Skip to main content
Lab Manufacturers China
Two-Dimensional UHPLC System — product photo

Two-Dimensional UHPLC System

二维超高效液相色谱系统

  • SKU UHP-0005
  • UHPLC
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • RoHS (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Maximum operating pressure100 MPa (1000 bar) per dimension
Flow rate range, 1st dimension0.01–2 mL/min
Flow rate range, 2nd dimension0.1–5 mL/min
Second-dimension gradient time0.5–2 min (comprehensive mode)
Valve type2-position, 8-, 10- or 12-port
Sample loop or trap volume20–500 µL
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <621>, Ph. Eur. 2.2.46, USP <1058>

What is the Two-Dimensional UHPLC System?

A two-dimensional UHPLC system couples two columns of different selectivity through a switching valve, transferring selected fractions from the first dimension to a fast second-dimension separation. It is used when a single column cannot resolve a complex sample, for example impurities co-eluting with a main peak or peptides in digests.

What is the Two-Dimensional UHPLC System?

A two-dimensional UHPLC system couples two columns of different selectivity through a switching valve, transferring selected fractions from the first dimension to a fast second-dimension separation. It is used when a single column cannot resolve a complex sample, for example impurities co-eluting with a main peak or peptides in digests.

Overview

The system contains two independent pump modules and column ovens linked by a two-position multi-port valve fitted with sample loops or trapping columns. In heart-cutting mode, one or several peaks of interest are captured from the first dimension and injected into the second, typically a different stationary phase or pH. In comprehensive mode, the entire first-dimension effluent is sampled in short intervals, for example every 30–60 seconds, into ultrafast second-dimension gradients of 30–60 seconds, producing a contour plot. Peak capacity can approach the product of both dimensions, often exceeding 1,000. Software synchronizes valve timing, pump gradients and detector acquisition. Typical uses include peak purity confirmation for regulated methods, desalting of first-dimension buffers before mass spectrometry, and deep profiling of polymers, peptides and natural products. Method development is more complex than single-dimension work.

How does it work?

UHPLC follows the same partition principle as HPLC but exploits the van Deemter relationship: smaller particles reduce eddy diffusion and mass-transfer resistance, so efficiency stays high at higher linear velocities. Columns of 1.7–1.9 µm particles, commonly 2.1 mm internal diameter and 50–150 mm long, generate backpressures of several hundred bar, which requires pumps, valves and injectors rated to 1000 bar or more. Extra-column dispersion must be minimized with narrow-bore tubing, low-volume mixers and flow cells of 1–2 µL, otherwise the narrow peaks, often 1–3 seconds wide, are broadened. Fast detector sampling rates of 40–200 Hz are needed to define peaks, and column ovens with active pre-heating control frictional heating. Mobile phases are filtered to 0.2 µm to protect frits. Runs that took 30 minutes on HPLC can often finish in 3–10 minutes.

What are the key features?

  • Two independent binary UHPLC pump modules with separate column ovens and gradients
  • Two-position multi-port switching valve with sample loops or trapping columns
  • Supports heart-cutting, multiple heart-cutting and comprehensive 2D-LC modes
  • Orthogonal selectivity through different stationary phases, pH or organic modifiers per dimension
  • Integrated software synchronizing valve timing, gradients and detector data into contour plots

What are the technical specifications?

A typical two-dimensional uhplc system configuration offers maximum operating pressure of 100 MPa (1000 bar) per dimension, flow rate range, 1st dimension of 0.01–2 mL/min, flow rate range, 2nd dimension of 0.1–5 mL/min, second-dimension gradient time of 0.5–2 min (comprehensive mode), valve type of 2-position, 8-, 10- or 12-port and sample loop or trap volume of 20–500 µL. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Two-Dimensional UHPLC System is used in peak purity confirmation in regulated methods, impurity profiling co-eluting with the main peak, peptide and protein digest mapping, polymer and natural product characterization and online desalting before mass spectrometry. Across the wider UHPLC class, laboratories also deploy it for high-throughput pharmaceutical impurity profiling, multi-residue pesticide and veterinary drug screening and bioanalytical PK sample analysis.

Which standards and certifications apply?

Relevant standards and methods include USP <621>, Ph. Eur. 2.2.46, USP <1058>, ICH Q2(R2), ChP 0512, IEC 61010-1, ISO/IEC 17025 and 21 CFR Part 11. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Use LC-MS grade solvents and filter samples to 0.2 µm, since particulates quickly block sub-2 µm frits. Replace inlet filters, seals and rotor seals on the manufacturer schedule because pressure cycling wears them faster than in HPLC. Keep tubing from injector to detector short and narrow, and use zero-dead-volume fittings rated for the operating pressure. Qualify the system following USP <1058> and verify detector sampling rate and gradient accuracy. When transferring HPLC methods, follow the allowable adjustments in USP <621>. Observe safety precautions for high-pressure components and never exceed column pressure limits, which are often 1000 bar for UHPLC columns.

How do I select the right Two-Dimensional UHPLC System?

Select a 2D-UHPLC system only when single-dimension optimization cannot resolve critical pairs, or when online desalting and mass spectrometry compatibility is needed. Decide between heart-cutting and comprehensive modes based on whether you need target fractions or a full profile. Specify valve configuration, pump modules, detectors and 2D software features.

How does it compare with other UHPLC variants?

The decisive parameters are pressure rating, dwell and extra-column volume, detector speed and whether methods will be transferred from legacy HPLC. A 600-bar system suffices for 2.7 µm solid-core columns; true sub-2 µm work on 100–150 mm columns calls for 1000–1300 bar. Binary pumps give low dwell volume for fast gradients, while quaternary designs suit screening and pH scouting at modest throughput cost. Micro-flow systems increase mass sensitivity for limited samples and mass spectrometer coupling. Two-dimensional configurations address complex matrices. Also evaluate autosampler cycle time and carryover, because fast runs make injection overhead a large share of total time. Confirm that the software supports method translation with dwell-volume compensation and that data integrity features meet your regulatory requirements.

Full specifications

Full specifications
Maximum operating pressure100 MPa (1000 bar) per dimension
Flow rate range, 1st dimension0.01–2 mL/min
Flow rate range, 2nd dimension0.1–5 mL/min
Second-dimension gradient time0.5–2 min (comprehensive mode)
Valve type2-position, 8-, 10- or 12-port
Sample loop or trap volume20–500 µL
Theoretical peak capacity>1,000 (comprehensive mode)
Column ovens2 independent, 5–90 °C
Detector data rateup to 200 Hz
Modes supportedHeart-cutting, multiple heart-cutting, comprehensive

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Peak purity confirmation in regulated methods
  • Impurity profiling co-eluting with the main peak
  • Peptide and protein digest mapping
  • Polymer and natural product characterization
  • Online desalting before mass spectrometry

Key features

  • Two independent binary UHPLC pump modules with separate column ovens and gradients
  • Two-position multi-port switching valve with sample loops or trapping columns
  • Supports heart-cutting, multiple heart-cutting and comprehensive 2D-LC modes
  • Orthogonal selectivity through different stationary phases, pH or organic modifiers per dimension
  • Integrated software synchronizing valve timing, gradients and detector data into contour plots

Frequently asked questions about Two-Dimensional UHPLC System

Heart-cutting transfers only selected first-dimension fractions to the second column, which is simpler and quantitative for target peaks. Comprehensive 2D-LC samples the whole first-dimension effluent continuously, giving a full two-dimensional map at the cost of higher method complexity and dilution.

Related models

All UHPLC
Binary UHPLC System — product photo

Binary UHPLC System

二元超高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote
Quaternary UHPLC System — product photo

Quaternary UHPLC System

四元超高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote
Micro-Flow UHPLC System — product photo

Micro-Flow UHPLC System

微流量超高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote
Capillary GC-FID System — product photo

Capillary GC-FID System

毛细管气相色谱 FID 系统

  • CE (to be confirmed)
  • +2
Request a quote

Request a quotation for the Two-Dimensional UHPLC System

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote