Online Process Turbidity Analyzer
在线过程浊度分析仪
- SKU TUM-0004
- Turbidity Meters
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Measuring range | 0–100 NTU or 0–4000 NTU |
|---|---|
| Resolution | 0.001 NTU (low range) |
| Accuracy | ±2% of reading or 0.02 NTU |
| Response time | ≤60 s (T90) |
| Sample flow rate | 0.25–1 L/min |
| Sample pressure | 0.1–0.6 MPa |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 7027-1, USEPA Method 180.1, GB/T 13200
What is the Online Process Turbidity Analyzer?
An online process turbidity analyzer continuously measures turbidity of a flowing sample in filtration, water treatment or production lines. It uses a flow cell with bubble trap, automatic cleaning and 4–20 mA output to give real-time readings without manual sampling.
What is the Online Process Turbidity Analyzer?
An online process turbidity analyzer continuously measures turbidity of a flowing sample in filtration, water treatment or production lines. It uses a flow cell with bubble trap, automatic cleaning and 4–20 mA output to give real-time readings without manual sampling.
Overview
The online turbidity analyzer is installed on a sample line, drawing a continuous flow of 0.25–1 L/min through a flow cell with an integral bubble trap that removes air before measurement. A 90° nephelometric sensor with infrared or white light source reports 0–100 NTU or 0–4000 NTU, depending on the model, with a response time of under 60 seconds. Automatic wiper or ultrasonic cleaning keeps optical windows free from fouling, while programmable alarm relays, 4–20 mA outputs and Modbus or Profibus communication integrate with SCADA systems. Calibration is verified with a built-in solid secondary standard and primary calibration by formazin. Typical uses include filter effluent monitoring at 0.01–1 NTU, settled water control and production streams in beverage and chemical processing. Regular cleaning and lamp replacement are needed.
How does it work?
A turbidity meter directs a beam of light through a sample and measures the portion scattered by suspended particles. In nephelometry the detector sits at 90° to the incident beam, and the signal is proportional to particle concentration within the linear range. Tungsten lamps at 400–600 nm are described in USEPA Method 180.1 and reported in NTU, while infrared LEDs at 860 nm are specified in ISO 7027 and reported in FNU. Ratio designs combine the 90° signal with transmitted and forward-scatter detectors to compensate for color and lamp drift. Calibration uses formazin or stabilized polymer standards such as 0.1, 20, 100 and 800 NTU. Some beverage instruments add a 25° detector to measure beer haze in EBC units, because larger particles scatter predominantly forward.
What are the key features?
- Flow cell with integral bubble trap reduces air-induced false readings.
- Automatic wiper or ultrasonic cleaning limits optical window fouling.
- Programmable alarm relays and 4–20 mA output support process control.
- Modbus RTU or Profibus DP communication connects to SCADA systems.
- Built-in solid secondary standard allows quick verification without reagents.
- Wall or panel mounting with sample pressure regulation.
What are the technical specifications?
A typical online process turbidity analyzer configuration offers measuring range of 0–100 NTU or 0–4000 NTU, resolution of 0.001 NTU (low range), accuracy of ±2% of reading or 0.02 NTU, response time of ≤60 s (T90), sample flow rate of 0.25–1 L/min and sample pressure of 0.1–0.6 MPa. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Online Process Turbidity Analyzer is used in filter effluent monitoring, drinking water treatment plants, beverage production lines, chemical process streams and cooling water and reuse systems. Across the wider Turbidity Meters class, laboratories also deploy it for filtered beverage and beer haze testing, pharmaceutical water clarity control and drinking water and wastewater monitoring.
Which standards and certifications apply?
Relevant standards and methods include ISO 7027-1, USEPA Method 180.1, GB/T 13200, Standard Methods 2130 B, ASBC Beer-9 and EBC Method 9.29. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Use clean, scratch-free and indexed sample vials, and wipe them with lint-free cloth and a thin film of silicone oil to hide minor imperfections. Degas samples and avoid bubbles, which produce false high readings. Calibrate with fresh formazin or certified stabilized standards, following the manufacturer's sequence, and verify with a secondary gel standard daily. Keep the optical chamber free of condensation by allowing cold samples to reach room temperature. Replace tungsten lamps according to hours of operation. Document calibration and verification records for GLP or ISO 17025 laboratories, and dispose of formazin according to local regulations because it contains hydrazine sulfate.
How do I select the right Online Process Turbidity Analyzer?
Select the online analyzer where continuous readings and alarms are needed, rather than periodic grab samples. It requires sample conditioning and more maintenance than a benchtop meter. In an RFQ, specify range, response time, sample flow and pressure, communication protocol, cleaning mode and the required ingress protection.
How does it compare with other Turbidity Meters variants?
Choose a turbidity meter by measuring range, regulatory method, sample type and portability. Low-level clarity work below 1 NTU, such as filtered pharmaceutical water, requires a ratio benchtop instrument with stray light below 0.01 NTU and a resolution of 0.001 NTU. Wide-range industrial samples up to 4000 NTU suit a robust ratio design. Choose a white-light instrument for USEPA reporting and an infrared instrument for ISO 7027 compliance and colored samples. Beverage laboratories need EBC or ASBC units and a 25° channel. Field work favors portable, battery-powered models with IP67 housing. Online measurement needs a bubble trap and automatic cleaning. Specify vial or flow-cell format, calibration standards supplied and data output requirements.
Full specifications
| Measuring range | 0–100 NTU or 0–4000 NTU |
|---|---|
| Resolution | 0.001 NTU (low range) |
| Accuracy | ±2% of reading or 0.02 NTU |
| Response time | ≤60 s (T90) |
| Sample flow rate | 0.25–1 L/min |
| Sample pressure | 0.1–0.6 MPa |
| Cleaning | Automatic wiper or ultrasonic |
| Outputs | 4–20 mA, 2 relays, Modbus RTU |
| Ingress protection | IP65 (controller) |
| Power supply | 100–240 V AC or 24 V DC |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Filter effluent monitoring
- Drinking water treatment plants
- Beverage production lines
- Chemical process streams
- Cooling water and reuse systems
Key features
- Flow cell with integral bubble trap reduces air-induced false readings.
- Automatic wiper or ultrasonic cleaning limits optical window fouling.
- Programmable alarm relays and 4–20 mA output support process control.
- Modbus RTU or Profibus DP communication connects to SCADA systems.
- Built-in solid secondary standard allows quick verification without reagents.
- Wall or panel mounting with sample pressure regulation.
Frequently asked questions about Online Process Turbidity Analyzer
Air bubbles scatter light like particles and cause spikes or false high readings. A bubble trap or debubbler in the flow cell removes entrained air before the sample reaches the optical chamber, which stabilizes readings especially in pressurized or degassing samples.
Cleaning intervals depend on the sample. Automatic wipers reduce manual work, but a full clean of the flow cell and optics is typically scheduled every 1–3 months. Heavily fouling streams may require weekly checks, and verification with the solid standard confirms performance.
Yes. Most models provide 4–20 mA analog output and relay contacts, with Modbus RTU or Profibus DP digital communication as an option. This allows trending, alarms and automatic control actions such as filter backwash initiation.
Both report scattered light at 90° calibrated against formazin. NTU is used with white-light instruments as in USEPA Method 180.1, while FNU is used with 860 nm infrared instruments under ISO 7027. Values are close for colorless samples but may differ for colored or particulate-rich samples.
Formazin is the primary reference, but many laboratories use certified stabilized polymer standards that are safer and stable for months. Use standards traceable to the manufacturer's certified values and a multi-point calibration covering your measurement range, including a low standard near 0.1 NTU.
Common causes are air bubbles, scratched or dirty vials, condensation on cold samples, settling particles and lamp aging. Degas the sample, clean and index the vial, allow temperature equilibration, mix gently before reading and check lamp hours if drift continues.