Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil
绝缘油介质损耗及体积电阻率测试仪
- SKU TOT-0004
- Transformer Oil Testing
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Tan delta range | 0.00001–100 % |
|---|---|
| Tan delta accuracy | ± (1 % reading + 0.00005) |
| Test voltage (AC) | 0–2000 V, 500 V/mm typical |
| Test voltage (DC) | 250 V/mm typical |
| Resistivity range | 1 × 10^8 to 1 × 10^15 Ω·m |
| Cell volume | ≈ 40–60 mL |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 60156, ASTM D1816, ASTM D877
What is the Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil?
A dielectric loss tangent and resistivity tester measures dissipation factor and DC resistivity of insulating oil in a heated three-terminal cell. It detects polar contaminants, aging by-products and moisture that reduce insulating quality before breakdown voltage changes, and is used to grade new and reclaimed oil.
What is the Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil?
A dielectric loss tangent and resistivity tester measures dissipation factor and DC resistivity of insulating oil in a heated three-terminal cell. It detects polar contaminants, aging by-products and moisture that reduce insulating quality before breakdown voltage changes, and is used to grade new and reclaimed oil.
Overview
The tester applies an AC test voltage, typically 500 V/mm at 50 Hz, to a three-terminal cell with a guard electrode that eliminates surface leakage. A bridge or digital phase-detection circuit measures capacitance and the phase angle between voltage and current to give the dissipation factor, tan delta. Relative permittivity is derived from capacitance relative to empty-cell capacitance. For volume resistivity, a DC voltage of 250 V/mm is applied and the current after one minute of electrification is measured. The cell is heated by a thermostat to 90 degrees Celsius with stability within 0.5 degrees, because both parameters vary strongly with temperature. Clean, dry cell assemblies are essential; thorough cleaning between samples keeps results repeatable.
How does it work?
Insulating oil must resist electrical stress, remove heat and protect paper insulation, so testing targets each of these functions. Dielectric breakdown voltage is measured by increasing AC voltage at a fixed rate between standard electrodes in a stirred test cell until a discharge occurs. Dissolved gas analysis extracts gases from oil by headspace, vacuum or stripping and separates hydrogen, methane, ethane, ethylene, acetylene, carbon monoxide and carbon dioxide by gas chromatography with thermal conductivity and flame ionization detectors. Karl Fischer coulometry titrates water electrochemically. Dissipation factor and resistivity are measured in a three-terminal cell heated to 90 degrees Celsius. Interfacial tension uses a platinum ring that is pulled through the oil-water interface, and the pulling force gives the tension in mN/m.
What are the key features?
- Three-terminal stainless steel cell with guard electrode
- Thermostatic heating to 90 °C with ± 0.5 °C stability
- Simultaneous tan delta, permittivity and resistivity results
- Automatic temperature correction and cell constant calibration
- Shielded measuring circuit for low-current accuracy
What are the technical specifications?
A typical dielectric loss tangent and volume resistivity tester for insulating oil configuration offers tan delta range of 0.00001–100 %, tan delta accuracy of ± (1 % reading + 0.00005), test voltage (AC) of 0–2000 V, 500 V/mm typical, test voltage (DC) of 250 V/mm typical, resistivity range of 1 × 10^8 to 1 × 10^15 Ω·m and cell volume of ≈ 40–60 mL. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil is used in insulating oil quality grading, oil reclamation efficiency checks and cable and capacitor oil testing. Across the wider Transformer Oil Testing class, laboratories also deploy it for transformer acceptance and commissioning tests, routine insulating oil monitoring for utilities and fault diagnosis by dissolved gas analysis.
Which standards and certifications apply?
Relevant standards and methods include IEC 60156, ASTM D1816, ASTM D877, IEC 60599, ASTM D3612, IEC 60814, IEC 60247, ASTM D971 and GB/T 507. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Sampling is as critical as analysis: use clean, dry glass syringes or sampling bottles, avoid air exposure for dissolved gas samples and record oil temperature. Test cells and electrodes must be cleaned and rinsed with the test oil, and gap distance verified with a gauge. High-voltage testers need grounded benches, interlocked lids and periodic verification of voltage and rate of rise. Gas chromatographs require carrier gas purity, column conditioning and calibration with certified gas mixtures. Karl Fischer reagents are replaced per usage. Follow facility safety rules, and dispose of oil and halogenated waste according to regulations.
How do I select the right Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil?
Select this tester when you need early detection of oil aging or polar contamination, or must grade new and reclaimed oil to IEC 60247 or ASTM D924 and D1169. Specify the cell constant, test frequency, temperature range, resistivity range and need for automatic cleaning accessories.
How does it compare with other Transformer Oil Testing variants?
Choose equipment by the decisions it supports. A breakdown voltage tester is the minimum for acceptance and routine screening; select the test voltage class (80 kV for most service oil, 100 kV or higher for new oil and EHV equipment) and the electrode type that your standard requires, VDE or spherical for IEC 60156, or ASTM D1816 or D877 types. Dissolved gas analysis is the key diagnostic for fault detection and suits utilities with many assets. Coulometric Karl Fischer is needed because moisture strongly affects paper insulation life. Tan delta and resistivity testers help detect polar contaminants and aging, while IFT indicates sludge precursors. Also check sample volume, automation level, test-cell cleaning, data logging and safe high-voltage interlocks.
Full specifications
| Tan delta range | 0.00001–100 % |
|---|---|
| Tan delta accuracy | ± (1 % reading + 0.00005) |
| Test voltage (AC) | 0–2000 V, 500 V/mm typical |
| Test voltage (DC) | 250 V/mm typical |
| Resistivity range | 1 × 10^8 to 1 × 10^15 Ω·m |
| Cell volume | ≈ 40–60 mL |
| Temperature range | Ambient to 100 °C (90 °C typical) |
| Frequency | 50 Hz (or 60 Hz) |
| Power supply | 220 V AC, 50 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Insulating oil quality grading
- Oil reclamation efficiency checks
- Cable and capacitor oil testing
Key features
- Three-terminal stainless steel cell with guard electrode
- Thermostatic heating to 90 °C with ± 0.5 °C stability
- Simultaneous tan delta, permittivity and resistivity results
- Automatic temperature correction and cell constant calibration
- Shielded measuring circuit for low-current accuracy
Frequently asked questions about Dielectric Loss Tangent and Volume Resistivity Tester for Insulating Oil
A high dissipation factor shows that the oil has significant polar contaminants, moisture, or aging by-products that cause dielectric losses. Even if breakdown voltage remains acceptable, rising tan delta indicates declining insulation quality and may prompt reclamation or replacement.
Dissipation factor and resistivity vary strongly with temperature, so standards fix 90 °C for service oil to give comparable results and greater sensitivity to contaminants. Some standards also specify 20 or 25 °C measurements for comparison.
Rinse with the next sample, then with solvent such as petroleum ether or heptane, and dry in a clean oven. For polluted oils use detergent washing followed by hot-water and solvent rinses, verifying cleanliness with a blank measurement.
Standard tests include dielectric breakdown voltage, water content, acidity, dissipation factor, resistivity, interfacial tension and dissolved gas analysis. Together they assess insulating strength, contamination, aging and incipient faults, with frequency depending on transformer voltage class and criticality.
Dissolved gas analysis extracts gases from transformer oil and measures hydrogen, hydrocarbons and carbon oxides by gas chromatography. Gas ratios and concentrations indicate fault types such as partial discharge, overheating or arcing, using interpretation methods in IEC 60599 or IEEE C57.104.
Water lowers dielectric strength and accelerates cellulose degradation, shortening transformer life. Because most water resides in the paper insulation, oil moisture is interpreted together with temperature, using equilibrium curves to estimate paper moisture content.