Automated Image-Based Cell Counter
自动化图像法细胞计数仪
- SKU CEC-0001
- Cell Counters
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Sample volume | 10–20 µL per count |
|---|---|
| Counting time | approx. 30 s |
| Cell size range | approx. 2–70 µm |
| Concentration range | approx. 1 x 10⁴–1 x 10⁷ cells/mL |
| Accuracy | within ±10% of reference method (typical) |
| Imaging method | Brightfield with fixed or autofocus optics |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 20387, ISO 15189, IEC 61010-1
What is the Automated Image-Based Cell Counter?
An automated image-based cell counter loads a small sample into a disposable slide, images a defined volume in transmitted light and reports concentration and viability from trypan blue exclusion. It replaces manual haemocytometer counting in routine culture work.
What is the Automated Image-Based Cell Counter?
An automated image-based cell counter loads a small sample into a disposable slide, images a defined volume in transmitted light and reports concentration and viability from trypan blue exclusion. It replaces manual haemocytometer counting in routine culture work.
Overview
The instrument combines a brightfield optics module, a motorised stage or fixed focus and an image analysis algorithm. A 10–20 µL sample mixed with trypan blue is drawn into a disposable slide with a defined chamber depth, and the software identifies objects by size and contrast, excluding debris below a size threshold and classifying blue-stained cells as non-viable. Results are displayed in about 30 seconds with a size distribution histogram. Typical sizing range is 2–70 µm, concentration linearity from about 10⁴ to 10⁷ cells/mL, and accuracy within 10% of a reference method when the sample is well mixed. The format is simple, requires little training and avoids the subjectivity of manual counting, but it can overestimate viability in samples with many dead-cell fragments or red blood cells.
How does it work?
Counting methods differ in how they distinguish a cell from debris and from a dead cell. Trypan blue exclusion relies on the intact membrane of viable cells, which excludes the dye, so live cells appear bright and dead cells blue in transmitted light; automated counters image a fixed volume in a counting chamber and classify objects by size, shape and brightness. Fluorescence methods use acridine orange and propidium iodide, which stain all nuclei and membrane-damaged nuclei respectively, so concentration and viability are derived from two channels and are less affected by red blood cells, platelets and debris. Electrical impedance counters draw cells through a small aperture and count voltage pulses, where pulse height reflects cell volume. Flow cytometry counters add a known number of reference beads and calculate absolute counts from the ratio of cell to bead events, which also provides marker information. All methods require a well-mixed, appropriately diluted sample to be accurate.
What are the key features?
- Brightfield imaging with trypan blue for concentration and viability
- Counting time of about 30 seconds per sample
- Size distribution histogram with adjustable gate limits
- Disposable slides with defined chamber depth
- Concentration linearity from about 10⁴ to 10⁷ cells/mL
- Results exportable to spreadsheet or laboratory records
What are the technical specifications?
A typical automated image-based cell counter configuration offers sample volume of 10–20 µL per count, counting time of approx. 30 s, cell size range of approx. 2–70 µm, concentration range of approx. 1 x 10⁴–1 x 10⁷ cells/mL, accuracy of within ±10% of reference method (typical) and imaging method of Brightfield with fixed or autofocus optics. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Automated Image-Based Cell Counter is used in routine cell line culture monitoring, seeding density calculations, transfection and infection normalisation and teaching laboratory cell culture. Across the wider Cell Counters class, laboratories also deploy it for seeding density determination before experiments, culture growth and viability monitoring and cell therapy product release counts.
Which standards and certifications apply?
Relevant standards and methods include ISO 20387, ISO 15189, IEC 61010-1, IEC 60825-1, USP <1046> and ISO 17025. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Accuracy depends more on sample handling than on the instrument. Cells settle quickly, so mix by gentle inversion or low-speed pipetting immediately before sampling, and avoid vigorous vortexing that damages membranes. Dilute into the method's linear range, since concentrated samples saturate the sensor and very dilute samples lose precision. Consumables such as counting slides or chambers should be used once and disposed of as biohazard waste. Clean the optics and sample path regularly, and avoid samples containing strong dyes or fluorescent compounds that interfere with fluorescence channels. Validate each new cell type or matrix against a reference method such as a haemocytometer or a validated flow count, and document the dilution and settings used, because viability definitions differ between laboratories.
How do I select the right Automated Image-Based Cell Counter?
Select an image-based counter for routine culture work with established cell lines where a trypan blue count is sufficient. Specify size range, concentration linearity, counting time, slide consumable cost and whether autofocus or fixed optics are supplied.
How does it compare with other Cell Counters variants?
Choose by sample type and required accuracy. For routine culture of established cell lines, an image-based counter with trypan blue is adequate and inexpensive. When samples contain debris, red cells, platelets or clumps, or when viability is low and must be measured precisely, choose a fluorescence counter with acridine orange and propidium iodide, which reports concentration and viability from nuclear staining. Impedance counters are useful for continuous monitoring of cell size distribution and for samples where optical clarity varies. Flow cytometry counters are the right choice when counts must be combined with surface marker measurement or when absolute counts are required for clinical reporting. Check the linear range, sample volume, time per count, consumable cost, cleaning requirements and whether the instrument reports aggregate or clump correction.
Full specifications
| Sample volume | 10–20 µL per count |
|---|---|
| Counting time | approx. 30 s |
| Cell size range | approx. 2–70 µm |
| Concentration range | approx. 1 x 10⁴–1 x 10⁷ cells/mL |
| Accuracy | within ±10% of reference method (typical) |
| Imaging method | Brightfield with fixed or autofocus optics |
| Viability dye | Trypan blue |
| Consumables | Single-use counting slides |
| Data output | CSV export with histogram data |
| Dimensions (W x D x H) | approx. 250 x 300 x 250 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Routine cell line culture monitoring
- Seeding density calculations
- Transfection and infection normalisation
- Teaching laboratory cell culture
Key features
- Brightfield imaging with trypan blue for concentration and viability
- Counting time of about 30 seconds per sample
- Size distribution histogram with adjustable gate limits
- Disposable slides with defined chamber depth
- Concentration linearity from about 10⁴ to 10⁷ cells/mL
- Results exportable to spreadsheet or laboratory records
Frequently asked questions about Automated Image-Based Cell Counter
Agreement is typically within 10% when the sample is well mixed and diluted into range, and reproducibility is usually better than manual counting because the imaged volume and object classification are fixed. Systematic differences appear with clumped or unusually small cells.
Most systems are factory calibrated, but periodic checks with reference beads or a counted suspension are recommended, especially before critical experiments. Checks confirm that optics, slide depth and software thresholds remain consistent.
Trypan blue can be excluded by cells with damaged but intact membranes, and small dead-cell fragments may fall below the size threshold and be ignored. Fluorescence nuclear dyes give a more sensitive viability assessment in such samples.
Most benchtop counters use 10–20 µL per count, and some fluorescence systems accept as little as 10 µL when viability is measured from nuclear staining. Samples are usually diluted with dye or buffer before counting.
Trypan blue measures membrane integrity, whereas fluorescence nuclear dyes also detect early apoptosis and can include cells that have lost nuclear structure. Instruments also differ in how they classify small or clumped objects, so methods should not be mixed within a study.
Gently pipette the suspension with a wide-bore tip or filter through a 40 µm mesh to break clumps before counting. Some instruments apply a clump correction factor, which should be verified against a reference count for each cell type.
Related models