Skip to main content
Lab Manufacturers China
Capillary GC-FID System — product photo

Capillary GC-FID System

毛细管气相色谱 FID 系统

  • SKU GAC-0001
  • Gas Chromatography
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • RoHS (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Oven temperature rangeAmbient +4 °C to 450 °C
Maximum oven ramp rateup to 100 °C/min
Inlet temperature limit400 °C
Split ratio range1:1 to 1000:1
FID minimum detectable level≤1.5 pg C/s
FID linear dynamic range≥10⁷
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <621>, USP <467>, Ph. Eur. 2.2.28

What is the Capillary GC-FID System?

A capillary GC-FID system pairs a split/splitless inlet and temperature-programmed oven with a flame ionization detector. It is the general-purpose configuration for quantifying hydrocarbons, solvents, fatty acid methyl esters and most organic compounds, offering a wide linear range and stable response for routine quality control.

What is the Capillary GC-FID System?

A capillary GC-FID system pairs a split/splitless inlet and temperature-programmed oven with a flame ionization detector. It is the general-purpose configuration for quantifying hydrocarbons, solvents, fatty acid methyl esters and most organic compounds, offering a wide linear range and stable response for routine quality control.

Overview

The flame ionization detector burns column effluent in a hydrogen and air flame, where carbon-containing compounds produce ions collected as a current proportional to the mass of carbon entering per second. It offers a linear dynamic range near 10⁷ and minimum detectable levels around 1–3 pg C/s, with low maintenance and tolerance for moisture. The system normally includes an electronic pneumatic control module, a split/splitless inlet up to 400 °C, an oven ramping at up to 100 °C/min, a liquid autosampler with 50–150 positions, and optionally a second inlet and detector for dual-column confirmation. FID does not respond to permanent gases, water or CS₂ well, and gives no structural information, so peak identity relies on retention time or a second column. It remains the workhorse for ASTM, USP and ISO methods that specify FID for hydrocarbons and residual solvents.

How does it work?

A gas chromatograph injects a liquid or gas sample into a heated inlet, commonly split/splitless at 150–350 °C, where it vaporizes and mixes with carrier gas such as helium, hydrogen or nitrogen. The vapor passes through a capillary column, typically 15–60 m long, 0.25 mm internal diameter, coated with a polysiloxane or polyethylene glycol film of 0.1–1.0 µm. Analytes partition between gas and film according to volatility and polarity. A programmable oven ramps temperature, often from 40 to 300 °C, eluting compounds in boiling-point order. Detectors respond differently: the flame ionization detector senses carbon-containing compounds, the thermal conductivity detector responds to all gases, the electron capture detector to electronegative halogens, and mass spectrometers give identification. Electronic pneumatic control regulates pressure and flow to keep retention times reproducible.

What are the key features?

  • Flame ionization detector with linear dynamic range near 10⁷ and MDL of about 1–3 pg C/s
  • Split/splitless inlet up to 400 °C with electronic pneumatic control of pressure and flow
  • Oven temperature ramps up to 100 °C/min with fast cool-down for high throughput
  • Liquid autosampler with 50–150 vials and dual-tower option for two injections in parallel
  • Optional second inlet and detector for dual-column confirmation of peak identity

What are the technical specifications?

A typical capillary gc-fid system configuration offers oven temperature range of Ambient +4 °C to 450 °C, maximum oven ramp rate of up to 100 °C/min, inlet temperature limit of 400 °C, split ratio range of 1:1 to 1000:1, FID minimum detectable level of ≤1.5 pg C/s and FID linear dynamic range of ≥10⁷. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Capillary GC-FID System is used in residual solvents in pharmaceuticals, fatty acid methyl ester analysis, petrochemical hydrocarbon purity, alcohol and beverage analysis and environmental hydrocarbons and BTEX. Across the wider Gas Chromatography class, laboratories also deploy it for residual solvent testing in pharmaceuticals, petrochemical hydrocarbon and purity analysis and environmental volatile organic compounds.

Which standards and certifications apply?

Relevant standards and methods include USP <621>, USP <467>, Ph. Eur. 2.2.28, ASTM D6729, ASTM D5134, US EPA Method 8015, ISO 6974, ChP 0521 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Provide high-purity carrier and detector gases, at least 99.999%, with moisture, oxygen and hydrocarbon traps, and install leak-tight fittings since oxygen damages columns at high temperature. If hydrogen is used, install a hydrogen sensor and follow flammable gas safety requirements. Condition new columns following the manufacturer's procedure and respect maximum temperature limits. Replace septa, inlet liners and O-rings regularly, and trim the column inlet to restore peak shape. Verify oven temperature, flow and detector sensitivity during qualification under USP <1058>. Ventilate exhaust, particularly when analyzing toxic solvents, and keep a stable laboratory temperature for electronic pneumatics.

How do I select the right Capillary GC-FID System?

Select GC-FID for routine quantitation of organic compounds where identity is established by retention time. Add ECD or TCD modules if halogenated or permanent-gas analytes are present, or move to GC-MS for unknown identification. In an RFQ specify inlet types, oven range, number of detectors, gas supply, autosampler capacity and methods that must be reproduced.

How does it compare with other Gas Chromatography variants?

Choose the detector first, because it defines what you can measure. FID covers hydrocarbons, solvents and most organic compounds at ppm to ppb levels; ECD is selective for halogenated pesticides and PCBs at ppt levels; TCD is universal and suits permanent gases and percent-level analysis. Headspace sampling is selected for volatile analytes in solid or liquid matrices without injecting the matrix. Portable GC suits field screening. Then consider inlet type, number of detectors, oven ramp rate, electronic pneumatic control precision, autosampler capacity, and carrier gas availability, since hydrogen generators can replace helium. Confirm software support for regulatory reporting, and compare column compatibility with your validated methods, including ASTM, EPA or pharmacopoeial procedures that specify particular phases and dimensions.

Full specifications

Full specifications
Oven temperature rangeAmbient +4 °C to 450 °C
Maximum oven ramp rateup to 100 °C/min
Inlet temperature limit400 °C
Split ratio range1:1 to 1000:1
FID minimum detectable level≤1.5 pg C/s
FID linear dynamic range≥10⁷
Retention time repeatability (RSD)≤0.008%
Peak area repeatability (RSD)≤1.0%
Pneumatic pressure control0–150 psi, 0.001 psi resolution
Autosampler capacity50–150 vials

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Residual solvents in pharmaceuticals
  • Fatty acid methyl ester analysis
  • Petrochemical hydrocarbon purity
  • Alcohol and beverage analysis
  • Environmental hydrocarbons and BTEX

Key features

  • Flame ionization detector with linear dynamic range near 10⁷ and MDL of about 1–3 pg C/s
  • Split/splitless inlet up to 400 °C with electronic pneumatic control of pressure and flow
  • Oven temperature ramps up to 100 °C/min with fast cool-down for high throughput
  • Liquid autosampler with 50–150 vials and dual-tower option for two injections in parallel
  • Optional second inlet and detector for dual-column confirmation of peak identity

Frequently asked questions about Capillary GC-FID System

An FID needs a carrier gas (helium, hydrogen or nitrogen), hydrogen fuel at roughly 30–40 mL/min, zero air at 300–400 mL/min, and a make-up gas, usually nitrogen, at about 25–30 mL/min. All should be at least 99.999% pure with hydrocarbon traps to minimize baseline noise.

Quaternary UHPLC System — product photo

Quaternary UHPLC System

四元超高效液相色谱系统

  • CE (to be confirmed)
  • +2
Request a quote

Request a quotation for the Capillary GC-FID System

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote