Coulometric Karl Fischer Titrator for Insulating Oil — product photo

Coulometric Karl Fischer Titrator for Insulating Oil

绝缘油库仑法卡尔费休水分测定仪

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Measuring range1 ppm–5 % water (10 µg–100 mg)
Resolution0.1 µg water
Accuracy± 3 µg (< 1000 µg), ± 0.3 % above
Repeatability (RSD)≤ 3 % at 10 ppm
Drift stability≤ 10 µg/min
Sample size1–10 g oil
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 60156, ASTM D1816, ASTM D877

What is the Coulometric Karl Fischer Titrator for Insulating Oil?

A coulometric Karl Fischer titrator measures water in insulating oil at ppm level by generating iodine electrochemically in a reagent cell. It is the referee method for moisture in transformer oil and is used with heated oven or direct injection.

What is the Coulometric Karl Fischer Titrator for Insulating Oil?

A coulometric Karl Fischer titrator measures water in insulating oil at ppm level by generating iodine electrochemically in a reagent cell. It is the referee method for moisture in transformer oil and is used with heated oven or direct injection.

Overview

In the titration cell, iodine is generated at the anode from iodide in the Karl Fischer reagent, and each mole of iodine reacts stoichiometrically with one mole of water in the presence of sulfur dioxide and a base. A double platinum indicator electrode detects excess iodine and stops generation; by Faraday's law, 10.71 coulombs correspond to 1 mg of water. Because insulating oil may be injected directly by weight, reagent exhaustion and poor solubility can occur, so instruments often offer oil-specific anolytes or an oven that heats the sample to 120 to 140 degrees Celsius and carries evaporated water to the cell by dry carrier gas. The method is standardized in IEC 60814 and ASTM D1533, with detection below 1 ppm.

How does it work?

Insulating oil must resist electrical stress, remove heat and protect paper insulation, so testing targets each of these functions. Dielectric breakdown voltage is measured by increasing AC voltage at a fixed rate between standard electrodes in a stirred test cell until a discharge occurs. Dissolved gas analysis extracts gases from oil by headspace, vacuum or stripping and separates hydrogen, methane, ethane, ethylene, acetylene, carbon monoxide and carbon dioxide by gas chromatography with thermal conductivity and flame ionization detectors. Karl Fischer coulometry titrates water electrochemically. Dissipation factor and resistivity are measured in a three-terminal cell heated to 90 degrees Celsius. Interfacial tension uses a platinum ring that is pulled through the oil-water interface, and the pulling force gives the tension in mN/m.

What are the key features?

  • Coulometric generation with double platinum indicator electrode
  • Oil-compatible anolyte and optional heated oven to 300 °C
  • Automatic drift compensation and end-point detection
  • Direct ppm or µg water result with weight entry
  • Memory for results and USB or RS-232 output

What are the technical specifications?

A typical coulometric karl fischer titrator for insulating oil configuration offers measuring range of 1 ppm–5 % water (10 µg–100 mg), resolution of 0.1 µg water, accuracy of ± 3 µg (< 1000 µg), ± 0.3 % above, repeatability (RSD) of ≤ 3 % at 10 ppm, drift stability of ≤ 10 µg/min and sample size of 1–10 g oil. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Coulometric Karl Fischer Titrator for Insulating Oil is used in transformer oil moisture measurement, cable and capacitor oil testing and lubricant and hydraulic oil water content. Across the wider Transformer Oil Testing class, laboratories also deploy it for transformer acceptance and commissioning tests, routine insulating oil monitoring for utilities and fault diagnosis by dissolved gas analysis.

Which standards and certifications apply?

Relevant standards and methods include IEC 60156, ASTM D1816, ASTM D877, IEC 60599, ASTM D3612, IEC 60814, IEC 60247, ASTM D971 and GB/T 507. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Sampling is as critical as analysis: use clean, dry glass syringes or sampling bottles, avoid air exposure for dissolved gas samples and record oil temperature. Test cells and electrodes must be cleaned and rinsed with the test oil, and gap distance verified with a gauge. High-voltage testers need grounded benches, interlocked lids and periodic verification of voltage and rate of rise. Gas chromatographs require carrier gas purity, column conditioning and calibration with certified gas mixtures. Karl Fischer reagents are replaced per usage. Follow facility safety rules, and dispose of oil and halogenated waste according to regulations.

How do I select the right Coulometric Karl Fischer Titrator for Insulating Oil?

Select the coulometric titrator for moisture measurement below roughly 1 percent, particularly for transformer oil where ppm accuracy is needed. For oils with high water content choose volumetric titration. In an RFQ, specify oven option, cell type, reagent grades and compliance with IEC 60814 or ASTM D1533.

How does it compare with other Transformer Oil Testing variants?

Choose equipment by the decisions it supports. A breakdown voltage tester is the minimum for acceptance and routine screening; select the test voltage class (80 kV for most service oil, 100 kV or higher for new oil and EHV equipment) and the electrode type that your standard requires, VDE or spherical for IEC 60156, or ASTM D1816 or D877 types. Dissolved gas analysis is the key diagnostic for fault detection and suits utilities with many assets. Coulometric Karl Fischer is needed because moisture strongly affects paper insulation life. Tan delta and resistivity testers help detect polar contaminants and aging, while IFT indicates sludge precursors. Also check sample volume, automation level, test-cell cleaning, data logging and safe high-voltage interlocks.

Full specifications

Full specifications
Measuring range1 ppm–5 % water (10 µg–100 mg)
Resolution0.1 µg water
Accuracy± 3 µg (< 1000 µg), ± 0.3 % above
Repeatability (RSD)≤ 3 % at 10 ppm
Drift stability≤ 10 µg/min
Sample size1–10 g oil
Oven temperature50–300 °C (optional)
Titration cellDiaphragm or diaphragm-free
Power supply220 V AC, 50 Hz

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Transformer oil moisture measurement
  • Cable and capacitor oil testing
  • Lubricant and hydraulic oil water content

Key features

  • Coulometric generation with double platinum indicator electrode
  • Oil-compatible anolyte and optional heated oven to 300 °C
  • Automatic drift compensation and end-point detection
  • Direct ppm or µg water result with weight entry
  • Memory for results and USB or RS-232 output

Frequently asked questions about Coulometric Karl Fischer Titrator for Insulating Oil

Karl Fischer titration reacts specifically with water, so it is not affected by other volatile compounds. Coulometric detection provides absolute measurement at ppm level, which is essential because even small moisture changes affect insulation performance.

Request a quotation for the Coulometric Karl Fischer Titrator for Insulating Oil