Perchloric Acid Hoods

高氯酸通风柜

PAH · Fume Hoods · 5 models

By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 14175-2 (fume cupboards, safety and performance), EN 14175-3 (type test methods), ASHRAE 110 (method of testing performance of laboratory fume hoods)

Standards referenced

  • EN 14175-2 (fume cupboards, safety and performance)
  • EN 14175-3 (type test methods)
  • ASHRAE 110 (method of testing performance of laboratory fume hoods)
  • NFPA 45 (fire protection for laboratories using chemicals)
  • GB/T 34269
  • IEC 61010-1

What are Perchloric Acid Hoods?

Perchloric acid hoods are dedicated fume hoods with water washdown systems and smooth, all-welded ductwork designed for digestions using perchloric acid. They prevent the accumulation of explosive perchlorate salts in the duct and are used in trace-metal and environmental laboratories.

5 models of Perchloric Acid Hoods

The 5 models below are current configurations; specifications are confirmed on quotation.

Water-Washdown Perchloric Acid Hood

水冲洗高氯酸通风柜

  • PAH-0001
  • Perchloric Acid Hoods
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Body material:
Polypropylene, PVC or stainless steel
Washdown coverage:
Interior, baffles and full duct
Washdown cycle:
3–10 min, timer controlled
Face velocity:
0.4–0.6 m/s
Airflow:
600–1500 m3/h
Width:
1200–1800 mm

A water-washdown perchloric acid hood floods the interior and full duct with water on a timed cycle to dissolve perchlorate deposits. It is the baseline design for any perchloric acid digestion application.

Request a quote
View details

Walk-In Perchloric Acid Hood

步入式高氯酸通风柜

  • PAH-0002
  • Perchloric Acid Hoods
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Body material:
Polypropylene or 316L stainless
Height:
2000–2500 mm
Width:
1500–2400 mm
Depth:
900–1500 mm
Face velocity:
0.4–0.6 m/s
Airflow:
1000–2500 m3/h

A walk-in perchloric acid hood provides floor-level access and full washdown for large digestion racks and tall glassware. It is used for high-throughput perchloric acid digestions in environmental and geological research laboratories.

Request a quote
View details

Benchtop Perchloric Acid Hood

台式高氯酸通风柜

  • PAH-0003
  • Perchloric Acid Hoods
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Body material:
Polypropylene or 316L stainless
Width:
900–1500 mm
Depth:
700–900 mm
Vessel capacity:
1–6 positions
Face velocity:
0.4–0.6 m/s
Airflow:
300–700 m3/h

A benchtop perchloric acid hood is a compact unit for one to six vessels with full washdown and a dedicated duct. It suits small perchloric acid digestions where floor space is limited.

Request a quote
View details

VAV Perchloric Acid Hood

变风量高氯酸通风柜

  • PAH-0004
  • Perchloric Acid Hoods
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Control type:
Variable air volume (VAV)
Face velocity:
0.4–0.6 m/s, controlled
Airflow range:
300–1500 m3/h
Sash sensor:
Linear position
Washdown cycle:
3–10 min, full duct
Body material:
Polypropylene or 316L

A VAV perchloric acid hood controls exhaust volume to keep face velocity constant as the sash moves, while retaining full washdown. It reduces energy use in laboratories with several perchloric exhaust points.

Request a quote
View details

Perchloric Acid Digestion Hood with Scrubber

洗涤式高氯酸消解通风柜

  • PAH-0005
  • Perchloric Acid Hoods
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
Scrubber type:
Packed column, wet
Removal efficiency:
90–99 % for acid mist
Face velocity:
0.4–0.6 m/s
Airflow:
600–2000 m3/h
Washdown cycle:
3–10 min, duct and scrubber
Liquid-to-gas ratio:
2–10 L/m3

A perchloric acid digestion hood with scrubber adds a wet scrubbing stage to the washdown duct to capture acid mist before discharge. It is used where perchloric acid emissions are limited and acid volumes are large.

Request a quote
View details

Compare Perchloric Acid Hoods models

ModelBody materialWashdown cycleFace velocityAirflowWidth
Water-Washdown Perchloric Acid HoodPAH-0001Polypropylene, PVC or stainless steel3–10 min, timer controlled0.4–0.6 m/s600–1500 m3/h1200–1800 mm
Walk-In Perchloric Acid HoodPAH-0002Polypropylene or 316L stainless5–15 min, full interior and duct0.4–0.6 m/s1000–2500 m3/h1500–2400 mm
Benchtop Perchloric Acid HoodPAH-0003Polypropylene or 316L stainless3–8 min, interior and duct0.4–0.6 m/s300–700 m3/h900–1500 mm
VAV Perchloric Acid HoodPAH-0004Polypropylene or 316L3–10 min, full duct0.4–0.6 m/s, controlled——
Perchloric Acid Digestion Hood with ScrubberPAH-0005—3–10 min, duct and scrubber0.4–0.6 m/s600–2000 m3/h—

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

How do Perchloric Acid Hoods work?

Perchloric acid hoods work like other fume hoods but add a washdown system that periodically sprays water through the interior, the rear baffle and the entire exhaust duct. Heated perchloric acid vaporises, and its vapour can condense with organic material to form perchlorate salts in the duct, which are shock-sensitive and can explode. Water washdown dissolves these deposits before they accumulate. The hood and duct are constructed from non-combustible, chemically resistant materials with smooth, coved and crevice-free interiors and no sharp bends or horizontal runs where liquid or salt could collect. The washdown cycle runs for several minutes, typically daily or after each use, and drains to a dedicated chemical waste line. A dedicated exhaust system, separate from other hoods, prevents cross contamination and dilution problems.

Which Perchloric Acid Hoods variants are available?

  • Water-Washdown Perchloric Acid Hood — A water-washdown perchloric acid hood floods the interior and full duct with water on a timed cycle to dissolve perchlorate deposits.
  • Walk-In Perchloric Acid Hood — A walk-in perchloric acid hood provides floor-level access and full washdown for large digestion racks and tall glassware.
  • Benchtop Perchloric Acid Hood — A benchtop perchloric acid hood is a compact unit for one to six vessels with full washdown and a dedicated duct.
  • VAV Perchloric Acid Hood — A VAV perchloric acid hood controls exhaust volume to keep face velocity constant as the sash moves, while retaining full washdown.
  • Perchloric Acid Digestion Hood with Scrubber — A perchloric acid digestion hood with scrubber adds a wet scrubbing stage to the washdown duct to capture acid mist before discharge.

How do I choose the right variant?

Every perchloric acid hood must have a full washdown system covering the interior and the whole duct, because partial washdown leaves deposition sites. Choose a dedicated exhaust path with no shared ductwork, and specify smooth, all-welded construction with coved corners and no internal fasteners or ledges. Select the liner for the acids used: polypropylene or PVC for mixed acid work, and stainless steel or special alloys where hot concentrated perchloric acid is handled. Match the internal height to reflux glassware and the digestion block, and select airflow so face velocity stays at 0.4–0.6 m/s. Include a washdown timer or automatic cycle, drainage with an acid-resistant trap, and airflow monitoring with an alarm. Confirm the duct material, slope and washdown coverage in writing.

Where are Perchloric Acid Hoods used?

Perchloric Acid Hoods are specified in trace-metal digestion with perchloric acid, environmental and geological analysis, food and biological sample decomposition, metallurgical and alloy analysis and forensic and toxicology digestion.

Which standards and certifications apply?

Work with Perchloric Acid Hoods is commonly governed by EN 14175-2 (fume cupboards, safety and performance), EN 14175-3 (type test methods), ASHRAE 110 (method of testing performance of laboratory fume hoods), NFPA 45 (fire protection for laboratories using chemicals), GB/T 34269 and IEC 61010-1. Instrument certifications such as CE marking are confirmed per configuration and never assumed.

What are the key considerations before purchase?

Never share the duct with other hoods, and never use the hood for organic solvents or for drying organic material, since perchloric acid and organics can form explosive mixtures. Run the washdown cycle before and after work and keep a log of cycles and cleaning. Inspect the duct interior for salt deposits on a defined schedule, using a borescope where needed, and clean any deposit with water rather than scraping. Keep all surfaces smooth and free of ledges, gaskets or insulation that could collect salts. Verify face velocity and alarms regularly, and train users to keep glassware away from baffles. Treat washdown effluent as acidic waste and confirm the trap and drain are compatible.

Frequently asked questions about Perchloric Acid Hoods

Heated perchloric acid vapour can condense with organic material to form perchlorate salts, which are shock-sensitive and may explode. Water washdown dissolves these deposits from the hood and duct before they accumulate.

Need help specifying Perchloric Acid Hoods?