GC-TCD System for Permanent Gases
永久性气体分析气相色谱 TCD 系统
- SKU GAC-0003
- Gas Chromatography
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Detector type | Dual-filament TCD |
|---|---|
| Sensitivity | ≥800 mV·mL/mg (propane in He) |
| Minimum detectable concentration | ≤400 pg/mL (propane) |
| Linear dynamic range | ≥10⁵ |
| Detector temperature | up to 400 °C |
| Gas sampling valve loop volume | 0.25–1 mL |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <621>, USP <467>, Ph. Eur. 2.2.28
What is the GC-TCD System for Permanent Gases?
A GC-TCD system uses a thermal conductivity detector that responds to any compound whose thermal conductivity differs from the carrier gas. It is the standard configuration for permanent gases such as hydrogen, oxygen, nitrogen, CO and CO₂, and for percent-level gas mixtures in industrial, energy and laboratory gas analysis.
What is the GC-TCD System for Permanent Gases?
A GC-TCD system uses a thermal conductivity detector that responds to any compound whose thermal conductivity differs from the carrier gas. It is the standard configuration for permanent gases such as hydrogen, oxygen, nitrogen, CO and CO₂, and for percent-level gas mixtures in industrial, energy and laboratory gas analysis.
Overview
The thermal conductivity detector compares heat loss from a heated filament in the column effluent with a reference flow of pure carrier gas, in a Wheatstone bridge circuit. It is non-destructive and universal, with a detectability of roughly 400 pg/mL for propane and a linear range of about 10⁵. Permanent-gas analysis typically uses packed columns or porous-layer open tubular columns with molecular sieve 5A and Porapak or HayeSep phases, often on a valved multi-column system with column switching and backflush so that CO₂ is kept off the molecular sieve. Helium is the usual carrier for most gases, while argon or nitrogen is used when hydrogen is the analyte. Gas sampling valves with 0.25–1 mL loops introduce gases reproducibly. Typical applications include biogas, natural gas, syngas, fuel cell gases and headspace oxygen, where percent and ppm-level quantitation is needed.
How does it work?
A gas chromatograph injects a liquid or gas sample into a heated inlet, commonly split/splitless at 150–350 °C, where it vaporizes and mixes with carrier gas such as helium, hydrogen or nitrogen. The vapor passes through a capillary column, typically 15–60 m long, 0.25 mm internal diameter, coated with a polysiloxane or polyethylene glycol film of 0.1–1.0 µm. Analytes partition between gas and film according to volatility and polarity. A programmable oven ramps temperature, often from 40 to 300 °C, eluting compounds in boiling-point order. Detectors respond differently: the flame ionization detector senses carbon-containing compounds, the thermal conductivity detector responds to all gases, the electron capture detector to electronegative halogens, and mass spectrometers give identification. Electronic pneumatic control regulates pressure and flow to keep retention times reproducible.
What are the key features?
- Dual-filament thermal conductivity detector with heated cell to 400 °C and reference flow
- Gas sampling valve with 0.25–1 mL sample loops for reproducible gas injection
- Multi-valve, multi-column configuration with backflush for molecular sieve and porous polymer columns
- Universal response for H₂, O₂, N₂, CO, CO₂, CH₄ and light hydrocarbons in one run
- Compatible with helium, argon or nitrogen carrier for hydrogen analysis
What are the technical specifications?
A typical gc-tcd system for permanent gases configuration offers detector type of Dual-filament TCD, sensitivity of ≥800 mV·mL/mg (propane in He), minimum detectable concentration of ≤400 pg/mL (propane), linear dynamic range of ≥10⁵, detector temperature of up to 400 °C and gas sampling valve loop volume of 0.25–1 mL. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The GC-TCD System for Permanent Gases is used in biogas and natural gas composition, syngas and hydrogen purity analysis, fuel cell and battery gas analysis, modified atmosphere packaging headspace oxygen and industrial gas purity control. Across the wider Gas Chromatography class, laboratories also deploy it for residual solvent testing in pharmaceuticals, petrochemical hydrocarbon and purity analysis and environmental volatile organic compounds.
Which standards and certifications apply?
Relevant standards and methods include USP <621>, USP <467>, Ph. Eur. 2.2.28, ASTM D6729, ASTM D5134, US EPA Method 8015, ISO 6974, ChP 0521 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Provide high-purity carrier and detector gases, at least 99.999%, with moisture, oxygen and hydrocarbon traps, and install leak-tight fittings since oxygen damages columns at high temperature. If hydrogen is used, install a hydrogen sensor and follow flammable gas safety requirements. Condition new columns following the manufacturer's procedure and respect maximum temperature limits. Replace septa, inlet liners and O-rings regularly, and trim the column inlet to restore peak shape. Verify oven temperature, flow and detector sensitivity during qualification under USP <1058>. Ventilate exhaust, particularly when analyzing toxic solvents, and keep a stable laboratory temperature for electronic pneumatics.
How do I select the right GC-TCD System for Permanent Gases?
Select GC-TCD when analyzing permanent gases or percent-level components that FID cannot detect. For ppm-level hydrocarbons in the same sample, combine TCD with FID or add a methanizer. In an RFQ specify target gases and ranges, valve and column configuration, carrier gas and sampling arrangement.
How does it compare with other Gas Chromatography variants?
Choose the detector first, because it defines what you can measure. FID covers hydrocarbons, solvents and most organic compounds at ppm to ppb levels; ECD is selective for halogenated pesticides and PCBs at ppt levels; TCD is universal and suits permanent gases and percent-level analysis. Headspace sampling is selected for volatile analytes in solid or liquid matrices without injecting the matrix. Portable GC suits field screening. Then consider inlet type, number of detectors, oven ramp rate, electronic pneumatic control precision, autosampler capacity, and carrier gas availability, since hydrogen generators can replace helium. Confirm software support for regulatory reporting, and compare column compatibility with your validated methods, including ASTM, EPA or pharmacopoeial procedures that specify particular phases and dimensions.
Full specifications
| Detector type | Dual-filament TCD |
|---|---|
| Sensitivity | ≥800 mV·mL/mg (propane in He) |
| Minimum detectable concentration | ≤400 pg/mL (propane) |
| Linear dynamic range | ≥10⁵ |
| Detector temperature | up to 400 °C |
| Gas sampling valve loop volume | 0.25–1 mL |
| Oven temperature range | Ambient +4 °C to 400 °C |
| Typical columns | Molecular sieve 5A, Porapak Q, HayeSep |
| Retention time repeatability (RSD) | ≤0.05% |
| Peak area repeatability (RSD) | ≤1.0% |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Biogas and natural gas composition
- Syngas and hydrogen purity analysis
- Fuel cell and battery gas analysis
- Modified atmosphere packaging headspace oxygen
- Industrial gas purity control
Key features
- Dual-filament thermal conductivity detector with heated cell to 400 °C and reference flow
- Gas sampling valve with 0.25–1 mL sample loops for reproducible gas injection
- Multi-valve, multi-column configuration with backflush for molecular sieve and porous polymer columns
- Universal response for H₂, O₂, N₂, CO, CO₂, CH₄ and light hydrocarbons in one run
- Compatible with helium, argon or nitrogen carrier for hydrogen analysis
Frequently asked questions about GC-TCD System for Permanent Gases
Hydrogen and helium have similar thermal conductivities, so with helium carrier the hydrogen peak is tiny or shows inverted polarity. Argon or nitrogen has much lower thermal conductivity than hydrogen, which gives a large, positive hydrogen peak and good sensitivity.
Backflush reverses carrier flow through a pre-column to vent heavier components, such as C3 and above or moisture, before they reach the main column. It shortens analysis time and protects molecular sieve columns, which are permanently poisoned by CO₂ and water.
No. The thermal conductivity detector does not consume or alter the analytes, so the effluent can pass to a second detector in series, for example FID or a mass spectrometer, or be collected after the detector.
Gas chromatography analyzes compounds that can be vaporized without decomposing, generally with boiling points below about 400 °C and molecular weights under roughly 500 Da. This includes hydrocarbons, solvents, pesticides, fatty acid esters and permanent gases. Polar or non-volatile compounds may need derivatization.
Helium is inert and widely specified in methods, but supply is variable. Hydrogen gives faster separations at optimum linear velocity and can be produced by a generator, but requires safety measures. Nitrogen is cheapest and suits some detectors but gives slower analysis and lower efficiency at high flows.
Split injection vents most of the vaporized sample, typically at ratios of 10:1 to 100:1, to prevent column overload with concentrated samples. Splitless injection transfers nearly the entire sample to the column for trace analysis, but requires solvent focusing and a longer inlet residence time.