Residual Stress X-Ray Diffractometer
残余应力 X 射线衍射仪
- SKU XRA-0005
- X-Ray Diffractometers
- CE (to be confirmed)
- Radiation safety certification (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Measurement method | sin-squared-psi |
|---|---|
| Psi range | 0–45° |
| Beam size | 0.5–5 mm |
| Depth range | 1–100 µm with layer removal |
| Stress accuracy | ±20 MPa typical |
| X-ray tube power | 30–300 W portable |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E915 Verifying the alignment of X-ray diffraction instrumentation, ASTM E1426 Determining the effective elastic parameter for X-ray diffraction measurements of residual stress, USP <941> X-Ray Diffraction
What is the Residual Stress X-Ray Diffractometer?
A residual stress X-ray diffractometer measures lattice strain by tracking the shift of a diffraction peak with sample tilt. It is chosen for welds, machined parts, coatings and heat-treated components in engineering.
What is the Residual Stress X-Ray Diffractometer?
A residual stress X-ray diffractometer measures lattice strain by tracking the shift of a diffraction peak with sample tilt. It is chosen for welds, machined parts, coatings and heat-treated components in engineering.
Overview
The instrument measures the interplanar spacing of a chosen reflection at several sample tilt angles, following the sin-squared-psi method. A shift in peak position with tilt indicates strain, and the stress is calculated from the strain together with the material elastic constants, often with the effective elastic parameter determined by ASTM E1426. Portable or gantry configurations can measure large components such as pipelines, rails and turbine parts on site, while laboratory systems provide higher precision on smaller samples. Sample positioning, beam size and oscillation determine spatial resolution and accuracy. Measurements are affected by texture, coarse grains and surface condition, so electropolishing and oscillation are often used. Results guide welding procedures, machining parameters and component lifing.
How does it work?
X-rays scatter from the electron density of a material. In a crystal, scattering from periodic lattice planes interferes constructively at angles given by Bragg's law, producing diffraction peaks. A diffractometer positions the sample and detector with high angular precision and records intensity versus angle or versus a two-dimensional detector image. Single-crystal instruments determine atomic structure from the full set of reflections, while powder and thin-film instruments analyze phase mixtures, preferred orientation and layer structure. Peak positions give lattice parameters and strain, peak widths relate to crystallite size and microstrain, and low-angle scattering reveals nanoscale features. The choice of optics, monochromator, detector and sample stage determines which measurements are possible, and instrument alignment with certified standards underpins accuracy.
What are the key features?
- Sin-squared-psi stress measurement method
- Tilt or side-inclination sample stage
- Portable or gantry options for large parts
- Oscillation to improve grain statistics
- Electropolishing for surface preparation
- Elastic parameter determination per ASTM E1426
What are the technical specifications?
A typical residual stress x-ray diffractometer configuration offers measurement method of sin-squared-psi, psi range of 0–45°, beam size of 0.5–5 mm, depth range of 1–100 µm with layer removal, stress accuracy of ±20 MPa typical and X-ray tube power of 30–300 W portable. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Residual Stress X-Ray Diffractometer is used in weld and heat-affected zone stress, machined and shot-peened surfaces, coatings and thermal spray layers and turbine, rail and pipeline components. Across the wider X-Ray Diffractometers class, laboratories also deploy it for single-crystal structure determination, thin-film thickness, texture and phase analysis and residual stress in welds, coatings and machined parts.
Which standards and certifications apply?
Relevant standards and methods include ASTM E915 Verifying the alignment of X-ray diffraction instrumentation, ASTM E1426 Determining the effective elastic parameter for X-ray diffraction measurements of residual stress, USP <941> X-Ray Diffraction, ISO 20203 Carbonaceous materials — XRD of calcined coke, ISO/IEC 17025 and IEC 61010-2-091. Certification candidates for this model are CE, Radiation safety certification and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
These are regulated radiation devices requiring shielding, interlocks, radiation surveys and trained operators. Single crystals must be mounted and centered precisely, and low-temperature or humidity devices often protect them. Thin films and wafers need flat mounting and careful alignment to the beam, and reflectometry requires very low background and precise angular steps. Residual stress methods need defined measurement directions and elastic constants. Small-angle systems require long evacuated paths and clean beam stops. Alignment, angular calibration and detector linearity are verified with standards, and X-ray tubes, goniometer motors and area detectors are wear items. Sample preparation and measurement conditions should be recorded for traceability.
How do I select the right Residual Stress X-Ray Diffractometer?
Select a residual stress diffractometer when strain and stress must be measured non-destructively on engineering components, and choose portable or gantry versions for parts that cannot come to the laboratory. Specify the material, grain size, beam size and depth of interest. Confirm that the method and elastic parameter match ASTM E1426 or your company procedure.
How does it compare with other X-Ray Diffractometers variants?
Select a diffractometer configuration by the structural question. Single-crystal systems with four-circle or kappa goniometers and area detectors determine atomic structures from small crystals. Thin-film instruments with grazing-incidence optics and parallel beams measure layer thickness, texture and phase in coatings. Reflectometry configurations measure electron density profiles, interface roughness and layer thickness at very low angles. Residual stress instruments use side-inclination or psi geometry with precise sample tilting to measure strain in metals and ceramics. Small-angle scattering systems with long camera lengths resolve nanometre-scale particles, pores and lamellae. Confirm the angular range and resolution, sample environment, detector type and software for your application, and verify that the instrument can follow any required standard method.
Full specifications
| Measurement method | sin-squared-psi |
|---|---|
| Psi range | 0–45° |
| Beam size | 0.5–5 mm |
| Depth range | 1–100 µm with layer removal |
| Stress accuracy | ±20 MPa typical |
| X-ray tube power | 30–300 W portable |
| Detector | position-sensitive or area |
| Portability | portable or gantry options |
| Sample oscillation | adjustable or automated |
| Interface | Ethernet / wireless |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Weld and heat-affected zone stress
- Machined and shot-peened surfaces
- Coatings and thermal spray layers
- Turbine, rail and pipeline components
Key features
- Sin-squared-psi stress measurement method
- Tilt or side-inclination sample stage
- Portable or gantry options for large parts
- Oscillation to improve grain statistics
- Electropolishing for surface preparation
- Elastic parameter determination per ASTM E1426
Frequently asked questions about Residual Stress X-Ray Diffractometer
The interplanar spacing is measured at several tilt angles and plotted against sine squared psi. For a stressed sample the points form a straight line whose slope is proportional to the strain, and the stress is calculated with the elastic constants. The method is standard for polycrystalline metals.
Yes, portable residual stress analyzers with small X-ray tubes and position-sensitive detectors are used on pipelines, rails, welded structures and large machined parts. They trade some precision and depth capability for portability. Radiation safety procedures and trained operators are required for field work.
When grains are large, only a few diffract crystallographically at each tilt, so the peak position fluctuates and results become unreliable. Sample oscillation over a larger area increases the number of diffracting grains, and sometimes a different reflection or a larger beam is used.
Single-crystal diffraction measures individual reflections from one crystal, giving enough data to solve the full atomic structure, including bond lengths and angles. Powder diffraction measures the overlapping reflections from many randomly oriented crystallites, which identifies and quantifies phases but cannot alone determine a complete atomic structure.
The sample is tilted to several angles and the shift of a chosen diffraction peak is measured at each tilt. Lattice strain is calculated from the peak shift, and stress follows from the strain and the material elastic constants. Methods follow standards such as ASTM E1426 and require defined measurement directions.
At very shallow incidence angles the X-ray beam probes only the near-surface region of a thin film or coating, which reduces signal from the substrate. This allows phase identification, texture and depth profiling of layers only nanometres to micrometres thick, as used for semiconductors, optical coatings and corrosion films.