Photometric Titrator
光度滴定仪
- SKU TIT-0004
- Titrators
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Wavelength options | 520, 550, 574, 590, 610, 640, 660 nm (filters) |
|---|---|
| Probe path length | 10–20 mm |
| Transmission measuring range | 0–1000 mV equivalent (0–100% T) |
| Burette volume | 10 mL or 20 mL |
| Dosing resolution | 1/20,000 burette volume |
| Endpoint detection | Fixed level or curve inflection |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D664, ASTM D974, ASTM D6304
What is the Photometric Titrator?
A photometric titrator detects the endpoint with an optical sensor that measures light transmission at a fixed wavelength, typically 520 to 660 nm. It is used for complexometric, turbidimetric and indicator-based titrations where electrodes are unsuitable, such as water hardness, nonaqueous systems and colored samples.
What is the Photometric Titrator?
A photometric titrator detects the endpoint with an optical sensor that measures light transmission at a fixed wavelength, typically 520 to 660 nm. It is used for complexometric, turbidimetric and indicator-based titrations where electrodes are unsuitable, such as water hardness, nonaqueous systems and colored samples.
Overview
The optical probe, often a fiber or dip-in design with a short path length, transmits light through the solution and measures intensity with a photodiode. As indicator color changes, for example eriochrome black T turning from red to blue in EDTA titration, transmission changes sharply and the endpoint is evaluated automatically from the curve. The approach removes subjective visual judgment and works in solvents that cause electrode drift or in samples that foul a glass membrane. Turbidimetric titrations such as surfactant or sulfate determination follow the same principle. Many instruments combine optical and potentiometric inputs, allowing either sensor per method. Performance depends on indicator choice, consistent sample volume and avoiding air bubbles in the light path.
How does it work?
A titrator combines a precision burette or coulometric generator electrode, a stirred titration vessel, a sensor and a control unit. In volumetric titration, a motor-driven piston burette dispenses titrant while a sensor, usually a pH glass electrode, redox platinum electrode or photometric probe, tracks the solution. The endpoint is identified as a preset potential or the inflection point of the titration curve, and the result is calculated from titrant volume, concentration and stoichiometry. In coulometric Karl Fischer titration, iodine is generated electrochemically at a generator electrode and the charge passed, according to Faraday's law, gives the mass of water directly without a burette. Dynamic dosing slows addition near the endpoint to improve accuracy, and automatic sample changers allow unattended series.
What are the key features?
- Optical probe with selectable wavelength filters across the visible range
- Automatic endpoint evaluation removing subjective visual color judgment
- Suitable for turbid, colored or electrode-fouling sample matrices
- Combined optical and potentiometric inputs for method flexibility
- Corrosion-resistant probe body compatible with common titration solvents
- Stored methods for EDTA, surfactant and turbidimetric titrations
What are the technical specifications?
A typical photometric titrator configuration offers wavelength options of 520, 550, 574, 590, 610, 640, 660 nm (filters), probe path length of 10–20 mm, transmission measuring range of 0–1000 mV equivalent (0–100% T), burette volume of 10 mL or 20 mL, dosing resolution of 1/20,000 burette volume and endpoint detection of Fixed level or curve inflection. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Photometric Titrator is used in total and calcium hardness of water, anionic surfactant content, sulfate by turbidimetric titration, chloride in nonaqueous solvents and complexometric titration of metal ions in plating baths. Across the wider Titrators class, laboratories also deploy it for water determination in oils and solvents, acid and base number of petroleum products and chloride and salt content in food and water.
Which standards and certifications apply?
Relevant standards and methods include ASTM D664, ASTM D974, ASTM D6304, ASTM E203, ISO 760, USP <921> and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the titrator on a vibration-free bench away from drafts and direct sunlight. Standardize titrants regularly against a primary standard such as potassium hydrogen phthalate or sodium oxalate, and record the titer. Verify electrodes with buffers and check the burette volume gravimetrically at defined intervals. For Karl Fischer work, replace desiccant, keep the cell sealed, check drift before each series and verify with a certified water standard. Rinse tubing when changing titrants to avoid cross-contamination, and inspect piston seals and valves as part of preventive maintenance. Dispose of reagents, particularly pyridine-free Karl Fischer solutions and organic solvents, according to local waste rules.
How do I select the right Photometric Titrator?
Select this variant when indicator-based or turbidimetric titrations dominate, or when electrodes drift or foul in the sample. Choose the potentiometric model for acid number and pH-governed titrations. In an RFQ, specify the titrations to be run, wavelength filters required, sample volume range, whether electrode input must also be available and the reference methods involved.
How does it compare with other Titrators variants?
The decisive choice is the titration chemistry. Water in oils, solvents and solids calls for Karl Fischer, volumetric above about 0.1% water and coulometric below it. Acid and base number in petroleum products and general acid-base, redox, argentometric and complexometric titrations call for a potentiometric titrator. Consider sample throughput, because an automatic sample changer is justified beyond roughly 20 samples per day. Check burette resolution, available sensors and whether the software supports the standard methods you run, such as ASTM D664 or ASTM D6304. Also weigh solvent consumption and waste handling, reagent storage for moisture-sensitive titrants, and whether audit-trail and user management features are required for accredited or regulated work.
Full specifications
| Wavelength options | 520, 550, 574, 590, 610, 640, 660 nm (filters) |
|---|---|
| Probe path length | 10–20 mm |
| Transmission measuring range | 0–1000 mV equivalent (0–100% T) |
| Burette volume | 10 mL or 20 mL |
| Dosing resolution | 1/20,000 burette volume |
| Endpoint detection | Fixed level or curve inflection |
| Stirring speed | 0–1500 rpm |
| Sample volume range | 20–150 mL |
| Probe material | PEEK / glass fiber |
| Data interface | USB, RS-232 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Total and calcium hardness of water
- Anionic surfactant content
- Sulfate by turbidimetric titration
- Chloride in nonaqueous solvents
- Complexometric titration of metal ions in plating baths
Key features
- Optical probe with selectable wavelength filters across the visible range
- Automatic endpoint evaluation removing subjective visual color judgment
- Suitable for turbid, colored or electrode-fouling sample matrices
- Combined optical and potentiometric inputs for method flexibility
- Corrosion-resistant probe body compatible with common titration solvents
- Stored methods for EDTA, surfactant and turbidimetric titrations
Frequently asked questions about Photometric Titrator
Eriochrome black T titrations are usually monitored near 610–640 nm, where the blue endpoint form absorbs least. The best wavelength is confirmed by recording a spectrum of the indicator at start and end of titration.
Neither is universally better. Optical detection excels for indicator systems, turbid samples and nonaqueous solvents, while pH electrodes are preferred for acid-base titrations governed by hydrogen ion activity and do not require an indicator.
Rinse the probe with deionized water or the titration solvent after each determination and wipe the optical windows with a lint-free cloth. For fouling by fats or polymers, use a compatible solvent, and never abrade the optical surfaces.
Volumetric Karl Fischer adds iodine-containing titrant from a burette and suits samples above roughly 0.1% water. Coulometric Karl Fischer generates iodine electrochemically and suits trace water from about 1 ppm to 5%, because it uses no burette and has lower detection limits.
The titrator monitors a sensor signal such as potential, pH or light transmission during titrant addition. The endpoint is either a preset signal value or the inflection point of the titration curve found by first or second derivative evaluation, which is more robust for weak acids or unknown samples.
Standardize titrants whenever a new batch is prepared and then at intervals set by stability, commonly weekly for sodium hydroxide and every few days for Karl Fischer titrant. Record the titer and investigate any drift, since titer errors transfer directly into every result.