Benchtop Vacuum Dissolution Media Degasser
台式真空溶出介质脱气机
- SKU MEP-0002
- Media Preparation
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Batch volume | 1–4 L |
|---|---|
| Heating temperature | ambient to 45 °C |
| Ultimate vacuum | about 100 mbar |
| Filter pore size | 0.45 µm |
| Filter diameter | 47 mm or 90 mm |
| Stirring speed | 100–1200 rpm |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <711>, USP <791>, USP <1092>
What is the Benchtop Vacuum Dissolution Media Degasser?
A benchtop vacuum degasser removes dissolved air from dissolution medium by heating it to about 41 °C and filtering it under vacuum, following the USP <711> approach. It is the simplest way to degas manually prepared media for routine tests.
What is the Benchtop Vacuum Dissolution Media Degasser?
A benchtop vacuum degasser removes dissolved air from dissolution medium by heating it to about 41 °C and filtering it under vacuum, following the USP <711> approach. It is the simplest way to degas manually prepared media for routine tests.
Overview
The benchtop unit consists of a heated stirring flask or reservoir, a vacuum pump capable of reaching about 100 mbar, a membrane filter holder (0.45 µm) and a receiving vessel. Medium is warmed, filtered through the membrane under vacuum, and then stirred vigorously under vacuum for about five minutes. The combination reduces dissolved gases so that bubbles do not form on tablets during testing. Some units include an integrated dissolved-oxygen probe that displays residual oxygen, often with a target below 6 mg/L, and a timer with automatic vacuum shutoff. Throughput is about 1–4 liters per batch, sufficient for one run of six 900 mL vessels when done in two batches. Because it uses single-use membranes or reusable filter holders, consumable cost is low. This approach is simple but depends on the operator, which makes procedures and timers essential for repeatable results.
How does it work?
Dissolution media such as 0.1 N HCl, pH 4.5 acetate buffer and pH 6.8 phosphate buffer must have correct composition, pH, volume, temperature and dissolved gas content. Preparation equipment automates these steps. Proportioning pumps or valves mix concentrate and purified water; pH electrodes and conductivity cells verify composition; degassing removes dissolved gases by vacuum with heat, vacuum membranes, or helium sparging, since gas bubbles on tablets or baskets change local hydrodynamics and surface area; and volumetric dispensers fill vessels to ±1% at 37–41 °C. Gas solubility falls with temperature and pressure, which is the basis for heated vacuum degassing, while helium displaces air because of its very low solubility. Materials contacting medium are PTFE, silicone or stainless steel to prevent leachables.
What are the key features?
- Heated stirring reservoir to about 41 °C
- Vacuum pump reaching about 100 mbar
- 0.45 µm membrane filter holder for particle removal
- Timer with automatic vacuum shutoff
- Optional dissolved-oxygen probe display
- Compact benchtop size with low consumable cost
What are the technical specifications?
A typical benchtop vacuum dissolution media degasser configuration offers batch volume of 1–4 L, heating temperature of ambient to 45 °C, ultimate vacuum of about 100 mbar, filter pore size of 0.45 µm, filter diameter of 47 mm or 90 mm and stirring speed of 100–1200 rpm. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Benchtop Vacuum Dissolution Media Degasser is used in routine tablet dissolution, small QC laboratories, acidic and buffer media and teaching laboratories. Across the wider Media Preparation class, laboratories also deploy it for dissolution medium preparation, buffer standardization and degassing for UV and HPLC methods.
Which standards and certifications apply?
Relevant standards and methods include USP <711>, USP <791>, USP <1092>, Ph. Eur. 2.9.3, Ph. Eur. 2.2.3, JP 6.10, ChP 0931 and ASTM E2503. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Calibrate pumps gravimetrically and pH sensors with at least two buffers on a regular schedule. Replace membrane filters, tubing and seals at defined intervals, and flush lines after each medium to prevent precipitation and microbial growth, especially with phosphate buffers. Use purified water meeting the method specification. Helium systems need leak-tight fittings and cylinder safety. Record concentrate lot numbers and expiry, and verify final pH at 37 °C where the monograph requires it. Include the equipment in IQ/OQ and preventive maintenance programs.
How do I select the right Benchtop Vacuum Dissolution Media Degasser?
Select the benchtop degasser for small to medium laboratories preparing less than about 20 liters per day and accepting manual handling. Specify batch volume, vacuum level, filter size, timer and whether an oxygen probe is required. For high volume or multi-operator sites, inline automated systems offer better repeatability and records.
How does it compare with other Media Preparation variants?
Match the equipment to daily medium volume, number of buffer recipes and compliance needs. A benchtop vacuum degasser suits laboratories preparing under 20 liters per day by hand. A helium station suits heat-labile or surfactant media. A volumetric dispenser removes filling error for many vessels. A dilution station with pH control reduces salt weighing. A fully automated system combines all steps and records batch data for GMP sites. Evaluate volume accuracy (±1%), dissolved oxygen endpoint, compatible recipes, cleaning and carry-over control, wetted materials, and software audit-trail features. Method validation must show that the degassing approach gives dissolution results equivalent to the compendial procedure.
Full specifications
| Batch volume | 1–4 L |
|---|---|
| Heating temperature | ambient to 45 °C |
| Ultimate vacuum | about 100 mbar |
| Filter pore size | 0.45 µm |
| Filter diameter | 47 mm or 90 mm |
| Stirring speed | 100–1200 rpm |
| Degassing time | about 5 min |
| Dissolved oxygen | < 6 mg/L typical |
| Pump type | diaphragm, oil-free |
| Power supply | 220 V AC, 50/60 Hz, about 0.4 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Routine tablet dissolution
- Small QC laboratories
- Acidic and buffer media
- Teaching laboratories
Key features
- Heated stirring reservoir to about 41 °C
- Vacuum pump reaching about 100 mbar
- 0.45 µm membrane filter holder for particle removal
- Timer with automatic vacuum shutoff
- Optional dissolved-oxygen probe display
- Compact benchtop size with low consumable cost
Frequently asked questions about Benchtop Vacuum Dissolution Media Degasser
Filtering under vacuum followed by about five minutes of vacuum stirring is typical. Total time per 1–2 L batch is around 10 minutes, depending on filter flow rate and medium viscosity.
Gas solubility decreases as temperature rises, so warming to about 41 °C helps release dissolved air, and the medium then cools to 37 °C in the vessel without reabsorbing significant gas.
Use a dissolved-oxygen probe to confirm levels below about 6 mg/L, or demonstrate by method validation that results from degassed and reference media are equivalent. Visual bubble absence on the tablet is a practical check.
Dissolved gases form bubbles on tablets, baskets and vessel walls during the test. Bubbles reduce surface contact and alter hydrodynamics, which lowers measured dissolution. USP <711> therefore requires degassing unless validated otherwise.
USP <711> requires that medium volume be within ±1% of the specified volume, such as 900 ± 9 mL. Volumetric dispensers or calibrated glassware are needed to meet this consistently.
Medium is typically preheated to 37–41 °C so vessels equilibrate to 37.0 ± 0.5 °C rapidly. The temperature is verified in each vessel before dosage forms are introduced.