Digital Tissue Flotation Bath
数显展片水浴
- SKU TFB-0001
- Tissue Float Baths
- CE (to be confirmed)
- IEC 61010-1 compliance (to be confirmed)
- ISO 13485 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | ambient to 90 °C |
|---|---|
| Temperature stability | ±1 °C |
| Display resolution | 0.1 °C |
| Bath volume | about 1.5–3 L |
| Working area | about 200 × 150 mm |
| Water depth | 40–70 mm |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-010, IEC 61326-1
What is the Digital Tissue Flotation Bath?
A digital tissue flotation bath uses a microprocessor controller and display to hold water within about ±1 °C of the setpoint. It is preferred where temperature must be documented and reproduced across operators.
What is the Digital Tissue Flotation Bath?
A digital tissue flotation bath uses a microprocessor controller and display to hold water within about ±1 °C of the setpoint. It is preferred where temperature must be documented and reproduced across operators.
Overview
The digital flotation bath replaces the dial of an analog unit with a thermistor or platinum sensor, a microprocessor controller and a digital readout, usually settable from ambient to 90 °C with a resolution of 0.1 or 1 °C. Accuracy of about ±1 °C and a stable setpoint remove the guesswork that causes wrinkled or over-melted sections. Most models offer an over-temperature cut-out and a low-water indicator, and some include a timer or audible alerts. A black or grey coated basin makes pale ribbons visible, and a drain valve simplifies daily water changes. Digital control suits laboratories under quality systems because the measured temperature can be read, checked and logged, and because new staff reach the correct setting without experience.
How does it work?
A tissue float bath is a thermostatted shallow basin, usually stainless steel or coated aluminium, heated by a resistive element under the base. Water is held at about 40 to 45 °C, comfortably above the paraffin melting point of roughly 56 to 58 °C when the water is only warming and below it when the surface is cooler, so the wax softens just enough to relax. The technician releases the section ribbon onto the water surface, where surface tension and gentle heat stretch it flat. Sections are then floated onto a clean, often albumin or polylysine-coated slide and drained. Because paraffin expands slightly on melting, the ribbon appears to relax, and gentle agitation with a brush helps remove trapped folds without tearing. Temperature stability and water purity determine whether sections spread evenly.
What are the key features?
- Microprocessor control with digital temperature display
- Temperature stability of about ±1 °C
- Over-temperature cut-out and low-water indication
- Coated dark basin improves ribbon visibility
- Drain valve for easy daily water changes
- Setpoint adjustable up to about 90 °C
What are the technical specifications?
A typical digital tissue flotation bath configuration offers temperature range of ambient to 90 °C, temperature stability of ±1 °C, display resolution of 0.1 °C, bath volume of about 1.5–3 L, working area of about 200 × 150 mm and water depth of 40–70 mm. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Digital Tissue Flotation Bath is used in routine paraffin histology, laboratories under ISO 15189, research histology and high-volume biopsy services. Across the wider Tissue Float Baths class, laboratories also deploy it for paraffin section flattening, routine histology and research histology.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-010, IEC 61326-1 and ISO 15189. Certification candidates for this model are CE, IEC 61010-1 compliance and ISO 13485 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Fill with purified or demineralised water, because mineral deposits stain sections and leave residue on slides. Change water daily or more often if many sections are floated, and clean the basin to prevent biofilm and carryover. Verify the setpoint with a calibrated thermometer at defined intervals, since a drift of a few degrees can wrinkle or melt sections. Place the bath on a stable bench away from draughts and keep the cable away from water. Record temperature checks, water changes and cleaning, and do not use the bath for other laboratory purposes where cross-contamination could occur.
How do I select the right Digital Tissue Flotation Bath?
Choose a digital bath when temperature must be reproducible and recorded, when several operators share the instrument, or when protocols specify an exact flotation temperature. Confirm accuracy, resolution, basin dimensions, warm-up time and whether a drain valve and low-water alarm are fitted. Specify the basin coating colour for ribbon visibility.
How does it compare with other Tissue Float Baths variants?
Select a float bath mainly by control type, size and how it will be used alongside other histology equipment. Digital models hold temperature within about ±1 °C and suit laboratories that must document controlled conditions; analog dial models cost less but need more attention. Choose a bath deep enough to submerge the ribbon fully, typically 40 to 70 mm of water, and wide enough for your largest sections. Illuminated black or blue backgrounds make pale ribbons easier to see. If space is limited or the workflow is high-throughput, a combination unit that pairs a float bath with a slide warmer saves bench space. Check warm-up time, wetted-material corrosion resistance, drain design and whether a lid is needed to limit evaporation.
Full specifications
| Temperature range | ambient to 90 °C |
|---|---|
| Temperature stability | ±1 °C |
| Display resolution | 0.1 °C |
| Bath volume | about 1.5–3 L |
| Working area | about 200 × 150 mm |
| Water depth | 40–70 mm |
| Warm-up time | about 20–30 min |
| Safety features | Over-temperature cut-out, low-water alarm |
| Power supply | 220 V, 50 Hz, about 0.5 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Routine paraffin histology
- Laboratories under ISO 15189
- Research histology
- High-volume biopsy services
Key features
- Microprocessor control with digital temperature display
- Temperature stability of about ±1 °C
- Over-temperature cut-out and low-water indication
- Coated dark basin improves ribbon visibility
- Drain valve for easy daily water changes
- Setpoint adjustable up to about 90 °C
Frequently asked questions about Digital Tissue Flotation Bath
About ±1 °C is sufficient for routine work, since the usable window is roughly 40 to 45 °C. Tighter control helps when protocols specify a temperature, and it reduces variation between operators and between days.
A 0.1 °C display is convenient for setting and verifying the flotation temperature, while a 1 °C display is usually adequate for routine use. Resolution of the display does not guarantee accuracy, so verify with a calibrated thermometer.
Yes. Evaporation during a working day can uncover the element, causing uneven heating or damage. A low-water indicator or cut-out protects the bath and warns the operator to top up with purified water before sections are floated.
Most laboratories set the bath between 40 and 45 °C, which softens the paraffin enough to relax the ribbon without melting it away. If sections spread poorly the water is too cool, while sections that disintegrate or show holes usually mean the water is too warm.
Use purified or demineralised water. Tap water leaves mineral deposits that can stain sections, contaminate slides and encourage biofilm in the basin. Change the water daily, and more often when many ribbons are floated.
Mount them promptly, usually within a minute or two. Prolonged floating lets the section hydrate and swell, causing folds, holes or loss of detail, and increases the risk of cross-contamination between specimens if water is not changed.