Open-Loop Flow-Through Dissolution System (USP Apparatus 4)
开环流通池溶出系统(USP 装置 4)
- CE (to be confirmed)
- +1
Welcome to Lab Manufacturers China — laboratory instruments direct from a Chinese manufacturer

带 HPLC 接口的自动多通道流通池溶出系统
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
| Apparatus type | USP Apparatus 4, automated |
|---|---|
| Number of channels | 6–8 |
| Flow rate | 1–50 mL/min programmable |
| Flow accuracy | ±3% |
| Pulse rate | 120 ± 10 pulses/min |
| Temperature control | 37.0 ± 0.5 °C |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <711>, USP <1092>, Ph. Eur. 2.9.3
This system runs six to eight flow-through cells in parallel under software control and delivers timed fractions to an HPLC autosampler vial tray or UV detector. It suits regulated laboratories needing unattended, data-integrity-compliant flow-through profiles. It supports compliant profile studies with many time points.
This system runs six to eight flow-through cells in parallel under software control and delivers timed fractions to an HPLC autosampler vial tray or UV detector. It suits regulated laboratories needing unattended, data-integrity-compliant flow-through profiles. It supports compliant profile studies with many time points.
The automated variant integrates independent pump modules, a thermostatted cell bath, valve manifolds and fraction collection. Software schedules flow rate, medium switching among reservoirs, and fraction timing, for example pH steps from 1.2 to 6.8. Eluate fractions are dispensed into vials for HPLC analysis or passed through a UV flow cell for continuous readout. Each channel pump is monitored for flow accuracy and pulse rate, with deviations logged. The integrated medium preparation option blends buffers online, which reduces manual preparation errors. Because all steps are scripted, runs are repeatable and traceable, and with audit trails, e-signatures and role-based access the system supports 21 CFR Part 11 needs. It is well suited to quality control of extended-release and poorly soluble products where profile data at many time points are required, and to regulatory submissions needing f2 comparisons.
A flow-through system uses a piston pump to deliver medium at a controlled flow rate, typically 4–16 mL/min, with a sinusoidal pulse of 120 ± 10 pulses per minute, upward through a vertical cell. The dosage form sits on a bed of glass beads, which stabilize laminar flow, and a filter at the top retains particles. The medium can go once through the cell (open loop), so fresh solvent continually maintains sink conditions, or return to a reservoir (closed loop). Because the cell is stationary, hydrodynamics are mild and reproducible, and the medium can be changed programmatically to create pH profiles. Key components are the pump, cells of 12 or 22.6 mm diameter, a heated bath holding 37 °C, filter heads, and a collector or detector. Release follows diffusion and erosion kinetics, and flow rate determines the boundary layer thickness around the dosage form.
A typical automated multi-channel flow-through dissolution system with hplc interface configuration offers apparatus type of USP Apparatus 4, automated, number of channels of 6–8, flow rate of 1–50 mL/min programmable, flow accuracy of ±3%, pulse rate of 120 ± 10 pulses/min and temperature control of 37.0 ± 0.5 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
The Automated Multi-Channel Flow-Through Dissolution System with HPLC Interface is used in regulated extended-release QC, regulatory f2 submission, pH-gradient profile studies and unattended overnight runs. Across the wider Flow-Through Systems class, laboratories also deploy it for extended-release product QC, poorly soluble drug testing and stent and implant release studies.
Relevant standards and methods include USP <711>, USP <1092>, Ph. Eur. 2.9.3, JP 6.10, ChP 0931, USP <724> and FDA Guidance on Dissolution Testing of Immediate Release Solid Oral Dosage Forms. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Degas media and prime the pump to remove air, since bubbles in cells cause flow variation and filter blockage. Verify flow rate gravimetrically and check pulse rate at installation and regularly. Use calibrated thermometers to confirm cell temperature at 37.0 ± 0.5 °C. Clean cells, beads and filters after each run and replace worn pump seals or check valves. Keep tubing lengths short to limit dead volume. Qualification should include IQ/OQ, flow calibration and dissolution performance verification using reference tablets. For regulated work, ensure software supports audit trail and data integrity.
Choose the automated multi-channel system when many time-points, pH-gradient profiles or overnight runs are routine, and compliance requires a full audit trail. Specify channel count, flow range, fraction vial capacity, HPLC or UV interface, and validation documents. Smaller labs with occasional flow-through work may find a manual four-channel system adequate.
Choose by the nature of the product and the regulatory requirement. Open-loop suits poorly soluble actives and extended-release forms needing sink conditions; closed-loop suits low-dose tablets and small media volumes. Implants, stents and microspheres need specialty cells and low flow rates. For high throughput or unattended runs, pick an automated multi-channel system with fraction collection and audit trail. A retrofit module is the lowest-cost route to meet an Apparatus 4 monograph. Evaluate flow accuracy (±5% or better), pulse rate control, cell diameter options, wetted-part material, number of channels and software data integrity features. Also consider medium volume and waste because open-loop runs can consume many liters.
| Apparatus type | USP Apparatus 4, automated |
|---|---|
| Number of channels | 6–8 |
| Flow rate | 1–50 mL/min programmable |
| Flow accuracy | ±3% |
| Pulse rate | 120 ± 10 pulses/min |
| Temperature control | 37.0 ± 0.5 °C |
| Fraction collection | up to 96 vials per channel |
| Detector interface | UV flow cell, HPLC autosampler |
| Media switching | up to 4 reservoirs per channel |
| Data integrity | audit trail, e-signature, user levels |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Yes. A valve manifold can switch between reservoirs on a programmed schedule, for example from pH 1.2 acid to pH 6.8 buffer, to simulate gastrointestinal transit without opening the cell.
Fractions are dispensed into vials that can be loaded into an HPLC autosampler, but HPLC analysis itself is run separately unless the system is integrated with a chromatography data system through software.
Flow is verified gravimetrically by collecting effluent over a timed interval and weighing it, with acceptance typically ±3 to 5% of setpoint. Pulse rate is checked with the instrument diagnostics.
It is a dissolution apparatus, USP Apparatus 4, in which medium is pumped through a vertical cell holding the dosage form. The medium either passes once (open loop) or recirculates (closed loop), providing gentle, reproducible hydrodynamics.
It is preferred for poorly soluble drugs needing sink conditions, for implants and stents, for extended-release forms, and for testing where medium changes are needed without handling the sample.
Compendial methods typically use 4, 8 or 16 mL/min with a pulsation of 120 ± 10 pulses per minute. Specialty methods for implants may use lower flow rates of 0.1 to 4 mL/min.

开环流通池溶出系统(USP 装置 4)