Triple-Gas Hypoxia CO2 Incubator
三气低氧 CO2 培养箱
- SKU CON-0004
- CO2 Incubators
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Oxygen range | 0.1–21% |
|---|---|
| Oxygen control accuracy | within ±0.2% at low set points |
| CO2 range | 0–20%, controlled at 5% |
| Temperature range | ambient +5 to 50 °C, set at 37 °C |
| Oxygen sensor | Zirconia or electrochemical, calibration every 1–3 months |
| Gas supply | CO2 and nitrogen, or premixed triple gas |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-010, IEC 61326-1
What is the Triple-Gas Hypoxia CO2 Incubator?
A triple-gas incubator controls oxygen as well as carbon dioxide by bleeding nitrogen into the chamber, holding oxygen from 0.1% to 21% for stem cell, embryo and tumour cultures that respond to physiological oxygen levels.
What is the Triple-Gas Hypoxia CO2 Incubator?
A triple-gas incubator controls oxygen as well as carbon dioxide by bleeding nitrogen into the chamber, holding oxygen from 0.1% to 21% for stem cell, embryo and tumour cultures that respond to physiological oxygen levels.
Overview
The instrument adds a nitrogen supply and an oxygen sensor, usually zirconia or electrochemical, to the standard temperature and CO2 control. Nitrogen is injected to displace oxygen until the set point is reached, and a controller maintains the balance while CO2 is regulated independently. Because dissolved oxygen affects metabolism, differentiation and drug response, hypoxic conditions of 1–5% oxygen are commonly used for stem cells, embryos, cartilage and tumour spheroids, and anoxic conditions near 0.1% for specific studies. Oxygen control adds cost, gas consumption and complexity, and sensors need periodic calibration. Chamber recovery after door opening is slower than for oxygen only controlled by ambient air, so inner doors and minimal access are important.
How does it work?
Cells in culture cannot regulate their own environment, so the incubator supplies what the medium cannot. Temperature is held at 37 °C by a water jacket, an air jacket or direct heating elements around the chamber, with a PID controller and an over-temperature cut-out. Carbon dioxide is measured by an infrared or thermal conductivity sensor and injected from a supply cylinder to maintain 5% CO2, which keeps bicarbonate-buffered medium at physiological pH. A water pan or humidity injection keeps relative humidity above 90%, preventing evaporation that would concentrate salts and osmolytes and change the pH set point. Copper-lined or antimicrobial interior panels and HEPA filters reduce contamination risk, and an inner glass door limits gas exchange when the outer door is opened. Some models add nitrogen to control oxygen for hypoxic culture.
What are the key features?
- Nitrogen injection for oxygen control from 0.1% to 21%
- Zirconia or electrochemical oxygen sensor with calibration routine
- Independent CO2 control at 5% for bicarbonate buffering
- Programmable oxygen profiles for differentiation studies
- Inner glass door to limit gas exchange on access
- Data logging of oxygen, CO2 and temperature for traceability
What are the technical specifications?
A typical triple-gas hypoxia co2 incubator configuration offers oxygen range of 0.1–21%, oxygen control accuracy of within ±0.2% at low set points, CO2 range of 0–20%, controlled at 5%, temperature range of ambient +5 to 50 °C, set at 37 °C, oxygen sensor of Zirconia or electrochemical, calibration every 1–3 months and gas supply of CO2 and nitrogen, or premixed triple gas. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Triple-Gas Hypoxia CO2 Incubator is used in stem cell and iPSC culture at physiological oxygen, embryo and in vitro fertilisation work, tumour spheroid and organoid culture and hypoxia signalling and metabolism research. Across the wider CO2 Incubators class, laboratories also deploy it for routine mammalian cell line culture, stem cell and primary cell maintenance and vaccine and recombinant protein production.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-010, IEC 61326-1, EN 12469, ISO 14644-1 and USP <1046>. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install the incubator on a level surface away from direct sunlight, heat sources and draughts, leaving clearance for the gas connection and ventilation. Fill the humidity pan with sterile distilled water and change it weekly, since stagnant water is a common contamination source; some models use humidity injection that avoids standing water. Verify temperature and CO2 accuracy at installation and at least every six months with a calibrated external instrument, and record recovery time after door opening. Do not overload shelves, because air circulation is needed for uniformity. Handle spills immediately, use only compatible disinfectants, and follow the manufacturer's decontamination procedure before servicing. Keep a log of cleaning, calibration and decontamination events so that culture problems can be traced.
How do I select the right Triple-Gas Hypoxia CO2 Incubator?
Select a triple-gas incubator when oxygen tension is part of the experimental variable or when physiological oxygen improves cell performance. Specify oxygen range and accuracy, sensor type and calibration interval, gas supply arrangement and data logging capability.
How does it compare with other CO2 Incubators variants?
Choose the chamber type first. Water-jacketed incubators hold temperature most stably through power interruptions and suit laboratories with frequent door openings, while air-jacketed and direct-heat models warm up faster, are lighter and need less maintenance. Then consider contamination control: copper interior surfaces, HEPA filtration of the air circuit and high-temperature or vaporised hydrogen peroxide decontamination cycles reduce losses in busy laboratories. Check CO2 and temperature accuracy and uniformity, recovery time after door opening, humidity control, and whether oxygen control is needed for stem cell or hypoxia work. Review alarm functions, data logging, access control and remote monitoring if the incubator holds irreplaceable cultures or is part of a regulated process. Confirm stackability, shelf adjustability and available chamber volumes before ordering.
Full specifications
| Oxygen range | 0.1–21% |
|---|---|
| Oxygen control accuracy | within ±0.2% at low set points |
| CO2 range | 0–20%, controlled at 5% |
| Temperature range | ambient +5 to 50 °C, set at 37 °C |
| Oxygen sensor | Zirconia or electrochemical, calibration every 1–3 months |
| Gas supply | CO2 and nitrogen, or premixed triple gas |
| Chamber volume | 100–200 L |
| Humidity | above 90% relative humidity |
| Recovery after door opening | approx. 5–15 min for oxygen set point |
| Dimensions (W x D x H) | approx. 650 x 700 x 950 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Stem cell and iPSC culture at physiological oxygen
- Embryo and in vitro fertilisation work
- Tumour spheroid and organoid culture
- Hypoxia signalling and metabolism research
Key features
- Nitrogen injection for oxygen control from 0.1% to 21%
- Zirconia or electrochemical oxygen sensor with calibration routine
- Independent CO2 control at 5% for bicarbonate buffering
- Programmable oxygen profiles for differentiation studies
- Inner glass door to limit gas exchange on access
- Data logging of oxygen, CO2 and temperature for traceability
Frequently asked questions about Triple-Gas Hypoxia CO2 Incubator
Many tissues experience 1–8% oxygen in vivo, so standard 21% air is hyperoxic for them. Physiological oxygen reduces oxidative stress, supports stem cell pluripotency and alters differentiation and metabolic pathways, making results more relevant to tissue conditions.
Every one to three months, depending on sensor type, and after any gas supply interruption or chamber decontamination. Zirconia sensors drift with use, so calibration against a reference gas should be recorded in the instrument log.
Yes. Nitrogen is consumed continuously to maintain low oxygen, and recovery after door openings uses additional gas. Laboratories should budget for both gases and confirm that regulators, lines and cylinder capacity suit the duty cycle.
Water-jacketed chambers hold temperature more stably during power cuts and heavy door use, but are heavier and need more maintenance. Air-jacketed and direct-heat units warm up faster, recover quickly and are easier to move, which suits most routine laboratories.
Without high humidity, medium evaporates and concentrates salts and proteins, raising osmolarity and shifting pH. This changes growth rates and can make results from long experiments inconsistent, especially in small-volume or multi-well formats.
Infrared sensors typically hold CO2 within ±0.1–0.3% of the set point and are unaffected by humidity, while thermal conductivity sensors are cheaper but sensitive to moisture. Accuracy should be verified with a calibrated analyser at least twice a year.