Combined Type I and Type II Water System
I 级和 II 级水组合系统
- SKU UWS-0002
- Ultrapure Water Systems
- CE (to be confirmed)
- RoHS (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Type II resistivity | 1–15 MΩ·cm |
|---|---|
| Type I resistivity | 18.2 MΩ·cm |
| Type II output | 5–30 L/h |
| Type I flow | 0.5–2 L/min |
| Storage reservoir | 10–100 L |
| Total organic carbon | < 10 ppb for Type I |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 3696, CLSI GP40, USP <643>
What is the Combined Type I and Type II Water System?
A combined Type I and Type II water system delivers both ultrapure and general-grade water from one unit. It suits laboratories that need high purity for analysis and larger volumes for general work.
What is the Combined Type I and Type II Water System?
A combined Type I and Type II water system delivers both ultrapure and general-grade water from one unit. It suits laboratories that need high purity for analysis and larger volumes for general work.
Overview
The combined system produces Type II water through reverse osmosis and ion exchange into a storage reservoir, and Type I water from a polishing loop fed by that reservoir. A valve or dispenser selects the grade. Type II output is typically 5 to 30 liters per hour at a resistivity of 1 to 15 megohm-centimeters, while Type I is drawn at 0.5 to 2 liters per minute at 18.2 megohm-centimeters with low organic carbon. This arrangement is efficient because the storage reservoir buffers demand and the polishing cartridges only treat what is actually needed at high purity, which extends their life. Monitoring usually covers both the reservoir conductivity and the Type I resistivity, with alarms. The unit is common in analytical and life science laboratories that need both grades daily.
How does it work?
An ultrapure water system chains several purification steps to reach Type I quality. Feed, usually reverse osmosis or distilled water, passes through activated carbon to remove organics and chlorine, ion exchange resin to remove ions, a 185 nanometer ultraviolet lamp to oxidize residual organics, a 254 nanometer lamp to inactivate microorganisms, and ultrafiltration or microfiltration to remove particles, bacteria and endotoxins. A final 0.22 micron filter or ultrafiltration membrane polishes the water at the point of use. Resistivity and total organic carbon are monitored continuously, and recirculation keeps the loop fresh. The combination of steps is essential because no single technique removes all contaminant classes, and care in materials, dead volume and loop design prevents the purified water from being re-contaminated before it is dispensed.
What are the key features?
- Type II production into a storage reservoir
- Type I polishing loop fed by the reservoir
- Type II output of 5–30 L/h at 1–15 MΩ·cm
- Type I output of 0.5–2 L/min at 18.2 MΩ·cm
- Grade selection by valve or dispenser
- Monitoring of reservoir and Type I quality
What are the technical specifications?
A typical combined type i and type ii water system configuration offers type II resistivity of 1–15 MΩ·cm, type I resistivity of 18.2 MΩ·cm, type II output of 5–30 L/h, type I flow of 0.5–2 L/min, storage reservoir of 10–100 L and total organic carbon of < 10 ppb for Type I. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Combined Type I and Type II Water System is used in mixed-purity laboratory water supply, analytical and general reagent water, glassware rinsing plus trace analysis, life science and clinical laboratories and multi-user quality control facilities. Across the wider Ultrapure Water Systems class, laboratories also deploy it for trace metal and ion analysis, HPLC, LC-MS and UHPLC mobile phases and molecular biology and PCR.
Which standards and certifications apply?
Relevant standards and methods include ISO 3696, CLSI GP40, USP <643>, USP <645>, IEC 61010-1 and ISO 9001. Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Maintain the system on the manufacturer's schedule: replace pretreatment and polishing cartridges, and the UV lamps, before performance declines. Keep the recirculation loop short and flowing, because stagnant water allows biofilm to form on wetted surfaces and raises total organic carbon. Sanitize on a defined interval and after any contamination event, and verify with resistivity and microbial testing. Handle the point-of-use filter carefully to avoid contamination, and touch only the connector. Draw water just before use and do not store Type I water for long, since its purity degrades in containers. Follow IEC 61010-1 for electrical safety, and document resistivity, total organic carbon and service dates for traceability.
How do I select the right Combined Type I and Type II Water System?
Choose a combined system when both ultrapure and general-grade water are needed daily, since it uses one feed and one control point. For Type I only, a benchtop polishing unit is cheaper. In the RFQ, state both grade volumes, feed and storage needs.
How does it compare with other Ultrapure Water Systems variants?
Start from the analytical method, because it determines the required resistivity, total organic carbon, bacterial and endotoxin limits. Trace metals and ion chromatography need low ionic content, LC-MS needs low organic carbon, and cell culture needs low endotoxin and bacteria. Choose a system with the purification steps that meet those requirements, and confirm that resistivity and total organic carbon are monitored, ideally with a calibration option. Size the unit by daily volume and instantaneous flow, and decide whether it should produce Type I only or both Type I and Type II. Consider the feed source, since reverse osmosis feed extends cartridge life compared with tap water. Check the point-of-use dispenser, dead volume, recirculation, sanitization procedure and consumable cost per liter before selecting.
Full specifications
| Type II resistivity | 1–15 MΩ·cm |
|---|---|
| Type I resistivity | 18.2 MΩ·cm |
| Type II output | 5–30 L/h |
| Type I flow | 0.5–2 L/min |
| Storage reservoir | 10–100 L |
| Total organic carbon | < 10 ppb for Type I |
| UV lamps | 185 nm and 254 nm |
| Monitoring | Reservoir and Type I |
| Power supply | 220–240 V, 50/60 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Mixed-purity laboratory water supply
- Analytical and general reagent water
- Glassware rinsing plus trace analysis
- Life science and clinical laboratories
- Multi-user quality control facilities
Key features
- Type II production into a storage reservoir
- Type I polishing loop fed by the reservoir
- Type II output of 5–30 L/h at 1–15 MΩ·cm
- Type I output of 0.5–2 L/min at 18.2 MΩ·cm
- Grade selection by valve or dispenser
- Monitoring of reservoir and Type I quality
Frequently asked questions about Combined Type I and Type II Water System
Polishing every liter to Type I wastes cartridges and increases cost, and most general work does not need it. Producing Type II in bulk and polishing only what is required is more economical.
It is vented with a filter, shielded from light, recirculated and sanitized on a schedule. Some systems include a UV lamp in the reservoir to limit microbial growth.
Yes, it recirculates continuously or on a timer so water does not stagnate in the loop. Keeping the loop short and using inert materials preserves resistivity and low organic carbon.
Type I water has resistivity near 18.2 megohm-centimeters at 25 degrees Celsius, total organic carbon below about 10 parts per billion, very low bacteria and, where specified, low endotoxin. These limits follow CLSI GP40 and ISO 3696.
The 185 nanometer lamp oxidizes organic molecules, lowering total organic carbon, while the 254 nanometer lamp inactivates microorganisms by damaging their nucleic acids. Together they control organics and microbial growth.
Only briefly, because containers leach ions and organics and allow microbial growth. Draw water fresh at the point of use and keep the system recirculating when idle to maintain quality.
Related models