Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions
浓悬浮液电声法 Zeta 电位分析仪
- SKU ZPA-0005
- Zeta Potential Analyzers
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Concentration range | 0.1–40 vol% |
|---|---|
| Zeta potential range | −200 to +200 mV |
| Particle size range (acoustic) | ≈ 5 nm–100 µm |
| Frequency range | 1–100 MHz (attenuation) |
| Sample volume | 20–200 mL |
| pH range | 2–12 |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 13099-1, ISO 13099-2, ISO 13099-3
What is the Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions?
An electroacoustic analyzer that measures zeta potential and particle size in concentrated, opaque suspensions and emulsions without dilution, using colloid vibration current or ultrasonic attenuation. It fits process-relevant slurries in ceramics, battery, mineral and coating production where dilution would change the sample.
What is the Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions?
An electroacoustic analyzer that measures zeta potential and particle size in concentrated, opaque suspensions and emulsions without dilution, using colloid vibration current or ultrasonic attenuation. It fits process-relevant slurries in ceramics, battery, mineral and coating production where dilution would change the sample.
Overview
Electroacoustic methods apply a high-frequency alternating field or ultrasound to a stirred sample. Particles with a density different from the fluid oscillate, generating a measurable colloid vibration current or electrokinetic sonic amplitude, from which dynamic mobility and zeta potential are obtained. Ultrasonic attenuation spectroscopy in the same probe provides particle size at volume fractions up to about 40 percent. There is no need to dilute, which preserves the real particle interactions in a slurry, and titration with pH and additive dosing is easy to add. Particle density, fluid density and viscosity are required inputs, and solids volume fraction should be above a minimum of around 0.1 to 1 percent. The technique is less sensitive to very fine nanoparticles, and sample volume is typically 20 to 200 mL.
How does it work?
Charged particles in a liquid are surrounded by an electrical double layer. The zeta potential is the potential at the slipping plane of this layer, and cannot be measured directly. Electrophoretic light scattering applies an electric field and measures particle velocity from the Doppler shift of scattered laser light; electrophoretic mobility is then converted to zeta potential by the Smoluchowski or Hückel approximation, or the Henry function for intermediate cases. Phase analysis light scattering improves sensitivity by measuring phase rather than frequency shift. Streaming potential instruments force electrolyte through a capillary or plug and measure the generated potential difference for solids. Electroacoustic methods detect sound waves generated by an applied alternating field in concentrated suspensions. Key components include the laser, electrodes and cell, temperature control and often a titrator.
What are the key features?
- Measures dispersions up to about 40 vol% without dilution
- Colloid vibration current or electrokinetic sonic amplitude detection
- Ultrasonic attenuation spectroscopy for size in the same probe
- Stirred vessel with integrated pH, conductivity and temperature probes
- Titration module for pH and additive dosing
- Suitable for opaque and dark slurries
What are the technical specifications?
A typical electroacoustic zeta potential analyzer for concentrated suspensions configuration offers concentration range of 0.1–40 vol%, zeta potential range of −200 to +200 mV, particle size range (acoustic) of ≈ 5 nm–100 µm, frequency range of 1–100 MHz (attenuation), sample volume of 20–200 mL and pH range of 2–12. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions is used in ceramic and alumina slurry dispersion control, lithium-ion battery electrode slurries, coal, ore and mineral tailings, paints, inks and coatings and chemical mechanical polishing slurries. Across the wider Zeta Potential Analyzers class, laboratories also deploy it for colloidal stability of emulsions and suspensions, protein and biologic formulation screening and water and wastewater coagulation control.
Which standards and certifications apply?
Relevant standards and methods include ISO 13099-1, ISO 13099-2, ISO 13099-3, ASTM D4187, GB/T 32671 and ISO 13320. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Zeta potential depends on pH, ionic strength, temperature and sample history, so each report should state dispersant, pH, conductivity and temperature. Use low-conductivity, filtered dispersants and avoid bubbles at the electrodes; high salt can cause electrode polarisation and sample heating. Verify performance periodically using a zeta potential transfer standard, typically around −42 mV, and follow the supplier's interval. Clean cells thoroughly, since contamination and electrode degradation alter results. Gold or palladium electrodes and disposable capillary cells simplify maintenance. For regulated use, carry out IQ/OQ/PQ and maintain audit trails. Report the model used to convert mobility to zeta potential.
How do I select the right Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions?
Choose the electroacoustic variant when samples are concentrated, opaque or cannot be diluted without changing their state. Specify solids loading, particle density, size range, slurry viscosity, pH range, wetted material compatibility, and whether in-line or batch use is planned. Dilute transparent colloids are cheaper to measure with light-scattering electrophoresis.
How does it compare with other Zeta Potential Analyzers variants?
Begin with the sample type. Dilute colloids and macromolecules suit electrophoretic light scattering; very low-mobility or high-conductivity samples benefit from phase analysis light scattering. Concentrated opaque suspensions, such as slurries, need electroacoustic methods, and flat or fibrous solids need streaming potential. If isoelectric point or pH dependence is required, choose a system with an automatic titrator. Check the required size range, conductivity limits, minimum sample volume and whether non-aqueous solvents are used, since the dielectric constant must be known. Decide if zeta potential alone is enough or whether combined size measurement is needed. Also weigh software validation support, cell durability and cost of disposable cells, and whether ISO 13099 conformity is expected in your methods.
Full specifications
| Concentration range | 0.1–40 vol% |
|---|---|
| Zeta potential range | −200 to +200 mV |
| Particle size range (acoustic) | ≈ 5 nm–100 µm |
| Frequency range | 1–100 MHz (attenuation) |
| Sample volume | 20–200 mL |
| pH range | 2–12 |
| Temperature range | 10–50 °C |
| Required inputs | Particle density, fluid density, viscosity |
| Stirring | Variable speed, ≤ 500 rpm |
| Probe | Immersion, steel or titanium wetted parts |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Ceramic and alumina slurry dispersion control
- Lithium-ion battery electrode slurries
- Coal, ore and mineral tailings
- Paints, inks and coatings
- Chemical mechanical polishing slurries
Key features
- Measures dispersions up to about 40 vol% without dilution
- Colloid vibration current or electrokinetic sonic amplitude detection
- Ultrasonic attenuation spectroscopy for size in the same probe
- Stirred vessel with integrated pH, conductivity and temperature probes
- Titration module for pH and additive dosing
- Suitable for opaque and dark slurries
Frequently asked questions about Electroacoustic Zeta Potential Analyzer for Concentrated Suspensions
Sound waves and alternating fields pass through opaque concentrated suspensions, so the method does not rely on light transmission. Measuring in the undiluted state preserves particle interactions that dilution can change.
Particle density, solvent density and viscosity, and usually the dielectric constant and solids fraction. Errors in particle density translate directly into zeta potential errors, so the values should be verified.
Sensitivity drops for very small particles because the acoustic signal is weak, so results may be noisy at low concentrations. Light-scattering electrophoresis on a diluted sample is better for dilute nanoparticle studies.
A large absolute zeta potential, commonly beyond ±30 mV, indicates strong electrostatic repulsion and usually good colloidal stability. Values near zero suggest particles can approach and aggregate. The guideline depends on whether steric stabilisers also contribute.
The isoelectric point is the pH at which zeta potential is zero and colloids are least stable. It is found by titrating the sample across a pH range with acid or base and recording zeta potential at each step, usually with an automatic titrator.
Zeta potential changes with pH, ionic strength and composition of the medium. A number without these conditions cannot be reproduced or compared, so method records must include dispersant, pH, conductivity and temperature.