Skip to main content
Lab Manufacturers China
Closed-Loop Flow-Through Dissolution System — product photo

Closed-Loop Flow-Through Dissolution System

闭环流通池溶出系统

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Apparatus typeUSP Apparatus 4, closed loop
Number of cells6–8
Reservoir volume100–1000 mL
Flow rate8–16 mL/min
Pulse rate120 ± 10 pulses/min
Temperature control37.0 ± 0.5 °C
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <711>, USP <1092>, Ph. Eur. 2.9.3

What is the Closed-Loop Flow-Through Dissolution System?

A closed-loop flow-through dissolution system recirculates a fixed volume of medium through the cell and back to a reservoir, so concentration builds up over time. It suits low-dose products and small-volume tests where medium use and waste must be limited.

What is the Closed-Loop Flow-Through Dissolution System?

A closed-loop flow-through dissolution system recirculates a fixed volume of medium through the cell and back to a reservoir, so concentration builds up over time. It suits low-dose products and small-volume tests where medium use and waste must be limited.

Overview

In closed-loop mode, medium circulates from a reservoir through the flow cell and back, typically at 8 to 16 mL/min with a total volume of 100 to 1000 mL. Since dissolved drug accumulates in the reservoir, the concentration can be sampled manually from the reservoir or measured continuously by a UV flow cell, making it suitable for low-dose tablets whose released concentration is detectable in the small volume. The same pump and cells used for open-loop tests are configured with a switching valve or dedicated tubing. Closed-loop operation behaves like a paddle test in terms of concentration but retains the gentle, reproducible hydrodynamics of the cell. It is frequently used for low-dose products, for testing with small medium volumes such as expensive buffers or biorelevant media, and for routine tests where cost of medium disposal matters. Sink conditions must be confirmed because saturation may develop.

How does it work?

A flow-through system uses a piston pump to deliver medium at a controlled flow rate, typically 4–16 mL/min, with a sinusoidal pulse of 120 ± 10 pulses per minute, upward through a vertical cell. The dosage form sits on a bed of glass beads, which stabilize laminar flow, and a filter at the top retains particles. The medium can go once through the cell (open loop), so fresh solvent continually maintains sink conditions, or return to a reservoir (closed loop). Because the cell is stationary, hydrodynamics are mild and reproducible, and the medium can be changed programmatically to create pH profiles. Key components are the pump, cells of 12 or 22.6 mm diameter, a heated bath holding 37 °C, filter heads, and a collector or detector. Release follows diffusion and erosion kinetics, and flow rate determines the boundary layer thickness around the dosage form.

What are the key features?

  • Recirculating medium loop with 100–1000 mL reservoir
  • Switching valve converts between open and closed loop
  • Inline UV flow-cell monitoring of reservoir concentration
  • Low medium consumption suited to biorelevant media
  • Pulsation-controlled pump for reproducible hydrodynamics
  • Temperature-controlled reservoir and tubing

What are the technical specifications?

A typical closed-loop flow-through dissolution system configuration offers apparatus type of USP Apparatus 4, closed loop, number of cells of 6–8, reservoir volume of 100–1000 mL, flow rate of 8–16 mL/min, pulse rate of 120 ± 10 pulses/min and temperature control of 37.0 ± 0.5 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Closed-Loop Flow-Through Dissolution System is used in low-dose tablet dissolution, biorelevant media testing, routine QC with reduced waste and method development. Across the wider Flow-Through Systems class, laboratories also deploy it for extended-release product QC, poorly soluble drug testing and stent and implant release studies.

Which standards and certifications apply?

Relevant standards and methods include USP <711>, USP <1092>, Ph. Eur. 2.9.3, JP 6.10, ChP 0931, USP <724> and FDA Guidance on Dissolution Testing of Immediate Release Solid Oral Dosage Forms. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Degas media and prime the pump to remove air, since bubbles in cells cause flow variation and filter blockage. Verify flow rate gravimetrically and check pulse rate at installation and regularly. Use calibrated thermometers to confirm cell temperature at 37.0 ± 0.5 °C. Clean cells, beads and filters after each run and replace worn pump seals or check valves. Keep tubing lengths short to limit dead volume. Qualification should include IQ/OQ, flow calibration and dissolution performance verification using reference tablets. For regulated work, ensure software supports audit trail and data integrity.

How do I select the right Closed-Loop Flow-Through Dissolution System?

Choose closed-loop when medium volume is limited, doses are low, or waste disposal costs matter. Specify reservoir volume range, pump flow accuracy, UV interface and valve type. For poorly soluble actives that need constant sink, select the open-loop version or a dual-mode system. Confirm the expected test duration and reservoir capacity.

How does it compare with other Flow-Through Systems variants?

Choose by the nature of the product and the regulatory requirement. Open-loop suits poorly soluble actives and extended-release forms needing sink conditions; closed-loop suits low-dose tablets and small media volumes. Implants, stents and microspheres need specialty cells and low flow rates. For high throughput or unattended runs, pick an automated multi-channel system with fraction collection and audit trail. A retrofit module is the lowest-cost route to meet an Apparatus 4 monograph. Evaluate flow accuracy (±5% or better), pulse rate control, cell diameter options, wetted-part material, number of channels and software data integrity features. Also consider medium volume and waste because open-loop runs can consume many liters.

Full specifications

Full specifications
Apparatus typeUSP Apparatus 4, closed loop
Number of cells6–8
Reservoir volume100–1000 mL
Flow rate8–16 mL/min
Pulse rate120 ± 10 pulses/min
Temperature control37.0 ± 0.5 °C
Cell diameter12 mm and 22.6 mm
Detectionmanual sampling or UV flow cell
Mixingreservoir stirrer, optional
Power supply220 V AC, 50/60 Hz, about 0.8 kW

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Low-dose tablet dissolution
  • Biorelevant media testing
  • Routine QC with reduced waste
  • Method development

Key features

  • Recirculating medium loop with 100–1000 mL reservoir
  • Switching valve converts between open and closed loop
  • Inline UV flow-cell monitoring of reservoir concentration
  • Low medium consumption suited to biorelevant media
  • Pulsation-controlled pump for reproducible hydrodynamics
  • Temperature-controlled reservoir and tubing

Frequently asked questions about Closed-Loop Flow-Through Dissolution System

Not directly. Both accumulate drug in a fixed medium volume, but the flow cell provides gentler, more reproducible hydrodynamics. Method equivalence must be demonstrated during development before replacing a paddle method.

Request a quotation for the Closed-Loop Flow-Through Dissolution System

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote