Skip to main content
Lab Manufacturers China
Optical Dissolved Oxygen Meter — product photo

Optical Dissolved Oxygen Meter

光学溶解氧仪

  • SKU DOM-0003
  • Dissolved Oxygen Meters
  • CE (to be confirmed)
  • RoHS (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Measuring range0–20 mg/L (0–200% saturation)
Accuracy±0.1 mg/L or ±1% of reading
Resolution0.01 mg/L
Response time≤60 s to 90%
Stirring dependenceNone (no oxygen consumption)
Temperature range0–50 °C
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 5814, ASTM D888, APHA 4500-O Standard Methods

What is the Optical Dissolved Oxygen Meter?

An optical dissolved oxygen meter uses a luminescent sensing cap instead of an electrochemical cell, so it needs no electrolyte, flow or polishing and holds calibration for months. It is preferred for unattended monitoring, low-maintenance field work and wastewater aeration control.

What is the Optical Dissolved Oxygen Meter?

An optical dissolved oxygen meter uses a luminescent sensing cap instead of an electrochemical cell, so it needs no electrolyte, flow or polishing and holds calibration for months. It is preferred for unattended monitoring, low-maintenance field work and wastewater aeration control.

Overview

The sensing cap contains a luminescent dye immobilized in a polymer film. Blue excitation light pulses the dye and the red emission is delayed by an amount that depends on the oxygen concentration; the instrument measures the phase shift and converts it to mg/L or percent saturation. Because there is no oxygen consumption and no membrane diffusion current, the measurement is independent of stirring, which removes a common source of error in stagnant or slow-moving waters. No electrolyte replacement, no membrane tensioning and no flow requirement mean maintenance is limited to cap replacement every one to two years and periodic recalibration. Response time is usually under 60 seconds, accuracy is often plus or minus 0.1 mg/L or 1 percent of reading, and fouling affects the cap less severely than electrochemical membranes. Sensors are available as benchtop probes, portable field probes, process probes with digital output and versions designed for high-temperature or CIP installation.

How does it work?

Electrochemical sensors are either galvanic or polarographic. Oxygen diffuses through a gas-permeable membrane into an electrolyte, where it is reduced at a cathode; the resulting current is proportional to the oxygen partial pressure. Polarographic cells need an applied voltage of about 0.7 V and a warm-up period, while galvanic cells generate their own current. Optical sensors use a luminescent dye immobilized in a film: blue light excites the dye, the emitted red light is quenched by oxygen, and the phase shift of the signal gives the oxygen concentration. Optical sensors need no electrolyte, flow, or polishing and hold calibration for months. Both types measure partial pressure, so salinity and barometric pressure corrections are applied to report concentration in mg/L, and a temperature sensor compensates the membrane permeability or dye response. Stirring or flow across the membrane is required for electrochemical sensors to avoid oxygen depletion at the surface.

What are the key features?

  • Luminescent sensing cap requiring no electrolyte or membrane service
  • Stirring-independent measurement that suits still water and slow flows
  • Calibration stable for months with automatic cap health diagnostics
  • Response time under 60 seconds with 0.1 mg/L typical accuracy
  • Digital or analog output options for process and laboratory use
  • Sensing cap life of one to two years with simple replacement

What are the technical specifications?

A typical optical dissolved oxygen meter configuration offers measuring range of 0–20 mg/L (0–200% saturation), accuracy of ±0.1 mg/L or ±1% of reading, resolution of 0.01 mg/L, response time of ≤60 s to 90%, stirring dependence of None (no oxygen consumption) and temperature range of 0–50 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Optical Dissolved Oxygen Meter is used in wastewater aeration control, long-term unattended monitoring, recirculating aquaculture systems, surface water and groundwater surveys and fermentation and bioprocess monitoring. Across the wider Dissolved Oxygen Meters class, laboratories also deploy it for municipal wastewater aeration control, fermentation and bioprocess monitoring and aquaculture pond and tank control.

Which standards and certifications apply?

Relevant standards and methods include ISO 5814, ASTM D888, APHA 4500-O Standard Methods, EN 25814, GB/T 11913 and USP <1231>. Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Calibrate electrochemical sensors in water-saturated air or against a reference method such as Winkler titration, and replace the membrane, electrolyte and O-ring at the interval the manufacturer specifies, commonly every one to three months. Keep the membrane free of deposits: biological growth in wastewater reduces response and must be cleaned without scratching the surface. Optical sensors need only periodic recalibration, but the sensing cap has a finite life of roughly one to two years and must be replaced. Zero-point checks with sodium sulfite solution confirm that the instrument reads low oxygen correctly. Record temperature, salinity and barometric pressure corrections with the data, and verify performance with a check standard or comparison measurement. For compliance work, follow the method required by the permit or pharmacopoeia and retain calibration and maintenance records.

How do I select the right Optical Dissolved Oxygen Meter?

Select an optical meter when maintenance access is limited, samples are slow-moving, or measurement stability over months matters more than initial cost. Electrochemical probes remain cheaper for occasional measurements. In an RFQ, specify cap life, calibration interval, output type, temperature range and whether CIP or high-temperature versions are required.

How does it compare with other Dissolved Oxygen Meters variants?

Choose between electrochemical and optical sensing first. Optical sensors suit long-term unattended monitoring and low-maintenance field work, while electrochemical sensors have lower initial cost and are well established in wastewater aeration control. Then match the format: benchtop meters for laboratory BOD, fermentation and validation work, portable meters with long cables for rivers, lakes and tanks, online analyzers with 4-20 mA outputs for aeration control, and BOD systems that measure depletion over five days. Check the measuring range, accuracy, temperature and salinity compensation, response time, minimum flow or stirring requirement, calibration method and logging capacity. Specify whether readings must be in mg/L, percent saturation or both, and which standards the laboratory follows.

Full specifications

Full specifications
Measuring range0–20 mg/L (0–200% saturation)
Accuracy±0.1 mg/L or ±1% of reading
Resolution0.01 mg/L
Response time≤60 s to 90%
Stirring dependenceNone (no oxygen consumption)
Temperature range0–50 °C
Temperature compensationAutomatic with integrated sensor
Calibration interval1–6 months typical
Sensing cap life12–24 months
OutputModbus RTU, 4-20 mA or meter display

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Wastewater aeration control
  • Long-term unattended monitoring
  • Recirculating aquaculture systems
  • Surface water and groundwater surveys
  • Fermentation and bioprocess monitoring

Key features

  • Luminescent sensing cap requiring no electrolyte or membrane service
  • Stirring-independent measurement that suits still water and slow flows
  • Calibration stable for months with automatic cap health diagnostics
  • Response time under 60 seconds with 0.1 mg/L typical accuracy
  • Digital or analog output options for process and laboratory use
  • Sensing cap life of one to two years with simple replacement

Frequently asked questions about Optical Dissolved Oxygen Meter

It does not consume oxygen. The dye responds to oxygen partial pressure by luminescence quenching, so there is no depleted layer at the surface. Stagnant water therefore gives the same reading as flowing water, unlike electrochemical cells that consume oxygen at the cathode.

Bipotentiostat — product photo

Bipotentiostat

双恒电位仪

  • CE (to be confirmed)
  • +2
Request a quote

Request a quotation for the Optical Dissolved Oxygen Meter

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote