Skip to main content
Lab Manufacturers China
UV-Fluorescence Total Sulfur Analyzer for Fuels — product photo

UV-Fluorescence Total Sulfur Analyzer for Fuels

燃料紫外荧光总硫分析仪

  • SKU SUA-0003
  • Sulfur Analyzers
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Test methodASTM D5453, ISO 20846, GB/T 34100
Measurement range1.0–8000 mg/kg (detection limit about 0.2 mg/kg)
Detection limitAbout 0.2 mg/kg
Furnace temperatureUp to 1100 °C
Injection volume5–50 µL
Analysis time3–6 min per sample
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D4294, ASTM D2622, ASTM D5453

What is the UV-Fluorescence Total Sulfur Analyzer for Fuels?

A UV-fluorescence sulfur analyzer combusts a fuel sample at about 1050 °C, converts sulfur to sulfur dioxide, and measures its ultraviolet-excited fluorescence. It follows ASTM D5453 and ISO 20846 for sulfur from about 1 to 8000 mg/kg in gasoline, diesel, jet fuel and other light hydrocarbons.

What is the UV-Fluorescence Total Sulfur Analyzer for Fuels?

A UV-fluorescence sulfur analyzer combusts a fuel sample at about 1050 °C, converts sulfur to sulfur dioxide, and measures its ultraviolet-excited fluorescence. It follows ASTM D5453 and ISO 20846 for sulfur from about 1 to 8000 mg/kg in gasoline, diesel, jet fuel and other light hydrocarbons.

Overview

The UV-fluorescence analyzer injects a small sample, typically 5 to 50 µL, into a combustion tube at about 1000 to 1100 °C in a stream of oxygen and argon or helium. Sulfur in the sample oxidizes to sulfur dioxide. After the gas stream passes a membrane dryer to remove water, it enters a fluorescence cell where pulsed ultraviolet light near 220 nm excites the SO2 molecules, which emit fluorescence at longer wavelengths. A photomultiplier measures the emission, and the signal is integrated to obtain total sulfur after calibration with certified standards. Liquid sampling uses a syringe drive, boat inlet or autosampler. The technique offers a wide dynamic range and low detection limit of about 0.2 mg/kg, making it widely used for ultra-low sulfur fuels. Nitrogen interference is minimal with proper design. The instrument needs oxygen and carrier gases and periodic maintenance of the combustion tube.

How does it work?

Petroleum sulfur analyzers use three main principles. In X-ray fluorescence, an X-ray tube excites sulfur atoms in a liquid sample held in a cell behind a thin film, and the detector measures the characteristic sulfur K-alpha line near 2.3 keV, whose intensity is proportional to concentration. Energy-dispersive systems read the whole spectrum with a silicon drift or proportional detector, while wavelength-dispersive systems use a crystal to select the line, giving better resolution. In UV fluorescence, the sample is burned in an oxygen-rich furnace at about 1000 to 1100 °C, converting sulfur to sulfur dioxide, which is excited by ultraviolet light and emits measurable fluorescence. Lamp combustion burns the sample in a wick lamp and titrates the sulfur oxides. Calibration with certified sulfur standards in a matching matrix is essential.

What are the key features?

  • Combustion furnace at 1000–1100 °C
  • Pulsed UV source and photomultiplier detector
  • Membrane dryer for water removal
  • Syringe or boat inlet, optional autosampler
  • Detection limit near 0.2 mg/kg
  • Wide dynamic range up to 8000 mg/kg

What are the technical specifications?

A typical uv-fluorescence total sulfur analyzer for fuels configuration offers test method of ASTM D5453, ISO 20846, GB/T 34100, measurement range of 1.0–8000 mg/kg (detection limit about 0.2 mg/kg), detection limit of About 0.2 mg/kg, furnace temperature of Up to 1100 °C, injection volume of 5–50 µL and analysis time of 3–6 min per sample. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The UV-Fluorescence Total Sulfur Analyzer for Fuels is used in ultra-low sulfur gasoline and diesel QC, jet fuel and naphtha testing, catalyst feed monitoring and biofuel and blend sulfur analysis. Across the wider Sulfur Analyzers class, laboratories also deploy it for ultra-low sulfur diesel and gasoline release, crude oil and residual fuel sulfur screening and marine fuel compliance (sulfur cap).

Which standards and certifications apply?

Relevant standards and methods include ASTM D4294, ASTM D2622, ASTM D5453, ASTM D7039, ASTM D1266, ISO 8754, ISO 20846, IP 336, GB/T 11140 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

X-ray based analyzers require radiation safety compliance, including registration or licensing according to local regulation and periodic leak checks of the tube housing. Calibrate with certified sulfur-in-oil standards matching the sample matrix, and verify daily with a check standard. Keep sample cells and films clean and replace films regularly, since pinholes or contamination bias results. For UV fluorescence, maintain the combustion furnace, inlet septa, oxygen and carrier gas supplies and dryer membranes. Control sample temperature and avoid evaporation of light samples before injection. Store low-sulfur samples in clean containers to prevent contamination. Record instrument drift and apply drift correction standards as specified. Follow solvent and waste handling rules for used cells, films and sample residues.

How do I select the right UV-Fluorescence Total Sulfur Analyzer for Fuels?

Select UV fluorescence when you must report sulfur below 10 mg/kg, need sub-mg/kg detection, or test naphtha and gasoline streams protecting catalysts. It needs gases and furnace maintenance. In an RFQ specify range, autosampler capacity, inlet type for liquids or gases, and gas supply arrangements.

How does it compare with other Sulfur Analyzers variants?

Choose by required detection limit and the methods in your specification. Low-cost energy-dispersive XRF (ASTM D4294) suits sulfur from about 16 mg/kg to 5 mass percent in crude oil, residual fuel and diesel. Wavelength-dispersive XRF (D2622) gives better resolution and lower detection limits for routine fuels. Monochromatic WDXRF (D7039) and UV fluorescence (D5453) are the usual choices for ultra-low sulfur gasoline and diesel at 10 mg/kg or below. The lamp method (D1266) is a classical check for light products. Also consider matrix effects from oxygenates and additives, sample throughput with autosamplers, helium or gas needs, radiation safety approvals for X-ray tubes, and availability of calibration standards in your concentration range. Service access and consumable cost matter as much as purchase price.

Full specifications

Full specifications
Test methodASTM D5453, ISO 20846, GB/T 34100
Measurement range1.0–8000 mg/kg (detection limit about 0.2 mg/kg)
Detection limitAbout 0.2 mg/kg
Furnace temperatureUp to 1100 °C
Injection volume5–50 µL
Analysis time3–6 min per sample
RepeatabilityPer method, about 5% at 10 mg/kg
GasesOxygen and argon or helium
AutosamplerUp to 100 vials (optional)
Power supply220 V, 50 Hz, 1000 W

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Ultra-low sulfur gasoline and diesel QC
  • Jet fuel and naphtha testing
  • Catalyst feed monitoring
  • Biofuel and blend sulfur analysis

Key features

  • Combustion furnace at 1000–1100 °C
  • Pulsed UV source and photomultiplier detector
  • Membrane dryer for water removal
  • Syringe or boat inlet, optional autosampler
  • Detection limit near 0.2 mg/kg
  • Wide dynamic range up to 8000 mg/kg

Frequently asked questions about UV-Fluorescence Total Sulfur Analyzer for Fuels

Typical instruments reach about 0.2 mg/kg, and ASTM D5453 applies from 1.0 to 8000 mg/kg. Actual limits depend on blank control, sample handling, inlet cleanliness and calibration quality.

Related models

All Sulfur Analyzers

Request a quotation for the UV-Fluorescence Total Sulfur Analyzer for Fuels

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote