Linear MALDI-TOF Mass Spectrometer for Microbial Identification
用于微生物鉴定的线性 MALDI-TOF 质谱仪
- SKU TOM-0001
- TOF-MS
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
- IEC 61010-2-101 IVD safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Mass range (working) | m/z 2000–20000 |
|---|---|
| Mass range (maximum) | up to m/z 100000 (linear mode) |
| Resolution (m/z 5000) | ≥ 500 FWHM |
| Mass accuracy | ≤ 500 ppm (external calibration) |
| Laser type and wavelength | Solid-state, 355 nm |
| Laser repetition rate | 60–200 Hz |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D7134, ISO 11260, ISO/TS 16940
What is the Linear MALDI-TOF Mass Spectrometer for Microbial Identification?
A linear MALDI-TOF mass spectrometer records protein fingerprints of whole bacterial or fungal cells, mostly ribosomal proteins between m/z 2000 and 20000, and matches them against a reference database. It gives species-level identification in minutes and is standard in clinical and industrial microbiology.
What is the Linear MALDI-TOF Mass Spectrometer for Microbial Identification?
A linear MALDI-TOF mass spectrometer records protein fingerprints of whole bacterial or fungal cells, mostly ribosomal proteins between m/z 2000 and 20000, and matches them against a reference database. It gives species-level identification in minutes and is standard in clinical and industrial microbiology.
Overview
This instrument is optimized for speed and robustness rather than high resolution. A colony is smeared onto a steel or disposable target, overlaid with matrix solution such as alpha-cyano-4-hydroxycinnamic acid, and ablated by a nitrogen or solid-state laser at 337 or 355 nm running at 60 to 200 Hz. Ions accelerate to about 20 kV and travel through a linear flight tube of 0.8 to 1.0 m without a reflectron, which maximizes sensitivity for large proteins though resolution is modest, about 500 to 1000 at m/z 5000. The resulting spectrum is compared with a curated database by software that calculates a score or confidence value. Target plates of 48 to 96 spots allow high throughput, and a full identification typically takes under 5 minutes of acquisition. Databases, validation claims and regulatory clearance depend on the software supplied and must be confirmed for the intended use.
How does it work?
In a time-of-flight mass spectrometer, ions are given the same kinetic energy by an accelerating potential of several hundred volts to 20 kV and then drift through a field-free tube. Flight time scales with the square root of m/z, so the detector, a microchannel plate or fast electron multiplier, records ions in order of mass. Resolution depends on flight path length and on correcting initial energy spread. A reflectron, an ion mirror, extends the path and refocuses ions of equal mass but different energies, raising resolution from about 1000 to 20000 or more. MALDI sources pulse ions naturally, while orthogonal acceleration converts continuous ESI or ICP beams into pulses. Time-to-digital converters or fast digitizers at gigahertz rates record the signals.
What are the key features?
- Linear flight tube optimized for sensitivity on proteins between m/z 2000 and 20000
- Solid-state laser at 355 nm with repetition rate of 200 Hz for fast spectrum accumulation
- Disposable or reusable target plates with 48 to 96 spots and barcode tracking
- Automated spectrum acquisition with quality check and score-based identification
- Compact benchtop footprint with a simple maintenance interface and load-lock target entry
- User-extendable reference database for in-house strains and environmental isolates
What are the technical specifications?
A typical linear maldi-tof mass spectrometer for microbial identification configuration offers mass range (working) of m/z 2000–20000, mass range (maximum) of up to m/z 100000 (linear mode), resolution (m/z 5000) of ≥ 500 FWHM, mass accuracy of ≤ 500 ppm (external calibration), laser type and wavelength of Solid-state, 355 nm and laser repetition rate of 60–200 Hz. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Linear MALDI-TOF Mass Spectrometer for Microbial Identification is used in clinical bacteria and yeast identification, food and beverage spoilage organism identification, pharmaceutical environmental monitoring isolates, veterinary and agricultural microbiology and culture collection authentication. Across the wider TOF-MS class, laboratories also deploy it for microbial identification in clinical microbiology, intact protein and peptide mass analysis and synthetic polymer molecular weight distribution.
Which standards and certifications apply?
Relevant standards and methods include ASTM D7134, ISO 11260, ISO/TS 16940, USP <1052> Biotechnology-Derived Articles, ASTM E2933, ISO 15189, ISO/IEC 17025 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing, IEC 61010-1 safety and IEC 61010-2-101 IVD safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
TOF instruments are sensitive to temperature changes, because drift-tube expansion shifts flight time, so install them in a room stable within ±1 to 2 °C and away from drafts. Calibrate frequently with standards or internal lock mass to sustain mass accuracy below 5 ppm. MALDI systems require matrix preparation, such as alpha-cyano-4-hydroxycinnamic acid or sinapinic acid, and careful target plate cleaning. Detector microchannel plates degrade over time, so monitor gain and plan replacement, usually after several years. Keep the vacuum system leak tight, with analyzer pressure below 10^-6 mbar. Regulated laboratories, particularly those using MALDI for clinical microbiology, should confirm validated databases and software supporting audit trail and access control.
How do I select the right Linear MALDI-TOF Mass Spectrometer for Microbial Identification?
Select a linear MALDI-TOF system when the task is high-throughput microbial identification against an established database. It is not suited to peptide sequencing or accurate-mass work. In an RFQ, specify database content and validation status, laser type and rate, target capacity, LIS connectivity and audit-trail support.
How does it compare with other TOF-MS variants?
Select by ionization source and required resolving power. MALDI-TOF suits intact proteins, polymers and microbial identification, where linear mode reaches m/z above 100000 and reflectron mode improves accuracy for peptides. Orthogonal-acceleration ESI-TOF gives accurate mass for LC-MS at rates compatible with UHPLC. Multi-reflecting designs reach resolving power of 50000 to 100000 in a compact footprint for demanding screening. Consider mass accuracy, typically 1 to 5 ppm with calibration, resolution at your m/z range, acquisition rate, and dynamic range, which is narrower than quadrupoles. Decide whether tandem MS is needed, since TOF/TOF adds fragmentation for sequencing. Check software for microbial databases or polymer analysis, calibration strategy, and service support.
Full specifications
| Mass range (working) | m/z 2000–20000 |
|---|---|
| Mass range (maximum) | up to m/z 100000 (linear mode) |
| Resolution (m/z 5000) | ≥ 500 FWHM |
| Mass accuracy | ≤ 500 ppm (external calibration) |
| Laser type and wavelength | Solid-state, 355 nm |
| Laser repetition rate | 60–200 Hz |
| Acceleration voltage | up to 20 kV |
| Flight tube length | ~0.8–1.0 m (linear) |
| Target capacity | 48–96 spots per plate |
| Time per identification | < 5 min acquisition |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Clinical bacteria and yeast identification
- Food and beverage spoilage organism identification
- Pharmaceutical environmental monitoring isolates
- Veterinary and agricultural microbiology
- Culture collection authentication
Key features
- Linear flight tube optimized for sensitivity on proteins between m/z 2000 and 20000
- Solid-state laser at 355 nm with repetition rate of 200 Hz for fast spectrum accumulation
- Disposable or reusable target plates with 48 to 96 spots and barcode tracking
- Automated spectrum acquisition with quality check and score-based identification
- Compact benchtop footprint with a simple maintenance interface and load-lock target entry
- User-extendable reference database for in-house strains and environmental isolates
Frequently asked questions about Linear MALDI-TOF Mass Spectrometer for Microbial Identification
It detects mainly abundant ribosomal proteins from whole cells, producing a spectrum of peaks between m/z 2000 and 20000. Software compares this fingerprint with reference spectra from known organisms and reports the closest match with a score, often within minutes and without biochemical tests.
Usually yes. A single colony from solid media, typically 18 to 24 hours old, is needed to obtain sufficient biomass and a clean spectrum. Direct analysis from positive blood cultures is possible with additional preparation, and requires specific validated protocols.
Often, but not always. Some groups, such as Streptococcus mitis and pneumoniae or Shigella and Escherichia coli, have nearly identical ribosomal protein profiles and need supplementary tests. Database depth and software algorithm affect performance for such cases.
TOF analyzers theoretically have no upper mass limit, since every ion eventually reaches the detector. In practice MALDI-TOF in linear mode measures intact proteins and polymers above m/z 100000, while ESI-TOF instruments typically scan up to m/z 10000 to 20000, with multiply charged ions extending the effective molecular weight range.
A reflectron uses an electrostatic mirror to turn ions around. Ions with higher energy penetrate deeper and take longer in the mirror, so ions of the same mass but different energy arrive at the detector together. This corrects energy spread and lengthens the flight path, improving resolution.
With proper calibration, TOF instruments typically achieve 1 to 5 ppm mass accuracy, and multi-reflecting models can reach below 1 ppm with lock mass correction. Accuracy depends on temperature stability, calibration frequency, ion statistics and peak shape, so it must be verified during validation.