Cryostat with Vacuum Section Transfer System
带真空转移系统的恒冷箱切片机
- SKU CRY-0005
- Cryostats
- CE (to be confirmed)
- IEC 61010-1 compliance (to be confirmed)
- IEC 61326-1 EMC (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Chamber temperature range | –5 to –35 °C |
|---|---|
| Specimen temperature range | –10 to –40 °C |
| Section thickness range | 1–100 µm |
| Transfer nozzle | Adjustable vacuum |
| Slide holder | Indexed, multi-position |
| Blade type | Disposable low-profile or high-profile |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-101, IEC 61326-1
What is the Cryostat with Vacuum Section Transfer System?
A cryostat with a vacuum section transfer system lifts and deposits cut sections onto slides or grids without manual handling. It is chosen for large, fragile or serial sections where manual transfer causes loss or folds.
What is the Cryostat with Vacuum Section Transfer System?
A cryostat with a vacuum section transfer system lifts and deposits cut sections onto slides or grids without manual handling. It is chosen for large, fragile or serial sections where manual transfer causes loss or folds.
Overview
Transferring fragile frozen sections by brush or forceps risks folding, tearing and loss, especially for large whole-mount sections or long series. A vacuum section transfer system uses a controlled suction nozzle that picks up the section inside the cold chamber and places it onto a slide held at a defined position, with adjustable vacuum so thick or thin sections are handled without damage. Some systems index the slide holder so serial sections land in a fixed order, which speeds up research workflows. The chamber must still be cold and the blade sharp, and the operator controls section thickness as usual. These instruments are used for large tissue sections, whole-organ studies, and studies where section loss would compromise quantitative analysis.
How does it work?
A cryostat places a microtome inside an insulated chamber cooled by one or two refrigeration compressors. The chamber and often a separate specimen stage are brought to between –15 and –35 °C, freezing the tissue to a suitable hardness. The operator attaches the frozen block to a chuck with embedding medium and sections it with a disposable blade held at a fixed angle, producing sections typically 1 to 100 µm thick. Sections adhere to room-temperature slides and are fixed immediately in alcohol or acetone. Anti-roll plates and brush techniques guide ribbons free of curling. Because the whole cutting zone stays cold, tissue morphology is preserved without paraffin infiltration, and total turnaround from specimen to stained slide can be under 20 minutes.
What are the key features?
- Controlled suction nozzle for section pickup
- Adjustable vacuum for different section thicknesses
- Automatic deposition onto slides or grids
- Indexed slide holder for ordered serial sections
- Reduces folding and loss of fragile sections
- Operates within the cold chamber
What are the technical specifications?
A typical cryostat with vacuum section transfer system configuration offers chamber temperature range of –5 to –35 °C, specimen temperature range of –10 to –40 °C, section thickness range of 1–100 µm, transfer nozzle of Adjustable vacuum, slide holder of Indexed, multi-position and blade type of Disposable low-profile or high-profile. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Cryostat with Vacuum Section Transfer System is used in large whole-mount frozen sections, long serial section series, whole-organ and whole-body studies and quantitative research workflows. Across the wider Cryostats class, laboratories also deploy it for intraoperative frozen sections, immunofluorescence on fresh tissue and enzyme histochemistry.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-101, IEC 61326-1, ISO 15189 and ISO 13485. Certification candidates for this model are CE, IEC 61010-1 compliance and IEC 61326-1 EMC; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the cryostat in a room kept below 25 °C with good ventilation, because condensers reject significant heat. Allow several hours for initial cool-down, then verify chamber and specimen temperatures against a calibrated thermometer. Defrost and clean the chamber on schedule; residual cryo-embedding medium degrades section quality. Use anti-roll plates and change blades often. Monitor for frost build-up, compressor noise and door seal condition, and service annually. Maintain temperature logs, cleaning records and staff training, since these documents support accreditation of frozen section services.
How do I select the right Cryostat with Vacuum Section Transfer System?
Specify a vacuum transfer system when sections are too large or fragile for reliable manual pickup, or when serial sections must be deposited in order for downstream analysis. Confirm nozzle compatibility with your section sizes, slide format and section thickness range. For routine diagnostic work the system is usually unnecessary.
How does it compare with other Cryostats variants?
Choose a cryostat by the work it must support. Clinical laboratories doing intraoperative frozen sections need fast cool-down, a stable chamber near –20 °C, reliable blade cooling and often UV sterilisation for contamination control. Research and immunohistochemistry laboratories value accurate independent specimen cooling, wide temperature range to –35 °С and a motorised or ergonomic cutting system for repetitive work. Confirm chamber and specimen temperature ranges, cool-down time, defrost cycle, section thickness range, blade type and whether the cutting area can be cleaned with the chamber at working temperature. Also check section transfer options and the footprint of the lab bench.
Full specifications
| Chamber temperature range | –5 to –35 °C |
|---|---|
| Specimen temperature range | –10 to –40 °C |
| Section thickness range | 1–100 µm |
| Transfer nozzle | Adjustable vacuum |
| Slide holder | Indexed, multi-position |
| Blade type | Disposable low-profile or high-profile |
| Cool-down time | about 2 h |
| Defrost | Programmable cycle |
| Power supply | 220 V, 50 Hz, about 1.3 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Large whole-mount frozen sections
- Long serial section series
- Whole-organ and whole-body studies
- Quantitative research workflows
Key features
- Controlled suction nozzle for section pickup
- Adjustable vacuum for different section thicknesses
- Automatic deposition onto slides or grids
- Indexed slide holder for ordered serial sections
- Reduces folding and loss of fragile sections
- Operates within the cold chamber
Frequently asked questions about Cryostat with Vacuum Section Transfer System
Large whole-mount sections such as whole brain, lung or embryo slices, and thin sections that tear easily. It is also useful when dozens of consecutive sections must be collected in a fixed order without loss.
With correct vacuum and nozzle size it should not. Suction is kept low enough to hold the section without compressing it, and the operator adjusts it for section thickness. Training and a short trial with representative sections are recommended.
Most systems accept standard 25 by 75 mm slides, and some support membrane slides for laser capture microdissection. Check that the slide holder, nozzle height and deposition position suit the slides and adhesion coatings used in your protocol.
A cryostat cuts frozen tissue sections inside a refrigerated chamber. It is used for rapid intraoperative diagnosis, for immunohistochemistry and enzyme work that need unfixed tissue, and for research when native morphology or antigen preservation is important.
Most frozen sections are cut with the chamber between –18 and –25 °C, and the specimen stage matched or set a few degrees colder for dense tissue. Fatty or soft tissue may need –30 °C, while delicate tissue such as lymph node is often cut nearer –15 °C.
Shredding usually means the tissue or chamber is too cold, the block is too thick or the blade is blunt. Crumpling can indicate the tissue is too warm. Adjust chamber and specimen temperature, use a fresh blade, trim to size and cut at a slow, steady speed.