Open-Loop Flow-Through Dissolution System (USP Apparatus 4)
开环流通池溶出系统(USP 装置 4)
- SKU FTS-0001
- Flow-Through Systems
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Apparatus type | USP Apparatus 4, open loop |
|---|---|
| Number of cells | 6–8 |
| Cell diameter | 12 mm and 22.6 mm |
| Flow rate | 4–16 mL/min |
| Pulse rate | 120 ± 10 pulses/min |
| Flow accuracy | ±5% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <711>, USP <1092>, Ph. Eur. 2.9.3
What is the Open-Loop Flow-Through Dissolution System (USP Apparatus 4)?
An open-loop flow-through dissolution system pumps fresh medium once through a thermostatted cell holding the dosage form and sends the eluate to a collector or detector. It is the preferred configuration for poorly soluble drugs, implants and products needing sink conditions throughout the test.
What is the Open-Loop Flow-Through Dissolution System (USP Apparatus 4)?
An open-loop flow-through dissolution system pumps fresh medium once through a thermostatted cell holding the dosage form and sends the eluate to a collector or detector. It is the preferred configuration for poorly soluble drugs, implants and products needing sink conditions throughout the test.
Overview
In open-loop mode, a piston or pulsed pump draws fresh medium from a reservoir, delivers it upward through the cell at a set flow rate, and discards or collects the effluent after filtration. Because the medium never recirculates, concentration in the cell stays low and sink conditions are maintained for drugs of low solubility. Flow rates of 4, 8 or 16 mL/min are typical, with a sinusoidal pulse of 120 ± 10 pulses per minute as USP <711> specifies. Cells come in 12 mm and 22.6 mm diameters, with ruby beads and glass beads at the bottom to create laminar flow. The cell and medium are kept at 37 °C by a heated bath. Open-loop systems are well suited to extended-release products, suspensions, suppositories and stents, and eluate fractions can be fed to UV or HPLC detectors. They consume more medium and require larger collection volumes than closed-loop systems.
How does it work?
A flow-through system uses a piston pump to deliver medium at a controlled flow rate, typically 4–16 mL/min, with a sinusoidal pulse of 120 ± 10 pulses per minute, upward through a vertical cell. The dosage form sits on a bed of glass beads, which stabilize laminar flow, and a filter at the top retains particles. The medium can go once through the cell (open loop), so fresh solvent continually maintains sink conditions, or return to a reservoir (closed loop). Because the cell is stationary, hydrodynamics are mild and reproducible, and the medium can be changed programmatically to create pH profiles. Key components are the pump, cells of 12 or 22.6 mm diameter, a heated bath holding 37 °C, filter heads, and a collector or detector. Release follows diffusion and erosion kinetics, and flow rate determines the boundary layer thickness around the dosage form.
What are the key features?
- Fresh medium delivered once through each cell
- Sinusoidal pulse pump at 120 ± 10 pulses per minute
- Flow rates 4–16 mL/min with flow accuracy ±5%
- Cells in 12 mm and 22.6 mm diameters with bead beds
- Inline filters at cell outlet for clear eluate
- Fraction collection or direct UV/HPLC detector interface
What are the technical specifications?
A typical open-loop flow-through dissolution system (usp apparatus 4) configuration offers apparatus type of USP Apparatus 4, open loop, number of cells of 6–8, cell diameter of 12 mm and 22.6 mm, flow rate of 4–16 mL/min, pulse rate of 120 ± 10 pulses/min and flow accuracy of ±5%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Open-Loop Flow-Through Dissolution System (USP Apparatus 4) is used in poorly soluble drug dissolution, extended-release dosage forms, implants and stents and suppositories. Across the wider Flow-Through Systems class, laboratories also deploy it for extended-release product QC, poorly soluble drug testing and stent and implant release studies.
Which standards and certifications apply?
Relevant standards and methods include USP <711>, USP <1092>, Ph. Eur. 2.9.3, JP 6.10, ChP 0931, USP <724> and FDA Guidance on Dissolution Testing of Immediate Release Solid Oral Dosage Forms. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Degas media and prime the pump to remove air, since bubbles in cells cause flow variation and filter blockage. Verify flow rate gravimetrically and check pulse rate at installation and regularly. Use calibrated thermometers to confirm cell temperature at 37.0 ± 0.5 °C. Clean cells, beads and filters after each run and replace worn pump seals or check valves. Keep tubing lengths short to limit dead volume. Qualification should include IQ/OQ, flow calibration and dissolution performance verification using reference tablets. For regulated work, ensure software supports audit trail and data integrity.
How do I select the right Open-Loop Flow-Through Dissolution System (USP Apparatus 4)?
Choose the open-loop configuration when sink conditions are critical or total release profiles of low-solubility actives are needed. Specify cell diameter, flow range, number of channels, filter type and reservoir size. If medium volume or disposal is limited, consider a closed-loop system instead. Confirm the expected test duration and reservoir capacity.
How does it compare with other Flow-Through Systems variants?
Choose by the nature of the product and the regulatory requirement. Open-loop suits poorly soluble actives and extended-release forms needing sink conditions; closed-loop suits low-dose tablets and small media volumes. Implants, stents and microspheres need specialty cells and low flow rates. For high throughput or unattended runs, pick an automated multi-channel system with fraction collection and audit trail. A retrofit module is the lowest-cost route to meet an Apparatus 4 monograph. Evaluate flow accuracy (±5% or better), pulse rate control, cell diameter options, wetted-part material, number of channels and software data integrity features. Also consider medium volume and waste because open-loop runs can consume many liters.
Full specifications
| Apparatus type | USP Apparatus 4, open loop |
|---|---|
| Number of cells | 6–8 |
| Cell diameter | 12 mm and 22.6 mm |
| Flow rate | 4–16 mL/min |
| Pulse rate | 120 ± 10 pulses/min |
| Flow accuracy | ±5% |
| Temperature control | 37.0 ± 0.5 °C |
| Medium reservoir | 5–20 L per channel |
| Tablet cell type | standard and large-volume |
| Power supply | 220 V AC, 50/60 Hz, about 0.9 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Poorly soluble drug dissolution
- Extended-release dosage forms
- Implants and stents
- Suppositories
Key features
- Fresh medium delivered once through each cell
- Sinusoidal pulse pump at 120 ± 10 pulses per minute
- Flow rates 4–16 mL/min with flow accuracy ±5%
- Cells in 12 mm and 22.6 mm diameters with bead beds
- Inline filters at cell outlet for clear eluate
- Fraction collection or direct UV/HPLC detector interface
Frequently asked questions about Open-Loop Flow-Through Dissolution System (USP Apparatus 4)
Because fresh medium continuously replaces the eluate, the concentration around the dosage form stays well below saturation. Sink conditions are therefore maintained without adding surfactants or large medium volumes, which gives a more discriminating release profile.
Standard methods use 4, 8 or 16 mL/min, selected to match the monograph and to keep the dosage form exposed to sufficient hydrodynamic force. The pulse rate should be 120 ± 10 pulses per minute.
At 8 mL/min, a 24 h run consumes about 11.5 L per cell, so reservoir capacity and waste handling must be planned. Shorter immediate-release runs need far less.
It is a dissolution apparatus, USP Apparatus 4, in which medium is pumped through a vertical cell holding the dosage form. The medium either passes once (open loop) or recirculates (closed loop), providing gentle, reproducible hydrodynamics.
It is preferred for poorly soluble drugs needing sink conditions, for implants and stents, for extended-release forms, and for testing where medium changes are needed without handling the sample.
Compendial methods typically use 4, 8 or 16 mL/min with a pulsation of 120 ± 10 pulses per minute. Specialty methods for implants may use lower flow rates of 0.1 to 4 mL/min.