Benchtop pH and Ion Meter
台式 pH 和离子计
- SKU ELC-0001
- Electrochemical Analyzers
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| pH range | −2.000 to 20.000 pH |
|---|---|
| pH resolution | 0.001 / 0.01 / 0.1 selectable |
| pH accuracy (meter) | ±0.002 pH |
| mV range | ±2000.0 mV |
| mV accuracy | ±0.2 mV |
| Ion concentration range | 1×10^-9 to 1×10^9 (units selectable) |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D1293, ASTM D1125, ASTM D888
What is the Benchtop pH and Ion Meter?
A benchtop pH and ion meter measures potential from glass or ion-selective electrodes and converts it to pH, mV or ion concentration with temperature compensation. It is the standard laboratory tool for buffer preparation, water analysis, fluoride or nitrate determination, and routine quality control.
What is the Benchtop pH and Ion Meter?
A benchtop pH and ion meter measures potential from glass or ion-selective electrodes and converts it to pH, mV or ion concentration with temperature compensation. It is the standard laboratory tool for buffer preparation, water analysis, fluoride or nitrate determination, and routine quality control.
Overview
The meter provides a high-impedance input above 10^12 ohms for glass membranes, a precise millivolt converter and a temperature input for an NTC or Pt1000 probe. Calibration supports up to five points with automatic buffer recognition for common series such as NIST, DIN and technical buffers, and the calibration slope and offset are logged. Ion modes use direct calibration, known addition or standard addition to measure ions including fluoride, sodium, ammonium and chloride. Dual-channel models read two electrodes simultaneously. Good laboratory practice features include user levels, calibration reminders, electrode identity and data export. Resolution is commonly 0.001 pH and 0.1 mV, but accuracy in practice is governed by electrode condition, buffer quality and sample temperature stability.
How does it work?
Electrochemical analysis links an electrical signal to analyte activity or concentration. Potentiometry measures the potential between an indicator electrode, such as a pH glass membrane or ion-selective membrane, and a reference electrode, related to ion activity by the Nernst equation, about 59.16 mV per decade at 25 °C for monovalent ions. Conductometry measures the ability of ions in solution to carry current between electrodes at alternating voltage, with temperature compensation. Amperometric sensors such as Clark cells measure the current from reduction of dissolved oxygen, whereas optical sensors measure luminescence quenching. Voltammetry sweeps potential across a working electrode and records current, with peak position identifying the species and peak height its concentration. Potentiostats control the potential of the working electrode against a reference while measuring current through a counter electrode.
What are the key features?
- Input impedance above 10^12 ohms suitable for glass and ion-selective electrodes
- Automatic buffer recognition with up to five-point calibration and slope logging
- Direct, known-addition and standard-addition modes for ion concentration
- Single or dual channel input with ATC probe for temperature compensation
- User levels, calibration reminders and electrode identification for GLP compliance
- Graphical display with USB and printer output for record keeping
What are the technical specifications?
A typical benchtop ph and ion meter configuration offers pH range of −2.000 to 20.000 pH, pH resolution of 0.001 / 0.01 / 0.1 selectable, pH accuracy (meter) of ±0.002 pH, mV range of ±2000.0 mV, mV accuracy of ±0.2 mV and ion concentration range of 1×10^-9 to 1×10^9 (units selectable). Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Benchtop pH and Ion Meter is used in buffer and reagent preparation, drinking water and wastewater pH, fluoride in toothpaste and water, nitrate and ammonium in fertilizer extracts and incoming inspection of chemicals. Across the wider Electrochemical Analyzers class, laboratories also deploy it for pH and ion control in water and process chemistry, conductivity of ultrapure and boiler water and dissolved oxygen in wastewater and fermentation.
Which standards and certifications apply?
Relevant standards and methods include ASTM D1293, ASTM D1125, ASTM D888, ISO 10523, ISO 7888, USP <645>, ASTM G5 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Electrode condition dominates accuracy. Calibrate pH meters with two or three fresh NIST-traceable buffers bracketing the sample, record slope, which should be 95–105% of theoretical, and replace electrodes with slow response. Store glass electrodes in proper storage solution, never distilled water. Calibrate conductivity cells with certified standards such as 1413 µS/cm at 25 °C and verify temperature compensation. Keep optical dissolved oxygen caps within their service life. Reference electrode fill solutions need topping up and periodic replacement. For potentiostats, use shielded cables, a Faraday cage for nA-level work and correct connection to the working, reference and counter electrodes. Maintain calibration records suited to ISO/IEC 17025.
How do I select the right Benchtop pH and Ion Meter?
Select this variant for routine pH, mV and ion measurement on the bench, where high resolution, GLP data handling and multiple electrodes are needed. Use a portable meter for field work and a titrator for higher-accuracy assays. In an RFQ, specify channels, resolution, ion-selective electrodes required, electrode types, GLP features, and standards such as ASTM D1293 or ISO 10523.
How does it compare with other Electrochemical Analyzers variants?
Choose by the measured parameter and accuracy demanded. Routine pH, mV and ion concentration call for a benchtop pH or ion meter with suitable electrodes, and resolution of 0.001 pH is worthwhile only when electrodes and buffers support it. Ultrapure water, boiler water and process streams need a conductivity meter with the right cell constant, with 0.01 for ultrapure water and 1.0 for general use. Dissolved oxygen in water and fermentation suit optical sensors, which avoid electrolyte maintenance. Trace heavy metals at ppb level suit stripping voltammetry. Corrosion, battery and materials research require a potentiostat with adequate current range and impedance capability. Also consider data integrity features, electrode availability and calibration automation.
Full specifications
| pH range | −2.000 to 20.000 pH |
|---|---|
| pH resolution | 0.001 / 0.01 / 0.1 selectable |
| pH accuracy (meter) | ±0.002 pH |
| mV range | ±2000.0 mV |
| mV accuracy | ±0.2 mV |
| Ion concentration range | 1×10^-9 to 1×10^9 (units selectable) |
| Temperature range | −5 to 110 °C (ATC) |
| Calibration points | Up to 5 (pH), up to 8 (ion) |
| Input impedance | > 10^12 ohm |
| Data interface | USB, RS-232 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Buffer and reagent preparation
- Drinking water and wastewater pH
- Fluoride in toothpaste and water
- Nitrate and ammonium in fertilizer extracts
- Incoming inspection of chemicals
Key features
- Input impedance above 10^12 ohms suitable for glass and ion-selective electrodes
- Automatic buffer recognition with up to five-point calibration and slope logging
- Direct, known-addition and standard-addition modes for ion concentration
- Single or dual channel input with ATC probe for temperature compensation
- User levels, calibration reminders and electrode identification for GLP compliance
- Graphical display with USB and printer output for record keeping
Frequently asked questions about Benchtop pH and Ion Meter
Two buffers bracketing the expected sample pH, such as 4.01 and 7.00 or 7.00 and 10.01, are the minimum. Three-point calibration is better for wide ranges, and buffers should be fresh, traceable and at the same temperature as samples.
It is a concentrated inert electrolyte added equally to standards and samples so ionic strength, and therefore activity coefficients, are the same. This allows the ion-selective electrode to respond to concentration instead of activity in direct measurements.
Common causes are a dirty or aged glass membrane, clogged reference junction, depleted reference electrolyte, poor temperature equilibrium and low-ionic-strength samples. Cleaning, refilling electrolyte and recalibrating usually restores stable readings.
A pH meter reads hydrogen ion activity from a glass electrode. An ion meter also reads other ions, such as fluoride, nitrate or sodium, with ion-selective electrodes and reports concentration using calibration standards and ionic strength adjusters. Most benchtop units combine both functions.
Ion mobility rises with temperature, increasing conductivity by roughly 2% per °C for most aqueous solutions. Meters use automatic temperature compensation to report conductivity at a reference of 25 °C, using a linear coefficient or nonlinear function for natural and ultrapure water.
With good care, pH glass electrodes last 6–24 months. Replace when calibration slope falls below about 92–95%, response is slow, or readings drift. Extreme pH, high temperature, fluoride and protein-rich samples shorten lifetime.