LC-MS/MS Toxicology Analyzer
LC-MS/MS 毒理学分析仪
- SKU TOX-0001
- Toxicology Analyzers
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Mass analyzer | Triple quadrupole |
|---|---|
| Mass range | 5–2000 m/z |
| LOQ for typical drugs | 0.1–5 ng/mL |
| Panel size | 100+ compounds |
| Run time per sample | 5–12 min |
| Ionization | ESI, polarity switching |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ANSI/ASB Standard 036, ISO 15189, ISO/IEC 17025
What is the LC-MS/MS Toxicology Analyzer?
An LC-MS/MS toxicology analyzer quantifies broad panels of drugs and metabolites in blood, urine and tissue with high sensitivity. It is the main confirmatory platform in clinical and forensic toxicology laboratories. It supports panels of many compounds.
What is the LC-MS/MS Toxicology Analyzer?
An LC-MS/MS toxicology analyzer quantifies broad panels of drugs and metabolites in blood, urine and tissue with high sensitivity. It is the main confirmatory platform in clinical and forensic toxicology laboratories. It supports panels of many compounds.
Overview
The platform couples a UHPLC with a triple quadrupole mass spectrometer and often an online sample preparation module such as turbulent flow or automated solid-phase extraction. Compounds are separated on reversed-phase columns, ionized by electrospray, and monitored by multiple reaction monitoring with two transitions per analyte. Isotope-labeled internal standards compensate for matrix effects. Panels covering 100 or more compounds, including opioids, benzodiazepines, antidepressants, stimulants and synthetic cannabinoids, can be run in a single method. Polarity switching and scheduled MRM maintain sensitivity. Limits of quantitation in the low ng/mL to pg/mL range are typical. The instrument requires skilled operation, regular source cleaning, and careful assessment of ion suppression. Some vendors supply validated kits and software for toxicology reporting and flagging.
How does it work?
Toxicology analyzers use several principles. Immunoassay analyzers detect drug classes through antibody binding and photometric or chemiluminescent signals, providing fast presumptive results. Gas chromatography with headspace sampling separates volatile compounds such as ethanol and detects them by flame ionization. GC-MS and LC-MS/MS separate compounds and identify them by mass spectra or specific ion transitions, giving confirmatory quantitation with isotope-labeled internal standards. Inductively coupled plasma mass spectrometry ionizes samples in argon plasma to measure metals at ng/L levels. Sample preparation steps such as protein precipitation, hydrolysis, liquid-liquid or solid-phase extraction remove interferences and improve sensitivity. Calibration, quality controls and proficiency testing are required to report results that stand up in clinical and legal settings.
What are the key features?
- Broad panels of 100 or more compounds in one method
- Online sample preparation option for automation
- Scheduled MRM for sensitivity across large panels
- Isotope-labeled internal standard correction of matrix effects
- Toxicology reporting with flagging and review workflows
What are the technical specifications?
A typical lc-ms/ms toxicology analyzer configuration offers mass analyzer of Triple quadrupole, mass range of 5–2000 m/z, LOQ for typical drugs of 0.1–5 ng/mL, panel size of 100+ compounds, run time per sample of 5–12 min and ionization of ESI, polarity switching. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The LC-MS/MS Toxicology Analyzer is used in comprehensive drug screening and confirmation, post-mortem toxicology, drug-facilitated crime casework and therapeutic drug monitoring. Across the wider Toxicology Analyzers class, laboratories also deploy it for clinical toxicology and poisoning cases, forensic post-mortem toxicology and workplace and driving-related testing.
Which standards and certifications apply?
Relevant standards and methods include ANSI/ASB Standard 036, ISO 15189, ISO/IEC 17025, CLSI C62 and CLSI EP15. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Validate each method for selectivity, accuracy, precision, carryover, matrix effects and stability, following recognized forensic or clinical toxicology guidance. Use certified reference materials, matrix-matched calibrators and independent quality controls, and participate in proficiency testing. Control carryover with blanks after high samples. Install instruments with stable power, gas purity, temperature and ventilation appropriate to solvents and acids used. Follow biosafety practices when handling blood and urine. Maintain sample retention and chain-of-custody records, and secure data with audit trails. Perform preventive maintenance on sources, columns, pumps and sample introduction systems, and document measurement uncertainty for quantitative results.
How do I select the right LC-MS/MS Toxicology Analyzer?
Choose LC-MS/MS when broad, sensitive confirmatory quantitation is needed. For alcohol only, headspace GC is simpler. In an RFQ, specify panel size, LOQ targets, sample volumes, automation level, software validation and service coverage. Also confirm software validation support.
How does it compare with other Toxicology Analyzers variants?
Choose by the question being asked and the required evidence level. Immunoassay analyzers give rapid presumptive results for common drug classes and suit high-volume screening. Headspace GC is the dedicated method for blood alcohol and volatile substances. LC-MS/MS is chosen for broad, sensitive confirmatory panels including therapeutic drugs and new substances, while GC-MS remains strong for volatile and derivatizable compounds. ICP-MS is required when metals and metalloids are the targets. Consider sample volume, matrix, turnaround time, automation level and staff expertise. Check calibration and QC software, LIS connectivity and the availability of validated assay kits. For forensic use, confirm data integrity features, chain-of-custody support and method validation documentation.
Full specifications
| Mass analyzer | Triple quadrupole |
|---|---|
| Mass range | 5–2000 m/z |
| LOQ for typical drugs | 0.1–5 ng/mL |
| Panel size | 100+ compounds |
| Run time per sample | 5–12 min |
| Ionization | ESI, polarity switching |
| Autosampler capacity | ≥ 200 vials |
| Sample volume | 50–200 µL |
| Linear dynamic range | ≥ 10^4 |
| Power supply | 220 V, 50/60 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Comprehensive drug screening and confirmation
- Post-mortem toxicology
- Drug-facilitated crime casework
- Therapeutic drug monitoring
Key features
- Broad panels of 100 or more compounds in one method
- Online sample preparation option for automation
- Scheduled MRM for sensitivity across large panels
- Isotope-labeled internal standard correction of matrix effects
- Toxicology reporting with flagging and review workflows
Frequently asked questions about LC-MS/MS Toxicology Analyzer
Well-designed methods cover 100 to 300 compounds, but the number depends on chromatography, dwell times and required sensitivity. More compounds usually need longer runs or scheduled acquisition.
It reduces manual work and improves reproducibility for routine matrices. It adds complexity and maintenance, so it is most valuable for high-volume laboratories.
Matrix effects, calibrator quality, internal standard choice and carryover are the main limits. Proper validation and matrix-matched calibrators control these factors.
A toxicology analyzer is an instrument that detects and quantifies drugs, poisons, alcohol or metals in biological samples. Options include immunoassay analyzers for screening and chromatography-mass spectrometry or ICP-MS systems for confirmation and quantitation.
Immunoassays detect drug classes through antibody cross-reactivity, so they can give false positives. A mass spectrometry confirmation identifies and quantifies the specific compound, which is essential for clinical decisions and legal use.
Blood alcohol is commonly measured by headspace gas chromatography with flame ionization detection. The sample is heated in a sealed vial and the vapor is injected, using an internal standard such as n-propanol.