Spray Chambers for ICP
ICP 喷雾室
- SKU IAC-0003
- ICP Accessories
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- Elemental cleanliness test (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Types | cyclonic, Scott double-pass |
|---|---|
| Materials | glass, quartz, PFA or PTFE |
| Dead volume | 10–50 mL |
| Transport efficiency | 2–8% |
| Washout time | 20–90 s |
| Heating option | up to 120 °C |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <730> Plasma Spectrochemistry, USP <1058> Analytical Instrument Qualification, ISO 11885 Water quality — ICP-OES determination of elements
What is the Spray Chambers for ICP?
Spray chambers remove large droplets from the aerosol before it reaches the plasma. Their design sets transport efficiency, washout time and matrix tolerance, making them key performance parts of the sample introduction path.
What is the Spray Chambers for ICP?
Spray chambers remove large droplets from the aerosol before it reaches the plasma. Their design sets transport efficiency, washout time and matrix tolerance, making them key performance parts of the sample introduction path.
Overview
After the nebulizer generates an aerosol, the spray chamber separates droplets by inertia and gravity, allowing only fine droplets to pass to the torch. Cyclonic chambers use a swirling flow and have low dead volume and fast washout, but slightly lower transport efficiency. Scott double-pass chambers use a longer path with a central tube to improve droplet removal and transport efficiency, at the cost of slower washout. Heated versions reduce solvent load and improve transport for some matrices, and cooled versions condense solvent for organic methods. Chamber material matters for hydrofluoric acid, and inert PFA or PTFE versions are used where glass is attacked. Chamber condition affects sensitivity and carry-over, so cleaning and replacement are part of routine maintenance.
How does it work?
The plasma and sample path are exposed to high temperature, acid fumes and salt deposits, so their components are consumables that wear, corrode or become contaminated. Quartz torches and injectors shape the plasma and must be replaced when they devitrify, crack or accumulate deposits that change sensitivity. Nebulizers and spray chambers set aerosol quality and transport efficiency. RF load coils and torch boxes deliver the plasma power and must be matched to the generator and torch. Peristaltic pump tubing controls sample flow and internal standard addition, and worn tubing changes volumes. Autosampler probes, trays and rinse stations determine carry-over and throughput. Using the correct part number and material for each instrument, element and matrix, and replacing parts on schedule, protects sensitivity, precision and safety.
What are the key features?
- Cyclonic or Scott double-pass geometry
- Low dead volume and fast washout options
- Heated or cooled versions for special matrices
- Glass, quartz or inert polymer construction
- Direct influence on transport efficiency
- Cleaning and replacement as routine maintenance
What are the technical specifications?
A typical spray chambers for icp configuration offers types of cyclonic, Scott double-pass, materials of glass, quartz, PFA or PTFE, dead volume of 10–50 mL, transport efficiency of 2–8%, washout time of 20–90 s and heating option of up to 120 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Spray Chambers for ICP is used in routine aqueous sample introduction, high-volume samples needing fast washout, organic solvent methods with cooling and hydrofluoric acid and inert requirements. Across the wider ICP Accessories class, laboratories also deploy it for replacement torches and injectors for plasma systems, nebulizer and spray chamber maintenance kits and RF coil and torch box service parts.
Which standards and certifications apply?
Relevant standards and methods include USP <730> Plasma Spectrochemistry, USP <1058> Analytical Instrument Qualification, ISO 11885 Water quality — ICP-OES determination of elements, ISO 17294-2 Water quality — ICP-MS determination of elements, IEC 61010-1 and ISO/IEC 17025. Certification candidates for this model are CE, ISO 9001 manufacturing and Elemental cleanliness test; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Accessory condition is the most common reason for drifting sensitivity, noisy baselines and failed quality checks. Establish a maintenance schedule based on sample load and matrix: inspect torches and injectors for deposits and cracks, clean or replace nebulizers when uptake falls, and replace pump tubing and probes at defined intervals. Keep spares of the parts most likely to fail, because a blocked nebulizer or cracked torch stops the laboratory. Handle quartz carefully, store it clean and dry, and never use abrasive tools on injectors or nebulizers. After any replacement, verify performance with a check standard, confirm plasma stability and update the instrument log with the part and date.
How do I select the right Spray Chambers for ICP?
Select a spray chamber to match your nebulizer, matrix and throughput. Choose cyclonic for fast washout and small samples, and double-pass for higher transport efficiency with larger samples. Add heating or cooling for unusual solvent loads, and choose inert materials for hydrofluoric acid. Confirm the drain arrangement suits your waste handling.
How does it compare with other ICP Accessories variants?
Select accessories by exact instrument model, torch type and application. Quartz torches are standard for aqueous and many organic methods, while ceramic or inert torches suit hydrofluoric acid and high-purity work. Injector bore diameter affects aerosol velocity and plasma stability, so match it to the sample introduction system and matrix. Nebulizers range from concentric for clean solutions to V-groove or parallel-path for high solids. Spray chambers differ in transport efficiency and washout time. RF coils and torch boxes must match the generator frequency and power. Pump tubing is chosen by internal diameter and chemical resistance. Autosampler probes and trays must fit the sampler. Confirm part numbers, materials and availability of performance data before ordering.
Full specifications
| Types | cyclonic, Scott double-pass |
|---|---|
| Materials | glass, quartz, PFA or PTFE |
| Dead volume | 10–50 mL |
| Transport efficiency | 2–8% |
| Washout time | 20–90 s |
| Heating option | up to 120 °C |
| Cooling option | down to -10 °C |
| Drain | gravity or pumped |
| Compatible instruments | model specific |
| Cleaning | dilute acid soak |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Routine aqueous sample introduction
- High-volume samples needing fast washout
- Organic solvent methods with cooling
- Hydrofluoric acid and inert requirements
Key features
- Cyclonic or Scott double-pass geometry
- Low dead volume and fast washout options
- Heated or cooled versions for special matrices
- Glass, quartz or inert polymer construction
- Direct influence on transport efficiency
- Cleaning and replacement as routine maintenance
Frequently asked questions about Spray Chambers for ICP
Not directly. Transport efficiency depends on design rather than size, and a chamber that is too large increases dead volume and slows washout. The best choice balances how much sample reaches the plasma against how quickly the signal returns to baseline between samples.
Heating the spray chamber and the aerosol path reduces the solvent load reaching the plasma, which stabilizes the plasma and improves transport of the remaining aerosol. It is useful for organic solvents, for samples with high water content that cool the plasma, and for some high-matrix methods.
Soak glass or quartz chambers in dilute acid and rinse with ultrapure water, or use a suitable solvent for organic residues. Inert polymer chambers tolerate most acids but not all solvents, so check compatibility. Inspect the drain and inner surfaces for residual deposits before reuse.
Replace the torch when it shows devitrification, cracks, a distorted tip, partial melting or deposits that cleaning cannot remove. Symptoms include unstable plasma, poor ignition, rising background and falling sensitivity. Torch life depends on sample matrix, power and cleaning practice, so inspect it at each maintenance interval.
High dissolved solids, high sample uptake and incomplete nebulization leave salt deposits on the injector tip and inside the bore. Deposits change aerosol velocity and can cause signal drift or plasma instability. Diluting samples, using a high-solids nebulizer and cleaning the injector in dilute acid reduce the problem.
Not usually, because torch dimensions, coil geometry, injector adapters and sampler fittings are instrument specific. Some third parties supply compatible parts for common models, but dimensional differences can cause poor plasma performance or damage. Confirm the exact model and part number, and prefer parts verified for your instrument.