Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer
惰性气体熔融氧氮氢分析仪
- SKU MET-0003
- Metal Analysis
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Measured elements | O, N, H, singly or combined |
|---|---|
| Oxygen range | From about 1 mg/kg to 0.5 % |
| Nitrogen range | From about 1 mg/kg to 0.5 % |
| Hydrogen range | From about 0.1 mg/kg to 50 mg/kg |
| Furnace temperature | Up to 3000 °C |
| Sample mass | 0.1–1 g |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E415, ASTM E1086, ASTM E1019
What is the Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer?
An inert gas fusion analyzer melts metal samples in a graphite crucible under inert gas and measures the released oxygen, nitrogen and hydrogen. It quantifies dissolved gases that affect steel cleanliness and weld quality.
What is the Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer?
An inert gas fusion analyzer melts metal samples in a graphite crucible under inert gas and measures the released oxygen, nitrogen and hydrogen. It quantifies dissolved gases that affect steel cleanliness and weld quality.
Overview
A weighed sample, often in a nickel or tin capsule with a flux, is dropped into a graphite crucible heated by an electrode furnace to 2000 to 3000 °C. The metal melts and reacts with the crucible carbon, so oxygen is released as carbon monoxide or dioxide while nitrogen and hydrogen are released as gases. A carrier gas of helium or argon sweeps them to detectors: infrared cells for oxygen, and thermal conductivity cells for nitrogen and hydrogen after separation columns. A single instrument can measure one gas or all three depending on configuration, and analysis takes one to three minutes per sample. Calibration uses certified gas standards or reference materials analysed under identical conditions, and crucible, flux and carrier gas blanks must be controlled because measured concentrations are typically in the single or tens of mg/kg range.
How does it work?
Spark optical emission spectrometry is the workhorse for solid metal. An argon-shielded electric discharge vaporises and excites atoms at the sample surface, the emitted light is dispersed by a spectrometer, and intensities are converted to concentrations using calibrations built from certified reference materials of similar composition. X-ray fluorescence offers the same idea without the need for a conductive, flat surface: a beam excites inner-shell electrons, and the characteristic fluorescence identifies and quantifies elements, which makes handheld units useful for alloy sorting. Carbon, sulfur, oxygen, nitrogen and hydrogen are not accessible to optical emission at low levels and are determined by combustion or inert gas fusion: the sample melts in a crucible under a stream of oxygen or inert gas, gases are released and measured by infrared cells or thermal conductivity. Inductively coupled plasma optical emission spectrometry is used after acid dissolution for trace elements and for certification work. All these techniques depend on calibration, matrix matching and clean, representative samples.
What are the key features?
- Electrode or resistance furnace reaching 2000–3000 °C
- Graphite crucibles with flux and capsule loading
- Infrared detector for oxygen and thermal conductivity for nitrogen and hydrogen
- Separation columns for gas speciation
- Carrier gas flow and blank monitoring
- Automatic crucible loading on some models
What are the technical specifications?
A typical inert gas fusion oxygen-nitrogen-hydrogen analyzer configuration offers measured elements of O, N, H, singly or combined, oxygen range of From about 1 mg/kg to 0.5 %, nitrogen range of From about 1 mg/kg to 0.5 %, hydrogen range of From about 0.1 mg/kg to 50 mg/kg, furnace temperature of Up to 3000 °C and sample mass of 0.1–1 g. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer is used in oxygen content control for steel cleanliness, hydrogen determination to prevent cracking, nitrogen control in stainless and alloy steels and titanium and superalloy gas content testing. Across the wider Metal Analysis class, laboratories also deploy it for steel and alloy grade verification, foundry melt control and charge optimisation and positive material identification of installed parts.
Which standards and certifications apply?
Relevant standards and methods include ASTM E415, ASTM E1086, ASTM E1019, ASTM E1479, ISO 14284, ASTM E1806 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Everything depends on calibration and reference materials, so keep a set of certified standards covering each element range and renew them before they expire. Condition the sample surface correctly: machined for spark emission, flat and clean for X-ray fluorescence, uncontaminated chips for combustion. Reference the instrument daily with a control sample and confirm drift before releasing results. Maintain gas supplies at the specified purity, since moisture and oxygen in argon degrade spark stability, and check crucibles, electrodes, filters and reagents on a schedule. Control contamination between samples: clean the spark stand, change crucibles and use new tools. Radiation and laser safety programs, fume extraction for acid digestion and safe handling of compressed gases are part of normal operation, and records of calibration, verification and maintenance should travel with the results.
How do I select the right Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer?
Select the inert gas fusion analyzer when oxygen, nitrogen or hydrogen limits are specified, for example in bearing, pipeline, titanium or superalloy grades. Specify which gases you need, detection limits, crucible and flux consumables and the reference materials used for calibration.
How does it compare with other Metal Analysis variants?
Decide from the elements, the levels and the sample form. For production control of solid metal, spark optical emission gives fast multi-element results from a machined surface and is the usual choice, with the spectrometer configuration matched to the base metals and element ranges you analyse. Combustion and inert gas fusion instruments are required when carbon, sulfur, oxygen, nitrogen or hydrogen form part of the specification, and they need their own sample preparation and reference materials. Handheld X-ray fluorescence suits incoming inspection, alloy sorting and positive material identification on installed parts, but cannot measure carbon or low atomic number elements and needs careful radiation management. Inductively coupled plasma is preferred for trace and tramp elements, for non-routine matrices and for validating other methods. Check detection limits, the number of elements and channels, calibration standards available, argon or gas supply and service support before committing.
Full specifications
| Measured elements | O, N, H, singly or combined |
|---|---|
| Oxygen range | From about 1 mg/kg to 0.5 % |
| Nitrogen range | From about 1 mg/kg to 0.5 % |
| Hydrogen range | From about 0.1 mg/kg to 50 mg/kg |
| Furnace temperature | Up to 3000 °C |
| Sample mass | 0.1–1 g |
| Analysis time | 1–3 min per sample |
| Carrier gas | Helium or argon, 99.999 % |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Oxygen content control for steel cleanliness
- Hydrogen determination to prevent cracking
- Nitrogen control in stainless and alloy steels
- Titanium and superalloy gas content testing
Key features
- Electrode or resistance furnace reaching 2000–3000 °C
- Graphite crucibles with flux and capsule loading
- Infrared detector for oxygen and thermal conductivity for nitrogen and hydrogen
- Separation columns for gas speciation
- Carrier gas flow and blank monitoring
- Automatic crucible loading on some models
Frequently asked questions about Inert Gas Fusion Oxygen-Nitrogen-Hydrogen Analyzer
Carbon in the crucible reduces oxides in the molten metal, releasing oxygen as carbon monoxide, which can then be measured by infrared cells. The crucible also provides the reducing environment needed to liberate nitrogen and hydrogen.
The carrier gas stream passes first through a detector or a column, depending on the configuration: oxygen is measured directly by infrared absorption, while nitrogen and hydrogen are separated chromatographically and quantified by thermal conductivity.
Blank contributions from crucible, flux and carrier gas, surface contamination on the sample, and the condition of the furnace and columns. Certified standards analysed under identical settings establish the calibration.
Spark optical emission spectrometry for solid metal, because it measures many elements quickly from a machined surface. Carbon, sulfur and dissolved gases are added by combustion or inert gas fusion analyzers, since optical emission cannot reach the required low levels.
Yes, for grades distinguished by elements from about magnesium upward, which covers most stainless and alloy steels, nickel alloys and copper alloys. Carbon, silicon and other light elements cannot be measured, so some grades require laboratory confirmation.
By combustion: a weighed sample burns with an accelerator in an oxygen stream in an induction furnace, and the carbon dioxide and sulfur dioxide released are measured with infrared cells. Results are reported in percent or in mg/kg for low levels.