Skip to main content
Lab Manufacturers China
Double-Well Glass Bead Sterilizer — product photo

Double-Well Glass Bead Sterilizer

双槽玻璃珠灭菌器

  • SKU GBS-0003
  • Glass Bead Sterilizers
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • EN 61010-1 test report (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Number of wells2
Working temperature250 °C nominal per well
Warm-up time to 250 °C20-30 min
Well diameter40-60 mm each
Well depth40-80 mm each
Bead capacity per well100-200 g
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 61010-1, EN 61010-2-010, ISO 11138-1

What is the Double-Well Glass Bead Sterilizer?

A sterilizer with two independent heated wells so one bed stays at temperature while the other recovers after use. It suits multi-user benches and busy laboratories that cannot afford waiting between insertions.

What is the Double-Well Glass Bead Sterilizer?

A sterilizer with two independent heated wells so one bed stays at temperature while the other recovers after use. It suits multi-user benches and busy laboratories that cannot afford waiting between insertions.

Overview

The double-well sterilizer houses two wells in one case, each with its own heating element and thermostat, or one controller with two independently switched circuits. Because the wells are separate, an instrument inserted into one bed does not pull down the temperature of the other, so throughput roughly doubles for a small increase in footprint and power. Capacity is usually 100 to 200 g of beads per well, and the wells can hold different bead sizes or be dedicated to different instrument families, for example loops on one side and forceps on the other. A single digital controller commonly displays both temperatures. Units draw 200 to 500 W, so a dedicated circuit or a socket with adequate rating is advisable, and the case becomes warmer than a single-well model.

How does it work?

Glass bead sterilization works by thermal conduction. A well in the unit holds soda-lime or borosilicate beads of about 1.5 to 4 mm diameter, and a resistive heating element raises them to the set temperature, commonly 250 °C. The bead bed has a high thermal mass, so when a small instrument is pushed into it the beads transfer heat directly to the metal and surface microorganisms are killed by dry heat within 10 to 30 seconds. A thermocouple or platinum sensor in the well feeds a controller that switches the heater and holds the temperature, while a timer or indicator lamp shows when the unit is ready. Insulation around the well limits heat loss and keeps the outer case cool. Because heat reaches only the surfaces touching the beads, deep lumens and hinges may stay unsterilized.

What are the key features?

  • Two independently heated wells with separate controls.
  • Throughput roughly doubled with a small footprint increase.
  • Wells can hold different bead sizes or dedicated tools.
  • Single display shows both well temperatures.
  • Ideal for multi-user benches and shared laboratories.
  • Available with interlocked covers for each well.

What are the technical specifications?

A typical double-well glass bead sterilizer configuration offers number of wells of 2, working temperature of 250 °C nominal per well, warm-up time to 250 °C of 20-30 min, well diameter of 40-60 mm each, well depth of 40-80 mm each and bead capacity per well of 100-200 g. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Double-Well Glass Bead Sterilizer is used in multi-user microbiology benches, animal facility surgery preparation, dental clinics with two operators and teaching laboratories with rotating groups. Across the wider Glass Bead Sterilizers class, laboratories also deploy it for rapid sterilization of forceps and scissors, inoculating loop and needle sterilization and animal research instrument sterilization.

Which standards and certifications apply?

Relevant standards and methods include EN 61010-1, EN 61010-2-010, ISO 11138-1, GB 4793.1 and ISO 9001 manufacturing. Certification candidates for this model are CE, ISO 9001 manufacturing and EN 61010-1 test report; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Place the unit on a heat-resistant bench away from flammable materials and leave at least 10 cm clearance on all sides. Use a dedicated socket, since benchtop units typically draw 100 to 300 W. Verify the bed temperature with an external thermometer at installation and annually afterwards, because a drifting thermostat can leave the bed below the effective temperature. Keep the bed filled to the marked level, as low beads reduce contact and overheat the element. Remove debris and metal chips with a sieve, and replace beads when they discolor, crack or sinter. Wear heat-resistant gloves when inserting instruments and never place solvent bottles near the well.

How do I select the right Double-Well Glass Bead Sterilizer?

Select the double-well model when several people share one bench or when insertions follow each other closely, as the second well removes recovery waiting. If only one operator works intermittently, a single-well unit is sufficient and uses less power. Specify whether the wells need independent temperature settings and which bead sizes are required for each side.

How does it compare with other Glass Bead Sterilizers variants?

Choose a glass bead sterilizer when small metal tools must be re-used quickly between bench steps, not when packaged loads, liquids or lumens need sterilization. Compare the well diameter and depth with the length of your instruments, because long forceps must reach the bead bed without the handle becoming too hot. Check the maximum temperature, the warm-up time and the recovery time after each insertion. A digital temperature display and timer make routine use more consistent than a simple indicator lamp. For heavy use, two wells or two units avoid waiting during warm-up. Confirm the bead size, the availability of replacement bead packs and whether a basket or heat shield is included.

Full specifications

Full specifications
Number of wells2
Working temperature250 °C nominal per well
Warm-up time to 250 °C20-30 min
Well diameter40-60 mm each
Well depth40-80 mm each
Bead capacity per well100-200 g
Typical hold time10-30 s per instrument
Temperature controlDual PID or independent thermostats
Safety devicesPer-well over-temperature cut-out, thermal fuse
Power supply220 V, 50 Hz, 200-500 W

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Multi-user microbiology benches
  • Animal facility surgery preparation
  • Dental clinics with two operators
  • Teaching laboratories with rotating groups

Key features

  • Two independently heated wells with separate controls.
  • Throughput roughly doubled with a small footprint increase.
  • Wells can hold different bead sizes or dedicated tools.
  • Single display shows both well temperatures.
  • Ideal for multi-user benches and shared laboratories.
  • Available with interlocked covers for each well.

Frequently asked questions about Double-Well Glass Bead Sterilizer

Not necessarily. Models with independent controllers allow different set points, which is useful when one well holds fine beads for loops and the other coarse beads for heavier instruments.

Request a quotation for the Double-Well Glass Bead Sterilizer

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote