Isocratic HPLC System

等度高效液相色谱系统

  • SKU HPL-0001
  • HPLC
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • RoHS (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Flow rate range0.01–10 mL/min
Flow accuracy±1%
Flow precision (RSD)≤0.1%
Maximum operating pressure40 MPa (400 bar)
Solvent channels1
Column oven temperature range5 °C above ambient to 80 °C
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <621>, Ph. Eur. 2.2.29, USP <1058>

What is the Isocratic HPLC System?

An isocratic HPLC system delivers a single premixed mobile phase at constant composition through a single-piston or dual-piston pump. It is the right choice for routine, validated assays of simple mixtures, where low cost, fast startup and minimal training matter more than gradient flexibility.

What is the Isocratic HPLC System?

An isocratic HPLC system delivers a single premixed mobile phase at constant composition through a single-piston or dual-piston pump. It is the right choice for routine, validated assays of simple mixtures, where low cost, fast startup and minimal training matter more than gradient flexibility.

Overview

The isocratic configuration uses one solvent line, a single pump head or a dual-piston serial pump, a manual or automated injector, a column compartment and a UV-Vis detector. Because there is no solvent proportioning, baseline stability is high and dwell volume is negligible, which makes it forgiving for refractive index detection and for transferring legacy pharmacopoeial methods written for isocratic operation. The typical flow range is 0.01–10 mL/min with working pressure up to about 400 bar. Method scope is limited to analyses where all peaks elute within a reasonable run time under fixed solvent strength, such as assay of active ingredients, preservatives in cosmetics or sugar profiling with a refractive index detector. Upgrade paths are limited, so users expecting method development or impurity profiling should consider a gradient model instead.

How does it work?

An HPLC system pumps solvent through an injector and a packed column, typically 50–250 mm long with 3–5 µm particles. The sample is injected as a narrow band and analytes partition between the mobile phase and the stationary phase, most commonly bonded C18 silica in reversed-phase mode. Retention depends on polarity, pH, temperature and solvent strength. Isocratic operation keeps composition constant; gradient operation changes it over time to elute compounds with widely different retention. Key modules are the solvent delivery pump with degasser, the autosampler, the thermostatted column compartment and the detector, commonly UV-Vis, diode-array, fluorescence or refractive index. Chromatography data software integrates peaks, applies calibration curves and reports retention time, area, plate count and tailing factor for system suitability.

What are the key features?

  • Single solvent line with low dwell volume and stable baseline suitable for refractive index detection
  • Dual-piston serial pump with flow range of 0.01–10 mL/min and pulsation damping
  • Thermostatted column compartment holding 5–80 °C for reproducible retention
  • Variable-wavelength UV-Vis detector with deuterium lamp and low-volume flow cell
  • Simple touch-screen or software control with short startup and equilibration time

What are the technical specifications?

A typical isocratic hplc system configuration offers flow rate range of 0.01–10 mL/min, flow accuracy of ±1%, flow precision (RSD) of ≤0.1%, maximum operating pressure of 40 MPa (400 bar), solvent channels of 1 and column oven temperature range of 5 °C above ambient to 80 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Isocratic HPLC System is used in routine assay of active ingredients, preservative testing in cosmetics, sugar analysis with refractive index detection and teaching laboratories. Across the wider HPLC class, laboratories also deploy it for assay and purity testing of drug substances, food additive and vitamin analysis and environmental pollutant monitoring.

Which standards and certifications apply?

Relevant standards and methods include USP <621>, Ph. Eur. 2.2.29, USP <1058>, ICH Q2(R2), ChP 0512, IEC 61010-1, ISO/IEC 17025 and 21 CFR Part 11. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Install the system on a stable bench away from drafts, with a steady power supply and ventilated solvent waste storage. Degas and filter mobile phases through 0.22–0.45 µm membranes to protect check valves and columns. Qualify the instrument (IQ/OQ/PQ) following USP <1058>, verifying flow accuracy, gradient composition accuracy, injection precision, detector linearity, wavelength accuracy and column oven temperature. Establish preventive maintenance for pump seals, piston washes, lamps and autosampler needle seats. Flush buffers with water before storing, and follow column manufacturer pH and pressure limits. For regulated work, configure audit trails and user roles in the data system.

How do I select the right Isocratic HPLC System?

Select the isocratic system when all your methods use a fixed mobile phase and you need a low-cost, easily validated platform. Choose a gradient system instead if you need impurity profiling, method development or solvent switching. In an RFQ specify flow range, maximum pressure, detector type, column oven range, injector type and required software compliance features.

How does it compare with other HPLC variants?

Within a chromatography laboratory, HPLC configurations differ mainly in pump design, detector options and scale. Isocratic systems suit single-solvent routine assays at the lowest cost. Binary gradient systems cover most method development and impurity work, while quaternary systems add on-line solvent blending for screening and pH or modifier adjustment. Preparative systems are chosen when the goal is to isolate milligrams to grams of a compound, and bio-inert systems are needed when mobile phases contain high salt or extreme pH, or when proteins must not contact stainless steel. Decide on maximum pressure, flow range, dwell volume, autosampler capacity, detector types, column oven range and data-integrity features. Check that column dimensions in your validated methods are supported without extensive re-plumbing.

Full specifications

Full specifications
Flow rate range0.01–10 mL/min
Flow accuracy±1%
Flow precision (RSD)≤0.1%
Maximum operating pressure40 MPa (400 bar)
Solvent channels1
Column oven temperature range5 °C above ambient to 80 °C
Detector wavelength range190–700 nm
Detector noise≤±0.5 × 10⁻⁵ AU
Injection volume1–100 µL
Injection carryover≤0.05%

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Routine assay of active ingredients
  • Preservative testing in cosmetics
  • Sugar analysis with refractive index detection
  • Teaching laboratories

Key features

  • Single solvent line with low dwell volume and stable baseline suitable for refractive index detection
  • Dual-piston serial pump with flow range of 0.01–10 mL/min and pulsation damping
  • Thermostatted column compartment holding 5–80 °C for reproducible retention
  • Variable-wavelength UV-Vis detector with deuterium lamp and low-volume flow cell
  • Simple touch-screen or software control with short startup and equilibration time

Frequently asked questions about Isocratic HPLC System

No. A single-solvent-line isocratic pump cannot proportion solvents, so gradient methods are not possible. Some users emulate a step change by manually switching reservoirs, but this is not reproducible or suitable for validated work. Use a binary or quaternary gradient system for any method that specifies a gradient program.

Related models

All HPLC
Preparative HPLC System — product photo

Preparative HPLC System

制备型高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote
Bio-Inert HPLC System — product photo

Bio-Inert HPLC System

生物惰性高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote
Quaternary UHPLC System — product photo

Quaternary UHPLC System

四元超高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote

Request a quotation for the Isocratic HPLC System

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote