Cryogenic Tissue Grinder
冷冻组织研磨仪
- SKU TIH-0004
- Tissue Homogenizers
- CE (to be confirmed)
- RoHS (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Cooling temperature | About -196 °C (liquid nitrogen) |
|---|---|
| Impact frequency | 5–30 Hz |
| Vial volume | 0.5–50 mL |
| Grinding time | 2–10 min |
| Pre-cooling | Automatic |
| Vial material | Stainless steel or polycarbonate |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-051, ISO 6887-1
What is the Cryogenic Tissue Grinder?
A cryogenic tissue grinder pulverizes frozen, brittle samples in a liquid-nitrogen-cooled vial using a magnetic impactor. It is the method of choice for hard tissues such as bone, seeds and tumor biopsies where RNA and metabolites must remain intact.
What is the Cryogenic Tissue Grinder?
A cryogenic tissue grinder pulverizes frozen, brittle samples in a liquid-nitrogen-cooled vial using a magnetic impactor. It is the method of choice for hard tissues such as bone, seeds and tumor biopsies where RNA and metabolites must remain intact.
Overview
The cryogenic grinder holds a sealed stainless steel or polycarbonate vial with a steel impactor in a chamber cooled by liquid nitrogen to about -196 degrees C. Electromagnetic coils drive the impactor back and forth at 5 to 30 impacts per second, pulverizing the frozen sample into a fine powder. Vials hold 0.5 to 50 mL of sample, and the process takes 2 to 10 minutes. Because the sample stays frozen, RNA, proteins and metabolites are preserved without enzymatic degradation. The technique works well for bone, seeds, tough plant tissue, tumor samples and animal organs. Pre-cooling of the vial is automatic. Users need a liquid nitrogen supply and must follow cryogenic safety practices. Throughput is low to moderate, with one vial or a few at a time. The equipment is larger and more costly than bead mills.
How does it work?
Tissue homogenizers apply mechanical force to rupture tissue structure and cell walls. Bead mills shake sealed tubes containing the sample and ceramic, glass or steel beads at high speed, typically 4 to 7 m/s, so that collisions crush and shear the tissue. Motorized Potter-Elvehjem homogenizers rotate a close-fitting pestle inside a glass tube at 200 to 2,000 rpm, forcing tissue through a narrow clearance of 0.1 to 0.15 mm. Cryogenic grinders use a magnetic impactor in liquid-nitrogen-cooled vials to pulverize brittle frozen tissue. Paddle blenders use reciprocating paddles to press and release material in a sterile bag. Key components are a drive unit, a vessel or tube, the grinding element such as beads or pestle, and a cooling or cryogenic system to preserve labile biomolecules.
What are the key features?
- Liquid-nitrogen cooling to about -196 °C keeps samples frozen
- Magnetic impactor pulverizes hard, brittle tissue to powder
- Vial volumes of 0.5–50 mL for varied sample sizes
- Automatic pre-cooling and programmed grinding cycles
- Preserves RNA, proteins and metabolites without degradation
- Sealed vials contain samples and reduce cross-contamination
What are the technical specifications?
A typical cryogenic tissue grinder configuration offers cooling temperature of About -196 °C (liquid nitrogen), impact frequency of 5–30 Hz, vial volume of 0.5–50 mL, grinding time of 2–10 min, pre-cooling of Automatic and vial material of Stainless steel or polycarbonate. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Cryogenic Tissue Grinder is used in bone and hard tissue grinding, seed and plant pulverization, tumor biopsy preparation, metabolomics of frozen tissue and forensic sample preparation. Across the wider Tissue Homogenizers class, laboratories also deploy it for DNA and RNA extraction from tissue, protein and enzyme extraction and plant and seed sample preparation.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-051, ISO 6887-1, ISO 7218 and GB 4789.1. Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Keep tissue cold and process quickly to preserve RNA, proteins and enzyme activity; add stabilizing buffers where appropriate. Load tubes symmetrically in bead mills and do not exceed the rated tube fill volume, which is usually half of the tube. Select bead size and material to match tissue hardness, and avoid contamination with metals if trace element analysis follows. Follow the supplier's method for each tissue type, and validate homogenization efficiency against a reference method. For infectious or biohazardous material, use sealed tubes and work in a biosafety cabinet. Clean glass homogenizers thoroughly, and handle liquid nitrogen with protective equipment. Comply with IEC 61010-1 and local biosafety rules.
How do I select the right Cryogenic Tissue Grinder?
Select a cryogenic grinder for hard, brittle or fibrous samples where frozen processing is essential. A bead mill has higher throughput for soft tissue. In the RFQ, specify vial sizes, sample type, liquid nitrogen supply and throughput.
How does it compare with other Tissue Homogenizers variants?
Choose a tissue homogenizer by sample type, throughput, sample size and downstream analysis. Bead mills handle soft and tough tissues, microbes and plant material in sealed tubes, and are the usual choice for high throughput with 24 or more samples at once. Choose cryogenic cooling for RNA and labile metabolites. Potter-Elvehjem homogenizers give gentle, reproducible disruption of soft tissue such as liver and brain, preserving organelles. Cryogenic grinders pulverize very hard or brittle samples such as bone and seeds. Paddle blenders suit food and microbiology samples of 10 to 400 mL. Check tube volumes, bead types, speed control, cooling and aerosol containment. For regulated testing, consider cross-contamination prevention and documented cleaning steps. Confirm noise, vibration and safety interlocks.
Full specifications
| Cooling temperature | About -196 °C (liquid nitrogen) |
|---|---|
| Impact frequency | 5–30 Hz |
| Vial volume | 0.5–50 mL |
| Grinding time | 2–10 min |
| Pre-cooling | Automatic |
| Vial material | Stainless steel or polycarbonate |
| Impactor material | Hardened stainless steel |
| Sample type | Bone, seeds, tough tissue |
| Programs | Up to 10 stored |
| Noise level | < 75 dB(A) |
| Power supply | 200–240 V, 50/60 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Bone and hard tissue grinding
- Seed and plant pulverization
- Tumor biopsy preparation
- Metabolomics of frozen tissue
- Forensic sample preparation
Key features
- Liquid-nitrogen cooling to about -196 °C keeps samples frozen
- Magnetic impactor pulverizes hard, brittle tissue to powder
- Vial volumes of 0.5–50 mL for varied sample sizes
- Automatic pre-cooling and programmed grinding cycles
- Preserves RNA, proteins and metabolites without degradation
- Sealed vials contain samples and reduce cross-contamination
Frequently asked questions about Cryogenic Tissue Grinder
Grinding at liquid nitrogen temperature keeps enzymes inactive and makes tissue brittle, which prevents RNA degradation and gives a fine powder for extraction.
Vials and impactors are cleaned between samples, and sealed vials contain the powder. Use dedicated vials for trace analysis and verify cleaning with blanks.
Liquid nitrogen requires ventilation, face shield, insulated gloves and oxygen monitoring. Vials may build pressure, so use rated vials and follow supplier procedures.
A tissue homogenizer breaks tissue into uniform suspensions or lysates by bead impact, grinding, pestle shear or paddle action. It prepares samples for DNA, RNA, protein and metabolite analysis in life science, clinical and food laboratories.
Bead mills with cooling and cryogenic grinders protect RNA by keeping samples cold and processing quickly. Use RNase-free tubes and lysis buffer that inactivates nucleases immediately.
A bead mill shakes sealed tubes with beads and handles many samples, while a rotor-stator probe is immersed in each sample. Bead mills avoid cross-contamination, whereas probes handle larger volumes.