor with interchangeable probes, covering expansion, compression, penetration, tension and bending. It is the general-purpose instrument for coefficients of thermal expansion and softening temperatures.

Request a quote
View details

Horizontal Push-Rod Dilatometer

水平推杆式膨胀仪

  • TMA-0002
  • TMA
  • CE (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Temperature range:
Ambient–1600 °C
Displacement resolution:
10–100 nm
Contact force:
< 1 N, adjustable
Specimen length:
25–50 mm
Heating rate:
1–10 °C/min
Temperature accuracy:
±1 °C

A horizontal push-rod dilatometer places the specimen on a flat bed with the probe moving horizontally, which eliminates errors from gravity and probe weight. It measures precise thermal expansion of ceramics, glasses, metals and refractories over long specimens.

Request a quote
View details

High-Temperature Thermomechanical Analyzer

高温热机械分析仪

  • TMA-0003
  • TMA
  • CE (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Temperature range:
Ambient–1700 °C
Displacement resolution:
10–50 nm
Heating rate:
1–20 °C/min
Contact load:
0.01–2 N, or defined stress for refractories
Drift at temperature:
< 10 nm/h
Probe materials:
Alumina, graphite, sapphire

A high-temperature thermomechanical analyzer reaches 1500 to 1700 degrees Celsius with an alumina or graphite furnace, measuring expansion, sintering shrinkage and softening of ceramics, refractories, glasses and metal oxides used in demanding processes.

Request a quote
View details

Low-Temperature Thermomechanical Analyzer

低温热机械分析仪

  • TMA-0004
  • TMA
  • CE (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Temperature range:
−150–400 °C
Cooling method:
Liquid nitrogen or mechanical cooler
Heating and cooling rate:
0.5–20 °C/min
Displacement resolution:
5–20 nm
Temperature accuracy:
±1 °C
Load range:
0.001–2 N

A low-temperature thermomechanical analyzer cools with liquid nitrogen or a mechanical cooler to minus 150 degrees Celsius or below, measuring contraction, glass transitions and low-temperature dimensional stability of polymers, elastomers, adhesives and cryogenic materials.

Request a quote
View details

Automated Thermomechanical Analyzer

自动进样热机械分析仪

  • TMA-0005
  • TMA
  • CE (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Autosampler capacity:
20–40 specimens
Temperature range:
Ambient–1000 °C
Displacement resolution:
5–20 nm
Load range:
0.001–5 N
Heating rate:
0.1–50 °C/min
Measurement modes:
Expansion, penetration, compression, bending

An automated thermomechanical analyzer loads specimens with a robot, changes probes when required and runs long sequences unattended. It is used where expansion or softening tests are routine and operator time must be spent elsewhere.

Request a quote
View details

Compare TMA models

ModelTemperature rangeDisplacement resolutionLoad rangeHeating rateTemperature accuracy
Benchtop Thermomechanical AnalyzerTMA-0001−150–1000 °C (cooling dependent)1 nm0.001–5 N programmable0.1–50 °C/min±1 °C
Horizontal Push-Rod DilatometerTMA-0002Ambient–1600 °C10–100 nm—1–10 °C/min±1 °C
High-Temperature Thermomechanical AnalyzerTMA-0003Ambient–1700 °C10–50 nm—1–20 °C/min—
Low-Temperature Thermomechanical AnalyzerTMA-0004−150–400 °C5–20 nm0.001–2 N—±1 °C
Automated Thermomechanical AnalyzerTMA-0005Ambient–1000 °C5–20 nm0.001–5 N0.1–50 °C/min—

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

How do TMA work?

A probe rests on the specimen and is connected to a linear variable differential transformer, which converts displacement into an electrical signal with resolution near one nanometre. A furnace programs the temperature while a defined static or oscillating load holds the probe in contact. Depending on the probe and holder fitted, the instrument measures expansion, compression, penetration, tension or three-point bending, and the same displacement channel serves all modes. Coefficient of thermal expansion is calculated from the slope of the expansion curve over a chosen temperature interval, and softening or glass transition temperatures are read from the onset of probe penetration. Calibration uses reference materials of known expansion such as fused silica or aluminium, plus melting point standards for the temperature channel. Probe material, load and contact geometry must be recorded because they influence the measured displacement, and sample preparation must give parallel, flat faces.

Which TMA variants are available?

  • Benchtop Thermomechanical Analyzer — A benchtop thermomechanical analyzer uses a vertical push-rod and LVDT displacement sensor with interchangeable probes, covering expansion, compression, penetration, tension and bending.
  • Horizontal Push-Rod Dilatometer — A horizontal push-rod dilatometer places the specimen on a flat bed with the probe moving horizontally, which eliminates errors from gravity and probe weight.
  • High-Temperature Thermomechanical Analyzer — A high-temperature thermomechanical analyzer reaches 1500 to 1700 degrees Celsius with an alumina or graphite furnace, measuring expansion, sintering shrinkage and softening of ceramics, refractories, glasses and metal oxides used in demanding processes.
  • Low-Temperature Thermomechanical Analyzer — A low-temperature thermomechanical analyzer cools with liquid nitrogen or a mechanical cooler to minus 150 degrees Celsius or below, measuring contraction, glass transitions and low-temperature dimensional stability of polymers, elastomers, adhesives and cryogenic materials.
  • Automated Thermomechanical Analyzer — An automated thermomechanical analyzer loads specimens with a robot, changes probes when required and runs long sequences unattended.

How do I choose the right variant?

Decide the measurement mode first, because it dictates the probe and holder: expansion requires a flat-ended probe and often a horizontal or vertical dilatometer configuration, penetration and compression suit polymers and soft materials, tension suits films and fibres, and three-point bending suits rigid bars. Then set the temperature range, since it determines whether the furnace uses quartz, alumina or graphite and whether cooling uses liquid nitrogen or mechanical refrigeration. Displacement resolution matters most for thin films and low-expansion ceramics, where nanometre sensitivity is needed. Load range and control quality decide whether soft, low-modulus samples can be measured without deformation caused by the probe itself. Check the atmosphere and vacuum options, calibration standards supplied, sample preparation tooling and whether the instrument can be upgraded with a humidity generator or an automatic sample changer.

Where are TMA used?

TMA are specified in coefficient of thermal expansion of metals and ceramics, polymer softening and glass transition, electronic substrate and packaging materials, glass and glaze thermal expansion matching and composite and fibre dimensional stability.

Which standards and certifications apply?

Work with TMA is commonly governed by ASTM E831, ASTM E1363, ASTM D696, ISO 11359-1, ISO 11359-2 and IEC 61010-1. Instrument certifications such as CE marking are confirmed per configuration and never assumed.

What are the key considerations before purchase?

Prepare specimens with parallel, flat and smooth faces, and record the original length precisely, because expansion results are normalized to it. Keep the load low enough to avoid deforming the sample, and allow the probe to settle at the starting temperature before beginning the ramp. Verify the expansion channel with a certified reference such as fused silica or aluminium and check the temperature channel with melting point standards. Beware of systematic artefacts: probe and holder expansion must be subtracted, and soft samples can show apparent shrinkage from creep under load. Keep the furnace and probe clean, replace worn probes, and use the correct purge gas to avoid oxidation or reaction with the specimen. Record mode, load, probe material, gas and calibration data with every curve.

Frequently asked questions about TMA

It measures dimensional change of a specimen under a controlled load as temperature changes. From that curve it determines thermal expansion coefficient, shrinkage on curing or sintering, softening and penetration onset, glass transition by expansion, and dimensional stability of films and composites.