Quaternary UHPLC System
四元超高效液相色谱系统
- SKU UHP-0002
- UHPLC
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Maximum operating pressure | 60–80 MPa (600–800 bar) |
|---|---|
| Flow rate range | 0.05–5 mL/min |
| Solvent channels | 4 |
| Dwell volume | 350–600 µL |
| Gradient composition accuracy | ±0.5% absolute |
| Flow precision (RSD) | ≤0.1% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <621>, Ph. Eur. 2.2.46, USP <1058>
What is the Quaternary UHPLC System?
A quaternary UHPLC system blends up to four solvents at low pressure before a single pump rated to about 600–800 bar. It is chosen by method-development and research laboratories that want automated solvent and pH scouting with sub-2 µm or core-shell columns, accepting slightly higher dwell volume than binary designs.
What is the Quaternary UHPLC System?
A quaternary UHPLC system blends up to four solvents at low pressure before a single pump rated to about 600–800 bar. It is chosen by method-development and research laboratories that want automated solvent and pH scouting with sub-2 µm or core-shell columns, accepting slightly higher dwell volume than binary designs.
Overview
This variant pairs a four-channel low-pressure proportioning valve and in-line degasser with a pump engineered for UHPLC-range backpressure, commonly 600–800 bar. Dwell volume is typically 350–600 µL, larger than binary systems, so very steep gradients show a delay, but modern methods can be transferred using software dwell-volume compensation. The configuration supports automated screening of columns through a multi-column valve, buffers, pH and organic modifiers without manual bottle changes, which shortens method development cycles. Typical flow is 0.05–5 mL/min and column compartments accommodate 2.1–4.6 mm columns up to 150 mm. It is well suited to 2.7 µm core-shell columns and to laboratories that run both HPLC-style and UHPLC-style methods on one platform. Users running very fast 1–2 minute gradients on 1.7 µm particles will generally be better served by a binary design.
How does it work?
UHPLC follows the same partition principle as HPLC but exploits the van Deemter relationship: smaller particles reduce eddy diffusion and mass-transfer resistance, so efficiency stays high at higher linear velocities. Columns of 1.7–1.9 µm particles, commonly 2.1 mm internal diameter and 50–150 mm long, generate backpressures of several hundred bar, which requires pumps, valves and injectors rated to 1000 bar or more. Extra-column dispersion must be minimized with narrow-bore tubing, low-volume mixers and flow cells of 1–2 µL, otherwise the narrow peaks, often 1–3 seconds wide, are broadened. Fast detector sampling rates of 40–200 Hz are needed to define peaks, and column ovens with active pre-heating control frictional heating. Mobile phases are filtered to 0.2 µm to protect frits. Runs that took 30 minutes on HPLC can often finish in 3–10 minutes.
What are the key features?
- Four-channel low-pressure proportioning with in-line vacuum degasser for automated solvent blending
- Pressure rating of 600–800 bar supporting 2.7 µm core-shell and many 1.8 µm columns
- Software dwell-volume compensation enabling method transfer to and from binary systems
- Multi-column selector valve enabling unattended screening of up to six columns
- Automated pH and buffer scouting sequences reduce manual mobile phase preparation
What are the technical specifications?
A typical quaternary uhplc system configuration offers maximum operating pressure of 60–80 MPa (600–800 bar), flow rate range of 0.05–5 mL/min, solvent channels of 4, dwell volume of 350–600 µL, gradient composition accuracy of ±0.5% absolute and flow precision (RSD) of ≤0.1%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Quaternary UHPLC System is used in automated method development screening, pH and buffer scouting, mixed HPLC and UHPLC method portfolios and academic and CRO research laboratories. Across the wider UHPLC class, laboratories also deploy it for high-throughput pharmaceutical impurity profiling, multi-residue pesticide and veterinary drug screening and bioanalytical PK sample analysis.
Which standards and certifications apply?
Relevant standards and methods include USP <621>, Ph. Eur. 2.2.46, USP <1058>, ICH Q2(R2), ChP 0512, IEC 61010-1, ISO/IEC 17025 and 21 CFR Part 11. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Use LC-MS grade solvents and filter samples to 0.2 µm, since particulates quickly block sub-2 µm frits. Replace inlet filters, seals and rotor seals on the manufacturer schedule because pressure cycling wears them faster than in HPLC. Keep tubing from injector to detector short and narrow, and use zero-dead-volume fittings rated for the operating pressure. Qualify the system following USP <1058> and verify detector sampling rate and gradient accuracy. When transferring HPLC methods, follow the allowable adjustments in USP <621>. Observe safety precautions for high-pressure components and never exceed column pressure limits, which are often 1000 bar for UHPLC columns.
How do I select the right Quaternary UHPLC System?
Select the quaternary UHPLC when flexibility in solvent blending matters more than absolute minimum dwell volume. Choose binary for maximum speed with 1.7 µm particles, or the 1300-bar model for extremely long columns. In an RFQ specify pressure rating, channel count, column selector needs, dwell-volume compensation software and detector data rate.
How does it compare with other UHPLC variants?
The decisive parameters are pressure rating, dwell and extra-column volume, detector speed and whether methods will be transferred from legacy HPLC. A 600-bar system suffices for 2.7 µm solid-core columns; true sub-2 µm work on 100–150 mm columns calls for 1000–1300 bar. Binary pumps give low dwell volume for fast gradients, while quaternary designs suit screening and pH scouting at modest throughput cost. Micro-flow systems increase mass sensitivity for limited samples and mass spectrometer coupling. Two-dimensional configurations address complex matrices. Also evaluate autosampler cycle time and carryover, because fast runs make injection overhead a large share of total time. Confirm that the software supports method translation with dwell-volume compensation and that data integrity features meet your regulatory requirements.
Full specifications
| Maximum operating pressure | 60–80 MPa (600–800 bar) |
|---|---|
| Flow rate range | 0.05–5 mL/min |
| Solvent channels | 4 |
| Dwell volume | 350–600 µL |
| Gradient composition accuracy | ±0.5% absolute |
| Flow precision (RSD) | ≤0.1% |
| Injection carryover | ≤0.01% |
| Column selector capacity | up to 6 columns |
| Detector data rate | up to 80 Hz |
| Column oven range | 5–90 °C |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Automated method development screening
- pH and buffer scouting
- Mixed HPLC and UHPLC method portfolios
- Academic and CRO research laboratories
Key features
- Four-channel low-pressure proportioning with in-line vacuum degasser for automated solvent blending
- Pressure rating of 600–800 bar supporting 2.7 µm core-shell and many 1.8 µm columns
- Software dwell-volume compensation enabling method transfer to and from binary systems
- Multi-column selector valve enabling unattended screening of up to six columns
- Automated pH and buffer scouting sequences reduce manual mobile phase preparation
Frequently asked questions about Quaternary UHPLC System
Yes, if its pressure rating covers the method. Short 1.7 µm columns at moderate flow often stay under 600 bar, so a 600–800 bar quaternary system can run them. Longer columns or high-viscosity solvents may exceed the limit, so check backpressure before selecting.
Software delays the injection or starts the gradient early by an amount equal to the volume difference between the source and target instrument, typically stated in microliters. This reproduces the effective gradient profile at the column inlet and restores retention time and selectivity when methods are transferred.
Yes, provided you use volatile buffers and LC-MS grade solvents. The four-channel blending is useful for screening additives such as formic acid, ammonium formate and ammonium acetate. For very fast gradients that feed a mass spectrometer, a binary pump reduces delay and improves reproducibility.
UHPLC uses sub-2 µm particles and hardware rated above 600 bar, typically 1000–1300 bar, which gives narrower peaks, higher resolution and runs three to ten times faster than conventional HPLC. It also requires lower system dispersion and faster detector sampling.
Usually yes. Keep the stationary phase chemistry, scale column length to particle size ratio, adjust flow rate and gradient time proportionally, and compensate for dwell volume differences. USP <621> defines permitted adjustments, but changes beyond those limits require revalidation.
Generally yes. Shorter columns of narrower internal diameter and faster runs typically reduce solvent use per sample by 60–90% compared with a 4.6 × 250 mm HPLC method. The saving depends on whether the method is fully redeveloped on a 2.1 mm column.