Spark Optical Emission Spectrometer for Metal Analysis
金属分析火花发射光谱仪
- SKU MET-0001
- Metal Analysis
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Wavelength range | 165–800 nm |
|---|---|
| Elements measured | 20–30 simultaneously |
| Detection limit | From about 1–10 mg/kg for many elements |
| Measurement time | 20–40 s including prespark |
| Argon purity | ≥ 99.999 % |
| Sample surface | Machined flat, Ra about 1–3 µm |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E415, ASTM E1086, ASTM E1019
What is the Spark Optical Emission Spectrometer for Metal Analysis?
A spark optical emission spectrometer excites a machined metal surface with an argon-shielded discharge and measures the emitted light to determine many elements at once. It is the standard instrument for steel and alloy composition control.
What is the Spark Optical Emission Spectrometer for Metal Analysis?
A spark optical emission spectrometer excites a machined metal surface with an argon-shielded discharge and measures the emitted light to determine many elements at once. It is the standard instrument for steel and alloy composition control.
Overview
A tungsten electrode strikes an arc or spark on the sample surface under an argon flow, vaporising a small volume of metal and exciting the atoms so they emit light at element-specific wavelengths. The light is separated by a polychromator or a set of photomultiplier and CCD channels, and intensities are converted to concentrations using calibrations from certified reference materials with similar base composition. Modern instruments cover twenty or more elements simultaneously, typically from low mg/kg to tens of percent, and a full measurement takes twenty to forty seconds after a prespark period that conditions the surface. Standardisation with reference materials corrects for drift, and type standardisation allows specific grades to be checked against a single standard. Results are used for melt control, grade verification and certificate of analysis, and the method is defined in standards such as ASTM E415 and ASTM E1086.
How does it work?
Spark optical emission spectrometry is the workhorse for solid metal. An argon-shielded electric discharge vaporises and excites atoms at the sample surface, the emitted light is dispersed by a spectrometer, and intensities are converted to concentrations using calibrations built from certified reference materials of similar composition. X-ray fluorescence offers the same idea without the need for a conductive, flat surface: a beam excites inner-shell electrons, and the characteristic fluorescence identifies and quantifies elements, which makes handheld units useful for alloy sorting. Carbon, sulfur, oxygen, nitrogen and hydrogen are not accessible to optical emission at low levels and are determined by combustion or inert gas fusion: the sample melts in a crucible under a stream of oxygen or inert gas, gases are released and measured by infrared cells or thermal conductivity. Inductively coupled plasma optical emission spectrometry is used after acid dissolution for trace elements and for certification work. All these techniques depend on calibration, matrix matching and clean, representative samples.
What are the key features?
- Argon shielded spark stand with tungsten electrode
- Simultaneous multi-element channels or CCD optics
- Prespark, integration and standardisation programs
- Calibration with certified reference materials
- Type standardisation for specific alloy grades
- Reconditioning of the electrode and spark stand
What are the technical specifications?
A typical spark optical emission spectrometer for metal analysis configuration offers wavelength range of 165–800 nm, elements measured of 20–30 simultaneously, detection limit of From about 1–10 mg/kg for many elements, measurement time of 20–40 s including prespark, argon purity of ≥ 99.999 % and sample surface of Machined flat, Ra about 1–3 µm. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Spark Optical Emission Spectrometer for Metal Analysis is used in melt control in steel and foundry production, final product grade verification, incoming inspection of alloy stock and certification of reference and control samples. Across the wider Metal Analysis class, laboratories also deploy it for steel and alloy grade verification, foundry melt control and charge optimisation and positive material identification of installed parts.
Which standards and certifications apply?
Relevant standards and methods include ASTM E415, ASTM E1086, ASTM E1019, ASTM E1479, ISO 14284, ASTM E1806 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Everything depends on calibration and reference materials, so keep a set of certified standards covering each element range and renew them before they expire. Condition the sample surface correctly: machined for spark emission, flat and clean for X-ray fluorescence, uncontaminated chips for combustion. Reference the instrument daily with a control sample and confirm drift before releasing results. Maintain gas supplies at the specified purity, since moisture and oxygen in argon degrade spark stability, and check crucibles, electrodes, filters and reagents on a schedule. Control contamination between samples: clean the spark stand, change crucibles and use new tools. Radiation and laser safety programs, fume extraction for acid digestion and safe handling of compressed gases are part of normal operation, and records of calibration, verification and maintenance should travel with the results.
How do I select the right Spark Optical Emission Spectrometer for Metal Analysis?
Select the spark spectrometer as the primary composition instrument where solid metal samples are routine, and specify the base metals, element ranges and detection limits you require. Check the gas supply, sample preparation equipment and the reference materials supplied with the calibration.
How does it compare with other Metal Analysis variants?
Decide from the elements, the levels and the sample form. For production control of solid metal, spark optical emission gives fast multi-element results from a machined surface and is the usual choice, with the spectrometer configuration matched to the base metals and element ranges you analyse. Combustion and inert gas fusion instruments are required when carbon, sulfur, oxygen, nitrogen or hydrogen form part of the specification, and they need their own sample preparation and reference materials. Handheld X-ray fluorescence suits incoming inspection, alloy sorting and positive material identification on installed parts, but cannot measure carbon or low atomic number elements and needs careful radiation management. Inductively coupled plasma is preferred for trace and tramp elements, for non-routine matrices and for validating other methods. Check detection limits, the number of elements and channels, calibration standards available, argon or gas supply and service support before committing.
Full specifications
| Wavelength range | 165–800 nm |
|---|---|
| Elements measured | 20–30 simultaneously |
| Detection limit | From about 1–10 mg/kg for many elements |
| Measurement time | 20–40 s including prespark |
| Argon purity | ≥ 99.999 % |
| Sample surface | Machined flat, Ra about 1–3 µm |
| Base metals supported | Fe, Ni, Cu, Al, Co, Ti bases |
| Data output | LAN, LIMS export |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Melt control in steel and foundry production
- Final product grade verification
- Incoming inspection of alloy stock
- Certification of reference and control samples
Key features
- Argon shielded spark stand with tungsten electrode
- Simultaneous multi-element channels or CCD optics
- Prespark, integration and standardisation programs
- Calibration with certified reference materials
- Type standardisation for specific alloy grades
- Reconditioning of the electrode and spark stand
Frequently asked questions about Spark Optical Emission Spectrometer for Metal Analysis
By running a control sample at the start of each session, and by re-standardising with one or more certified reference materials when the control drifts outside limits. Type standardisation re-calibrates for a specific grade using a single standard.
Moisture, oxygen and nitrogen in the argon change the discharge conditions and absorb ultraviolet light, which affects sensitivity and precision. Impure gas causes unstable intensities, poor reproducibility and shifted calibration.
A flat machined surface free from cracks, porosity, scale and coatings, produced by milling or turning, with thickness sufficient to dissipate heat. Preparation quality usually limits precision more than the instrument itself.
Spark optical emission spectrometry for solid metal, because it measures many elements quickly from a machined surface. Carbon, sulfur and dissolved gases are added by combustion or inert gas fusion analyzers, since optical emission cannot reach the required low levels.
Yes, for grades distinguished by elements from about magnesium upward, which covers most stainless and alloy steels, nickel alloys and copper alloys. Carbon, silicon and other light elements cannot be measured, so some grades require laboratory confirmation.
By combustion: a weighed sample burns with an accelerator in an oxygen stream in an induction furnace, and the carbon dioxide and sulfur dioxide released are measured with infrared cells. Results are reported in percent or in mg/kg for low levels.