Skip to main content
Lab Manufacturers China
Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water — product photo

Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water

水中阴离子分析用抑制电导离子色谱仪

  • SKU IOA-0002
  • Ion Analyzers
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Pump pressure limit35 MPa (5000 psi)
Flow rate range0.2–2.0 mL/min
EluentCarbonate/bicarbonate or KOH (electrolytic)
Injection volume10–100 µL
DetectorSuppressed conductivity, 0–15,000 µS/cm
Column oven temperatureAmbient +5 to 60 °C
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 10304-1, ISO 10304-2, EPA Method 300.1

What is the Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water?

An ion chromatograph configured with an anion-exchange column, carbonate or hydroxide eluent, chemical or electrolytic suppression and a conductivity detector. It determines fluoride, chloride, nitrite, bromide, nitrate, phosphate and sulfate in one 15 to 25 minute run at µg/L to mg/L levels.

What is the Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water?

An ion chromatograph configured with an anion-exchange column, carbonate or hydroxide eluent, chemical or electrolytic suppression and a conductivity detector. It determines fluoride, chloride, nitrite, bromide, nitrate, phosphate and sulfate in one 15 to 25 minute run at µg/L to mg/L levels.

Overview

This variant is the standard configuration for drinking water and environmental anion testing under ISO 10304-1 and EPA 300.1. The system comprises an isocratic or gradient pump delivering 0.2–2.0 mL/min at up to 35 MPa, an injection valve with a 10–100 µL loop, a guard and analytical column, a suppressor and a thermostatted conductivity cell. Electrolytic eluent generation with a hydroxide cartridge removes manual eluent preparation and improves baseline stability. Detection limits are typically 1–20 µg/L per anion with RSD below 2% for retention time and below 3% for peak area. An autosampler with 50–100 positions and 0.45 µm inline filtration supports overnight runs. Software handles integration, calibration, reporting and audit trails. Bromate or perchlorate at sub-µg/L need large loop injection or a dedicated column, and chromatograms can be affected by very high chloride in brackish samples.

How does it work?

Two technologies dominate. Ion-selective electrode (ISE) analyzers use a membrane that responds selectively to one ion, such as a LaF3 crystal for fluoride. The potential against a reference electrode follows the Nernst equation, changing about 59 mV per decade for monovalent ions at 25 °C, and concentration is found from calibration with standards and an ionic strength adjuster. Ion chromatography (IC) separates ions by injecting the sample onto an ion-exchange column, where an eluent such as carbonate-bicarbonate or hydroxide moves them at different rates. A suppressor reduces eluent background conductivity, and a conductivity detector records each ion as a peak whose retention time identifies it and whose area gives concentration. Autosamplers, eluent generators and chromatography software automate multi-ion runs.

What are the key features?

  • Electrolytic hydroxide eluent generation removes manual eluent preparation
  • Suppressed conductivity detection gives low background and µg/L sensitivity
  • Thermostatted column compartment stabilizes retention times
  • Autosampler with 50 to 100 positions and inline 0.45 µm filtration
  • Gradient capability for resolving fluoride from early-eluting organic acids
  • Chromatography software with audit trail and method templates for ISO and EPA methods

What are the technical specifications?

A typical suppressed-conductivity ion chromatograph for anion analysis in water configuration offers pump pressure limit of 35 MPa (5000 psi), flow rate range of 0.2–2.0 mL/min, eluent of Carbonate/bicarbonate or KOH (electrolytic), injection volume of 10–100 µL, detector of Suppressed conductivity, 0–15,000 µS/cm and column oven temperature of Ambient +5 to 60 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water is used in drinking water anion compliance under GB 5749, groundwater and surface water surveys, wastewater chloride, sulfate and nutrient profiling and condensate and boiler water analysis. Across the wider Ion Analyzers class, laboratories also deploy it for drinking water anion compliance testing, fluoride and nitrate monitoring and boiler and cooling water chemistry.

Which standards and certifications apply?

Relevant standards and methods include ISO 10304-1, ISO 10304-2, EPA Method 300.1, ASTM D4327, ASTM D1179, HJ 84, GB/T 5750.5 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

ISE systems need stable temperature, fresh standards and proper electrode conditioning, with slope checks in the 54–60 mV/decade range and membrane replacement when response slows. Ion chromatographs require deionized water of 18.2 MΩ·cm, filtered eluents degassed to avoid baseline noise, and sample filtration through 0.45 µm filters to protect the column. Keep guard columns in place and flush the system when storing. Calibrate with at least five standards and verify with a mid-range check standard each batch. Suppressors and columns have finite life, so track system pressure and retention times. Laboratories working to ISO/IEC 17025 should keep calibration records and control charts.

How do I select the right Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water?

Select the anion ion chromatograph when a full anion panel is reported, detection limits are in µg/L or interferences rule out ISE. If only cations are needed, choose the cation configuration. In an RFQ, specify target anions and ranges, eluent generation preference, autosampler size, column guard requirements, reference method and regulatory audit-trail needs.

How does it compare with other Ion Analyzers variants?

The choice starts with how many ions you need and at what limit. For one or two ions such as fluoride or ammonium in routine samples, an ISE analyzer is fast and inexpensive. For simultaneous determination of chloride, nitrate, sulfate, phosphate and other anions, or cations like sodium, potassium and calcium, ion chromatography is required because it resolves interferences and gives full profiles at µg/L levels. Consider sample throughput and whether an autosampler is needed, as well as matrix complexity, since high-salt samples may require dilution or matrix elimination. Field work favors portable ISE meters, and continuous process control favors online analyzers with automated calibration. Confirm the reference method, such as ISO 10304-1 or HJ 84, and compare running costs for eluents, suppressors, columns and electrodes.

Full specifications

Full specifications
Pump pressure limit35 MPa (5000 psi)
Flow rate range0.2–2.0 mL/min
EluentCarbonate/bicarbonate or KOH (electrolytic)
Injection volume10–100 µL
DetectorSuppressed conductivity, 0–15,000 µS/cm
Column oven temperatureAmbient +5 to 60 °C
Detection limit (Cl−, NO3−, SO42−)1–20 µg/L
Retention time RSD≤ 1%
Typical run time15–25 min
Autosampler capacity50–100 vials

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Drinking water anion compliance under GB 5749
  • Groundwater and surface water surveys
  • Wastewater chloride, sulfate and nutrient profiling
  • Condensate and boiler water analysis

Key features

  • Electrolytic hydroxide eluent generation removes manual eluent preparation
  • Suppressed conductivity detection gives low background and µg/L sensitivity
  • Thermostatted column compartment stabilizes retention times
  • Autosampler with 50 to 100 positions and inline 0.45 µm filtration
  • Gradient capability for resolving fluoride from early-eluting organic acids
  • Chromatography software with audit trail and method templates for ISO and EPA methods

Frequently asked questions about Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water

The suppressor exchanges eluent cations for hydronium ions, converting carbonate eluent to weakly conducting carbonic acid or hydroxide to water. This lowers background conductivity dramatically while analyte ions remain conductive as acids, giving high signal-to-noise and µg/L detection limits.

Related models

All Ion Analyzers

Request a quotation for the Suppressed-Conductivity Ion Chromatograph for Anion Analysis in Water

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote