Fluorescence Cell Imager
荧光细胞成像仪
- SKU CEI-0005
- Cell Imaging
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Magnification | Fixed 4x or 10x equivalent |
|---|---|
| Imaging channels | Brightfield plus 2–4 fluorescence channels |
| Excitation | approx. 365, 470 and 550 nm options |
| Camera | Colour and monochrome CMOS, 5–12 MP |
| Vessel compatibility | Flasks, dishes, multi-well plates and slides |
| Confluence measurement | 0–100% from brightfield contrast |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 60825-1, IEC 61010-1, IEC 61010-2-010
What is the Fluorescence Cell Imager?
A benchtop fluorescence cell imager is a compact instrument with fixed optics and pre-set channels that captures brightfield and fluorescence images of cells in plates, giving quick visual confirmation of confluence, viability and transfection without a full microscope setup.
What is the Fluorescence Cell Imager?
A benchtop fluorescence cell imager is a compact instrument with fixed optics and pre-set channels that captures brightfield and fluorescence images of cells in plates, giving quick visual confirmation of confluence, viability and transfection without a full microscope setup.
Overview
The imager replaces the adjustable optics of a microscope with fixed magnification and pre-aligned illumination and filter combinations for common dyes such as DAPI, GFP and propidium iodide. Users load a plate, flask or slide, select a protocol, and the instrument focuses, captures and stores images together with counts of labelled objects. Confluence measurement uses brightfield contrast to estimate the percentage of the surface covered by cells, which helps decide when to passage or harvest. Because optics are fixed, comparative images taken at different times are directly comparable, and the small footprint allows installation in a cell culture hood. Resolution and flexibility are lower than a full microscope, and analysis is limited to the measurements offered in the software.
How does it work?
All cell imaging systems collect light from a specimen and form an image on a detector. In widefield microscopy, the whole field is illuminated and the image is limited by out-of-focus light, so thin cells look sharp while thick specimens appear hazy. Confocal microscopy places a pinhole in front of the detector so that only light from the focal plane reaches it, and it scans point by point or line by line to build an optical section; this improves contrast and allows three-dimensional reconstruction. Fluorescence excitation comes from LEDs, lasers or arc lamps, and dichroic mirrors and emission filters separate the excitation from the emitted signal. Live-cell systems add an incubated stage or chamber to hold temperature, CO2 and humidity, and software captures time-lapse sequences. High-content systems combine automated plate handling, autofocus and image analysis to extract numerical features from thousands of images.
What are the key features?
- Fixed optics with pre-set brightfield and fluorescence channels
- Confluence measurement in percent from brightfield images
- Object counting for labelled nuclei or fluorescent cells
- Protocol-based operation requiring minimal training
- Compact housing suitable for placement in a culture hood
- Side-by-side comparison of images from different time points
What are the technical specifications?
A typical fluorescence cell imager configuration offers magnification of Fixed 4x or 10x equivalent, imaging channels of Brightfield plus 2–4 fluorescence channels, excitation of approx. 365, 470 and 550 nm options, camera of Colour and monochrome CMOS, 5–12 MP, vessel compatibility of Flasks, dishes, multi-well plates and slides and confluence measurement of 0–100% from brightfield contrast. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Fluorescence Cell Imager is used in confluence checks before passaging or harvest, transfection and transduction efficiency verification, viability and cytotoxicity visual checks and documentation of culture status for records. Across the wider Cell Imaging class, laboratories also deploy it for cell morphology and confluence assessment, immunofluorescence and protein localisation and live-cell dynamics and time-lapse studies.
Which standards and certifications apply?
Relevant standards and methods include IEC 60825-1, IEC 61010-1, IEC 61010-2-010, ISO 17025 and ISO 20387. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Optical cleanliness is essential, so objectives should be cleaned with lens tissue and the recommended solvent, and immersion oil removed after every session. Live-cell imaging requires careful control of temperature, CO2 and humidity, because evaporation changes osmolarity and focus drift spoils long time-lapse runs; hardware autofocus or a focus-maintenance routine is usually needed. Phototoxicity and photobleaching limit live experiments, so use the lowest illumination power and shortest exposure that gives an acceptable signal, and reduce the number of channels and time points. Laser systems fall under IEC 60825-1 and must remain interlocked. After installation, verify alignment with a calibration slide, then check stage repeatability and focus stability periodically, and keep records so that images from different sessions remain comparable.
How do I select the right Fluorescence Cell Imager?
Select a compact fluorescence imager for routine culture checks and documentation when a full microscope is unnecessary or unavailable in the laboratory. Specify available channels, magnification, confluence measurement, vessel compatibility and whether counts of labelled objects are reported by the software.
How does it compare with other Cell Imaging variants?
Match the system to the spatial and temporal resolution the experiment needs. Routine morphology, transfection checks and immunofluorescence on coverslips are well served by an inverted fluorescence microscope with LED illumination and a scientific camera. Confocal systems are needed when optical sections, thick specimens, three-dimensional reconstruction or low background matter, and spinning-disk confocals add speed for live imaging. Live-cell work requires an incubation chamber with CO2 and temperature control, plus low phototoxicity and a camera sensitive enough to limit exposure. High-content platforms are chosen when hundreds to thousands of wells must be imaged and quantified automatically. Check objective quality and numerical aperture, available fluorescence channels, detector sensitivity, stage automation, autofocus reliability, incubation performance and the analysis software, which often matters more than optics for quantitative results.
Full specifications
| Magnification | Fixed 4x or 10x equivalent |
|---|---|
| Imaging channels | Brightfield plus 2–4 fluorescence channels |
| Excitation | approx. 365, 470 and 550 nm options |
| Camera | Colour and monochrome CMOS, 5–12 MP |
| Vessel compatibility | Flasks, dishes, multi-well plates and slides |
| Confluence measurement | 0–100% from brightfield contrast |
| Capture time | approx. 5–30 s per image |
| Data output | PNG or TIFF images with confluence and count values |
| Cleaning | Wipe-down surface compatible with 70% ethanol |
| Dimensions (W x D x H) | approx. 200 x 250 x 200 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Confluence checks before passaging or harvest
- Transfection and transduction efficiency verification
- Viability and cytotoxicity visual checks
- Documentation of culture status for records
Key features
- Fixed optics with pre-set brightfield and fluorescence channels
- Confluence measurement in percent from brightfield images
- Object counting for labelled nuclei or fluorescent cells
- Protocol-based operation requiring minimal training
- Compact housing suitable for placement in a culture hood
- Side-by-side comparison of images from different time points
Frequently asked questions about Fluorescence Cell Imager
It typically agrees within a few percentage points of manual estimates for healthy monolayers, but accuracy falls for cells with elongated morphology, dense overgrowth or debris. Consistency improves when the same protocol and vessel type are used.
Yes, models with a compact footprint and wipe-down surfaces are designed for that use, which avoids transporting cultures and reduces contamination risk. Confirm the housing is resistant to 70% ethanol and that heat output is low.
No. It gives a fixed magnification and a small set of channels, so it cannot perform detailed morphology, high-resolution imaging or spectral separation. It is best used alongside a microscope for routine checks.
Confocal is needed when out-of-focus light obscures the signal, for example in thick specimens, three-dimensional reconstructions, thick tissue and sections with high background. For thin cultured cells and bright markers, widefield imaging with a good camera is often sufficient.
An incubation chamber maintaining 37 °C, CO2 and humidity, hardware autofocus to correct focus drift, and low illumination intensity to keep phototoxicity acceptable. Objectives with a long working distance also help when using thick culture vessels.
Use the lowest light intensity and shortest exposure that still gives adequate signal, reduce the number of fluorescence channels and time points, choose fluorophores excited by longer wavelengths, and use widefield or spinning-disk systems where confocal illumination would be excessive.