Wavelength-Dispersive XRF Spectrometer for Ore and Cement
矿石和水泥分析波长色散 XRF 光谱仪
- SKU SPO-0005
- Spectrometers
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
- Radiation safety type approval (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Element range | Be (or B) to U |
|---|---|
| X-ray tube power | 1–4 kW, up to 60 kV |
| Tube anode | Rh (end-window) |
| Analyzing crystals | LiF200, PET, Ge, multilayers (up to 8) |
| Detectors | Flow proportional, scintillation |
| Concentration range | ppm to 100% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E415, ASTM E1019, ISO 15350
What is the Wavelength-Dispersive XRF Spectrometer for Ore and Cement?
A wavelength-dispersive XRF spectrometer separates fluorescence X-rays with analyzer crystals and measures each wavelength with a goniometer or fixed channels. It provides high resolution and precision for major and minor oxides in ores, cement, slag and ceramics, typically on fused beads or pressed pellets.
What is the Wavelength-Dispersive XRF Spectrometer for Ore and Cement?
A wavelength-dispersive XRF spectrometer separates fluorescence X-rays with analyzer crystals and measures each wavelength with a goniometer or fixed channels. It provides high resolution and precision for major and minor oxides in ores, cement, slag and ceramics, typically on fused beads or pressed pellets.
Overview
An X-ray tube of 1–4 kW excites the sample, and a collimator, analyzing crystals and gas-flow or scintillation detectors measure characteristic wavelengths with spectral resolution of roughly 10–40 eV, several times better than energy-dispersive detection. Sequential instruments scan with a goniometer to measure many elements flexibly. Simultaneous instruments use fixed channels for high-speed production control of up to 30 elements in under a minute, as in cement plants. Fusion with lithium borate removes particle and mineralogical effects and gives accuracy to 0.05–0.2% for major oxides. Light elements from beryllium or boron up are measurable under vacuum. The technique needs stable cooling water, a P-10 counting gas supply and regulated radiation safety.
How does it work?
Atomic spectrometry relies on the fact that each element emits, absorbs or fluoresces radiation at characteristic wavelengths or energies. In optical emission, an arc, spark or argon plasma excites atoms, and a grating disperses the emitted light onto photomultiplier tubes or CCD/CMOS detectors. In atomic absorption, a hollow cathode lamp passes light through an atomized sample, and absorbance follows Beer–Lambert behavior. In X-ray fluorescence, a primary X-ray beam ejects inner-shell electrons, and the characteristic X-rays emitted as outer electrons fill the vacancies are measured by energy-dispersive or wavelength-dispersive detection. Quantification requires calibration with certified reference materials and correction for matrix effects and spectral interferences, handled by the instrument software.
What are the key features?
- High-power end-window X-ray tube of 1 to 4 kW for heavy and light elements
- Multiple analyzing crystals covering beryllium to uranium under vacuum
- Sequential goniometer or simultaneous fixed-channel configurations
- Gas-flow proportional and scintillation detectors with automatic counting-range switching
- Fusion-bead calibration programs for ores, cement, slag and ceramics
- Sample changer with 48 to 100 positions and vacuum load lock
What are the technical specifications?
A typical wavelength-dispersive xrf spectrometer for ore and cement configuration offers element range of Be (or B) to U, X-ray tube power of 1–4 kW, up to 60 kV, tube anode of Rh (end-window), analyzing crystals of LiF200, PET, Ge, multilayers (up to 8), detectors of Flow proportional, scintillation and concentration range of ppm to 100%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Wavelength-Dispersive XRF Spectrometer for Ore and Cement is used in cement raw meal and clinker control, iron ore and bauxite assay, slag and refractory composition, ceramic and glass raw materials and reference-method analysis for EDXRF calibration. Across the wider Spectrometers class, laboratories also deploy it for alloy grade verification in foundries and steel mills, sulfur in fuels and oils and wear metals in used lubricating oil.
Which standards and certifications apply?
Relevant standards and methods include ASTM E415, ASTM E1019, ISO 15350, ASTM D4294, ISO 8754, ASTM D5185, ISO 11885 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing, RoHS and Radiation safety type approval; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install spectrometers in a temperature-stable room, typically 18–25 °C with controlled humidity, on a stable bench with adequate exhaust for plasma or spark gases. Argon supply of suitable purity is needed for spark OES and ICP. XRF systems require radiation safety registration under local regulations and periodic leak checks. Calibrate with certified reference materials, then monitor drift with control samples and apply type standardization or recalibration at defined intervals. Keep sample surfaces consistent, since grinding finish and pellet density affect results. Replace spark stand electrodes, nebulizers, lamps and windows as maintenance schedules dictate, and document validation in line with ISO/IEC 17025.
How do I select the right Wavelength-Dispersive XRF Spectrometer for Ore and Cement?
Select this variant when precision, light-element performance and resolution of overlapping lines matter, such as cement and mining assay with fused beads. EDXRF suits screening and sulfur-in-oil at lower cost. In an RFQ, define elements and ranges, sequential versus simultaneous layout, fusion machine compatibility, sample handling capacity and utilities including cooling water and P-10 gas supply.
How does it compare with other Spectrometers variants?
Sample form and required detection limit narrow the choice quickly. Solid metals for production control suit spark OES, which reads major and trace alloying elements within seconds. Liquids and digested solids measured to ppb levels suit ICP-OES for multi-element work or atomic absorption for one or two elements at lower capital cost. Powders, liquids and solids needing minimal preparation suit XRF, with energy-dispersive models for sulfur in fuels and screening, and wavelength-dispersive models for higher resolution in ore and cement. Consider elements of interest, since light elements such as carbon and boron are difficult by XRF. Also weigh argon consumption, sample preparation effort, operator skill and available reference materials matching your matrix.
Full specifications
| Element range | Be (or B) to U |
|---|---|
| X-ray tube power | 1–4 kW, up to 60 kV |
| Tube anode | Rh (end-window) |
| Analyzing crystals | LiF200, PET, Ge, multilayers (up to 8) |
| Detectors | Flow proportional, scintillation |
| Concentration range | ppm to 100% |
| Repeatability (major oxides, fused bead) | RSD ≤ 0.1–0.3% |
| Sample positions | 48 to 100 |
| Counting gas | P-10 (90% Ar / 10% CH4) |
| Cooling water | Closed-loop chiller, 4–6 L/min |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Cement raw meal and clinker control
- Iron ore and bauxite assay
- Slag and refractory composition
- Ceramic and glass raw materials
- Reference-method analysis for EDXRF calibration
Key features
- High-power end-window X-ray tube of 1 to 4 kW for heavy and light elements
- Multiple analyzing crystals covering beryllium to uranium under vacuum
- Sequential goniometer or simultaneous fixed-channel configurations
- Gas-flow proportional and scintillation detectors with automatic counting-range switching
- Fusion-bead calibration programs for ores, cement, slag and ceramics
- Sample changer with 48 to 100 positions and vacuum load lock
Frequently asked questions about Wavelength-Dispersive XRF Spectrometer for Ore and Cement
Fusing the sample with lithium tetraborate or metaborate dissolves it into homogeneous glass, eliminating particle size and mineralogical effects that limit pressed pellets. The dilution reduces matrix interference and permits calibration with synthetic standards, at the cost of losing trace sensitivity.
Sequential WDXRF uses one scanning goniometer, so it can measure any element and suits varied samples. Simultaneous WDXRF has a fixed channel for each element, measuring all in seconds, which suits high-volume production control of a fixed element set.
With fused beads and matrix-matched certified standards, WDXRF typically matches classical gravimetric or ICP results for major oxides within 0.1–0.3% absolute. Accuracy depends on calibration quality, bead preparation consistency and stable instrument conditions.
ICP-OES measures many elements simultaneously with a wide linear range from ppb to percent, while atomic absorption measures one element at a time using a dedicated lamp. Flame AAS has lower capital cost, and graphite furnace AAS reaches ppb detection with very small sample volumes.
A spark optical emission spectrometer is the standard for solid metal alloys, giving results for major, minor and trace elements including carbon, sulfur and phosphorus in steel within about 30 seconds. XRF is used when non-destructive or portable analysis is preferred.
Standard XRF has poor sensitivity below sodium because low-energy X-rays are absorbed in air and sample. Wavelength-dispersive systems under vacuum can measure elements down to beryllium or boron, but carbon in steel is determined by spark OES or combustion analysis instead.