Semi-Automated Biochemistry Analyzer
半自动生化分析仪
- SKU CLA-0002
- Clinical Analyzers
- CE (to be confirmed)
- ISO 13485 quality system (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Test principle | Absorbance photometry |
|---|---|
| Wavelength selection | Interference filters, 340–630 nm |
| Measurement modes | End-point, fixed-time, kinetic, differential |
| Flow cell volume | 32 µL (typical) |
| Sample volume | 10–50 µL |
| Photometric range | 0–2.5 Abs |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 15189, CLSI EP05-A3, CLSI EP06
What is the Semi-Automated Biochemistry Analyzer?
A semi-automated biochemistry analyzer is a photometer with a flow cell or cuvette holder in which the operator pipettes samples and reagents manually or aspirates them by pump. It fits low-volume laboratories, research use, veterinary clinics and teaching.
What is the Semi-Automated Biochemistry Analyzer?
A semi-automated biochemistry analyzer is a photometer with a flow cell or cuvette holder in which the operator pipettes samples and reagents manually or aspirates them by pump. It fits low-volume laboratories, research use, veterinary clinics and teaching.
Overview
A semi-automated analyzer measures absorbance in a thermostatted flow cell or cuvette block, while the operator prepares the reaction by pipetting sample and reagent. Some models aspirate the mixed reaction through a peristaltic pump into a 32 µL flow cell, which keeps reagent use low. Wavelength selection is by interference filters, commonly 340, 405, 505, 546, 578 and 620 nm, or by a grating. The instrument supports end-point, fixed-time, kinetic and differential measurement modes and stores test parameters for dozens of assays. Calibration and quality control functions are simpler than on automated systems and results print on an internal thermal printer or export by USB. The design is robust, inexpensive to maintain and well suited to veterinary clinics, small hospitals, training laboratories and as a back-up for automated analyzers.
How does it work?
Most clinical analyzers quantify an analyte by converting its concentration into a measurable signal. Photometric analyzers mix sample with reagent in a reaction cuvette, hold it at 37 °C, and measure absorbance at selected wavelengths through a grating or filter wheel; concentration follows the Beer–Lambert relation against a calibration curve. Ion-selective electrode modules measure the potential of Na+, K+ and Cl- across a selective membrane relative to a reference electrode. Turbidimetric and nephelometric modules detect light scattered by antigen-antibody complexes. Chromatographic analyzers separate hemoglobin fractions by cation-exchange HPLC. Key components include a sample and reagent probe with liquid-level detection, a precision syringe, a thermostatted reaction unit, a detection optic, and software that applies calibration, quality control rules and result flagging.
What are the key features?
- Flow cell of about 32 µL with peristaltic aspiration
- Interference filters for 6 to 8 common wavelengths
- End-point, fixed-time and kinetic measurement modes
- Thermostatted cell holder at 25, 30 or 37 °C
- Internal thermal printer and USB data export
- Low reagent consumption with a simple operation menu
What are the technical specifications?
A typical semi-automated biochemistry analyzer configuration offers test principle of Absorbance photometry, wavelength selection of Interference filters, 340–630 nm, measurement modes of End-point, fixed-time, kinetic, differential, flow cell volume of 32 µL (typical), sample volume of 10–50 µL and photometric range of 0–2.5 Abs. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Semi-Automated Biochemistry Analyzer is used in small hospital and clinic laboratories, veterinary clinical chemistry, teaching and training laboratories and back-up testing during analyzer downtime. Across the wider Clinical Analyzers class, laboratories also deploy it for routine serum chemistry, electrolyte and blood gas testing and diabetes monitoring (HbA1c).
Which standards and certifications apply?
Relevant standards and methods include ISO 15189, CLSI EP05-A3, CLSI EP06, CLSI C24, IEC 61010-2-101 and IEC 61010-1. Certification candidates for this model are CE and ISO 13485 quality system; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install on a level bench away from direct sunlight, vibration and strong electromagnetic sources, with stable power, a dedicated earth and a supply of purified water that meets CLSI Type II quality where the analyzer requires it. Verify performance before clinical use through precision, linearity, carryover and method comparison studies in line with CLSI EP protocols. Run internal quality control every day and join an external quality assessment scheme. Schedule daily probe cleaning, weekly lamp and cuvette checks, and replacement of tubing, syringes and lamps per the maintenance plan. Treat all samples and waste as biohazardous and follow local regulations for disposal.
How do I select the right Semi-Automated Biochemistry Analyzer?
Choose this variant when daily chemistry volume is below about 100 tests, when budget is limited, or where a back-up to an automated analyzer is needed. In the RFQ specify filter wavelengths, flow cell volume, temperature setting, number of stored programs and printer type. Move to an automated benchtop unit as volume grows.
How does it compare with other Clinical Analyzers variants?
Start with the test menu and daily workload, because throughput and reagent format set the platform. Low volumes of routine chemistry suit a semi-automated photometer, while several hundred samples a day call for a benchtop random-access analyzer. Decide whether you need open reagent channels, which lower cost per test but require you to validate each assay, or closed reagent systems that simplify operation. Check sample volume needs for pediatric work, onboard reagent capacity, cuvette type and waste handling. For electrolytes, blood gas or HbA1c, choose a dedicated analyzer rather than stretching a general chemistry unit. Confirm LIS connectivity through HL7 or ASTM, barcode reading, and the availability of service engineers and spare parts in your region.
Full specifications
| Test principle | Absorbance photometry |
|---|---|
| Wavelength selection | Interference filters, 340–630 nm |
| Measurement modes | End-point, fixed-time, kinetic, differential |
| Flow cell volume | 32 µL (typical) |
| Sample volume | 10–50 µL |
| Photometric range | 0–2.5 Abs |
| Temperature control | 25/30/37 °C, ±0.2 °C |
| Stored test programs | 100–200 programs |
| Light source | Halogen lamp, 6 V 10 W |
| Display | Color LCD or monochrome LCD |
| Power supply | 100–240 V AC, 50/60 Hz, about 100 VA |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Small hospital and clinic laboratories
- Veterinary clinical chemistry
- Teaching and training laboratories
- Back-up testing during analyzer downtime
Key features
- Flow cell of about 32 µL with peristaltic aspiration
- Interference filters for 6 to 8 common wavelengths
- End-point, fixed-time and kinetic measurement modes
- Thermostatted cell holder at 25, 30 or 37 °C
- Internal thermal printer and USB data export
- Low reagent consumption with a simple operation menu
Frequently asked questions about Semi-Automated Biochemistry Analyzer
Yes, when the operator pipettes carefully with calibrated pipettes and controls are run daily. The photometric stage is comparable to automated systems; most variation comes from manual pipetting, so staff training and regular pipette verification are important.
A trained operator typically completes 40 to 80 tests per hour on a semi-automated analyzer, depending on incubation time and the number of samples batched. It is not suited to high-throughput work or to tests needing tight timing across many samples.
Generally yes. Users enter wavelength, mode, factor or calibrator concentration, and incubation time from the kit insert. Since there is no barcode or vendor lock, every new assay should be verified with control material before patient results are reported.
A clinical analyzer measures concentrations of substances such as glucose, enzymes, electrolytes, lipids and proteins in patient samples. Results support diagnosis, treatment decisions and monitoring in hospitals, clinics and reference laboratories. Different platforms cover chemistry, electrolytes, blood gases, specific proteins and HbA1c.
An open system accepts reagents from any supplier once channel parameters are programmed and validated, which gives flexibility and lower cost per test. A closed system uses reagents supplied for that platform with pre-set parameters and barcodes, which simplifies operation but limits supplier choice.
Calibration frequency depends on the assay and reagent lot. Many photometric tests are recalibrated at each new reagent lot, after major maintenance, or when quality control shifts beyond limits. Electrolyte electrodes are typically calibrated automatically every few hours. Follow the reagent instructions for use and your laboratory procedure.