Cryostat with Independent Specimen Cooling
独立标本冷却恒冷箱切片机
- SKU CRY-0004
- Cryostats
- CE (to be confirmed)
- IEC 61010-1 compliance (to be confirmed)
- IEC 61326-1 EMC (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Chamber temperature range | –5 to –35 °C |
|---|---|
| Specimen temperature range | –10 to –40 °C |
| Stage control accuracy | ±1 °C |
| Section thickness range | 1–100 µm |
| Stage freeze-down time | about 15–30 min |
| Blade type | Disposable low-profile |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-101, IEC 61326-1
What is the Cryostat with Independent Specimen Cooling?
A cryostat with independent specimen cooling controls the specimen stage temperature separately from the chamber. It is chosen when tissue types with different optimal cutting temperatures must be handled in one session.
What is the Cryostat with Independent Specimen Cooling?
A cryostat with independent specimen cooling controls the specimen stage temperature separately from the chamber. It is chosen when tissue types with different optimal cutting temperatures must be handled in one session.
Overview
Optimal cutting temperature varies widely: lymph node may cut best near –15 °C, while dense fibrous tissue or fat needs –30 °C or colder. In a conventional cryostat the block can only be cooled to the chamber setpoint. An independent specimen cooling circuit adds a separately controlled stage, usually covering –10 to –40 °C, so the block temperature can be tuned without disturbing the chamber, blade and anti-roll plate environment. This improves section quality for mixed workloads and lets protocols specify exact tissue temperatures for reproducible research. The stage reaches its setpoint faster than the whole chamber, which is useful when several different specimens are cut in sequence. Look for accurate stage control, uniform stage surface temperature and a fast freeze-down feature.
How does it work?
A cryostat places a microtome inside an insulated chamber cooled by one or two refrigeration compressors. The chamber and often a separate specimen stage are brought to between –15 and –35 °C, freezing the tissue to a suitable hardness. The operator attaches the frozen block to a chuck with embedding medium and sections it with a disposable blade held at a fixed angle, producing sections typically 1 to 100 µm thick. Sections adhere to room-temperature slides and are fixed immediately in alcohol or acetone. Anti-roll plates and brush techniques guide ribbons free of curling. Because the whole cutting zone stays cold, tissue morphology is preserved without paraffin infiltration, and total turnaround from specimen to stained slide can be under 20 minutes.
What are the key features?
- Separate specimen stage temperature control
- Stage range about –10 to –40 °C
- Chamber temperature held independently
- Fast stage freeze-down for sequential specimens
- Improved quality for mixed tissue workloads
- Accurate setpoint with uniform stage surface
What are the technical specifications?
A typical cryostat with independent specimen cooling configuration offers chamber temperature range of –5 to –35 °C, specimen temperature range of –10 to –40 °C, stage control accuracy of ±1 °C, section thickness range of 1–100 µm, stage freeze-down time of about 15–30 min and blade type of Disposable low-profile. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Cryostat with Independent Specimen Cooling is used in mixed tissue workloads in one session, immunofluorescence protocols, fatty and dense tissue sections and reproducible research sectioning. Across the wider Cryostats class, laboratories also deploy it for intraoperative frozen sections, immunofluorescence on fresh tissue and enzyme histochemistry.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-101, IEC 61326-1, ISO 15189 and ISO 13485. Certification candidates for this model are CE, IEC 61010-1 compliance and IEC 61326-1 EMC; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the cryostat in a room kept below 25 °C with good ventilation, because condensers reject significant heat. Allow several hours for initial cool-down, then verify chamber and specimen temperatures against a calibrated thermometer. Defrost and clean the chamber on schedule; residual cryo-embedding medium degrades section quality. Use anti-roll plates and change blades often. Monitor for frost build-up, compressor noise and door seal condition, and service annually. Maintain temperature logs, cleaning records and staff training, since these documents support accreditation of frozen section services.
How do I select the right Cryostat with Independent Specimen Cooling?
Choose independent specimen cooling when one session covers tissues with different optimal cutting temperatures, or when a research protocol fixes tissue temperature. Confirm the stage range, accuracy and freeze-down time, and check that the chamber remains stable while the stage changes. It is not essential for single-tissue clinical lists.
How does it compare with other Cryostats variants?
Choose a cryostat by the work it must support. Clinical laboratories doing intraoperative frozen sections need fast cool-down, a stable chamber near –20 °C, reliable blade cooling and often UV sterilisation for contamination control. Research and immunohistochemistry laboratories value accurate independent specimen cooling, wide temperature range to –35 °С and a motorised or ergonomic cutting system for repetitive work. Confirm chamber and specimen temperature ranges, cool-down time, defrost cycle, section thickness range, blade type and whether the cutting area can be cleaned with the chamber at working temperature. Also check section transfer options and the footprint of the lab bench.
Full specifications
| Chamber temperature range | –5 to –35 °C |
|---|---|
| Specimen temperature range | –10 to –40 °C |
| Stage control accuracy | ±1 °C |
| Section thickness range | 1–100 µm |
| Stage freeze-down time | about 15–30 min |
| Blade type | Disposable low-profile |
| Defrost | Programmable cycle |
| Cool-down time | about 2 h |
| Power supply | 220 V, 50 Hz, about 1.2 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Mixed tissue workloads in one session
- Immunofluorescence protocols
- Fatty and dense tissue sections
- Reproducible research sectioning
Key features
- Separate specimen stage temperature control
- Stage range about –10 to –40 °C
- Chamber temperature held independently
- Fast stage freeze-down for sequential specimens
- Improved quality for mixed tissue workloads
- Accurate setpoint with uniform stage surface
Frequently asked questions about Cryostat with Independent Specimen Cooling
Because hardness depends on water and lipid content. Fat and fibrous tissue must be colder to cut cleanly, while lymph node and brain are better slightly warmer to avoid shattering. A single chamber temperature cannot suit all of them in one run.
No. The chamber stays at its setpoint and only the specimen stage changes, which reaches its new temperature in 15 to 30 minutes. This avoids the long re-stabilisation that would follow a full chamber temperature change.
Check it with a calibrated thermocouple or resistance thermometer placed on the stage surface, at a defined interval and after any service. Record the reading against the setpoint and investigate deviations beyond the stated accuracy, typically about ±1 °C.
A cryostat cuts frozen tissue sections inside a refrigerated chamber. It is used for rapid intraoperative diagnosis, for immunohistochemistry and enzyme work that need unfixed tissue, and for research when native morphology or antigen preservation is important.
Most frozen sections are cut with the chamber between –18 and –25 °C, and the specimen stage matched or set a few degrees colder for dense tissue. Fatty or soft tissue may need –30 °C, while delicate tissue such as lymph node is often cut nearer –15 °C.
Shredding usually means the tissue or chamber is too cold, the block is too thick or the blade is blunt. Crumpling can indicate the tissue is too warm. Adjust chamber and specimen temperature, use a fresh blade, trim to size and cut at a slow, steady speed.