Skip to main content
Lab Manufacturers China
Benchtop ED-XRF Sulfur Analyzer for Petroleum Products — product photo

Benchtop ED-XRF Sulfur Analyzer for Petroleum Products

石油产品台式 ED-XRF 硫分析仪

  • SKU SUA-0001
  • Sulfur Analyzers
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • Radiation safety registration per local rules (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Test methodASTM D4294, ISO 8754, GB/T 17040
Measurement range16 mg/kg–5 mass %
DetectorSilicon drift or proportional counter
X-ray tubeLow-power, up to 50 W
Measurement time100–300 s
RepeatabilityPer method table, about 3–5% at 500 mg/kg
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D4294, ASTM D2622, ASTM D5453

What is the Benchtop ED-XRF Sulfur Analyzer for Petroleum Products?

A benchtop energy-dispersive XRF sulfur analyzer measures total sulfur in crude oil, diesel, residual fuel and lubricants from about 16 mg/kg to 5 mass percent under ASTM D4294. It needs little sample preparation, runs in minutes and is widely used in refinery and terminal laboratories for routine screening.

What is the Benchtop ED-XRF Sulfur Analyzer for Petroleum Products?

A benchtop energy-dispersive XRF sulfur analyzer measures total sulfur in crude oil, diesel, residual fuel and lubricants from about 16 mg/kg to 5 mass percent under ASTM D4294. It needs little sample preparation, runs in minutes and is widely used in refinery and terminal laboratories for routine screening.

Overview

The benchtop ED-XRF analyzer uses a low-power X-ray tube, often under 50 W, to irradiate a liquid sample in a disposable cell sealed with a thin polymer film. A silicon drift or proportional counter measures the sulfur K-alpha X-ray intensity, which software converts to concentration using a calibration curve from sulfur-in-oil standards. A measurement typically lasts 100 to 300 seconds, with longer times improving detection at lower concentrations. Because results depend on matrix, instruments apply compensation for carbon-to-hydrogen ratio variation, and calibrations may be established separately for diesel, crude oil and residual fuels. The design requires no combustion gases, although helium purge may be optional for lower limits. Basic ED-XRF is a cost-effective choice for higher sulfur levels, but performance below about 20 mg/kg depends on model and does not typically meet ultra-low sulfur needs.

How does it work?

Petroleum sulfur analyzers use three main principles. In X-ray fluorescence, an X-ray tube excites sulfur atoms in a liquid sample held in a cell behind a thin film, and the detector measures the characteristic sulfur K-alpha line near 2.3 keV, whose intensity is proportional to concentration. Energy-dispersive systems read the whole spectrum with a silicon drift or proportional detector, while wavelength-dispersive systems use a crystal to select the line, giving better resolution. In UV fluorescence, the sample is burned in an oxygen-rich furnace at about 1000 to 1100 °C, converting sulfur to sulfur dioxide, which is excited by ultraviolet light and emits measurable fluorescence. Lamp combustion burns the sample in a wick lamp and titrates the sulfur oxides. Calibration with certified sulfur standards in a matching matrix is essential.

What are the key features?

  • Low-power X-ray tube with shielded housing
  • Silicon drift or proportional counter detector
  • Disposable sample cells with thin film windows
  • Matrix compensation for carbon-to-hydrogen variation
  • Measurement time of 100–300 s per sample
  • Optional helium purge and multi-position sample changer

What are the technical specifications?

A typical benchtop ed-xrf sulfur analyzer for petroleum products configuration offers test method of ASTM D4294, ISO 8754, GB/T 17040, measurement range of 16 mg/kg–5 mass %, detector of Silicon drift or proportional counter, X-ray tube of Low-power, up to 50 W, measurement time of 100–300 s and repeatability of Per method table, about 3–5% at 500 mg/kg. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Benchtop ED-XRF Sulfur Analyzer for Petroleum Products is used in crude oil and heating oil screening, marine fuel sulfur checks, lubricant and additive sulfur analysis and terminal incoming inspection. Across the wider Sulfur Analyzers class, laboratories also deploy it for ultra-low sulfur diesel and gasoline release, crude oil and residual fuel sulfur screening and marine fuel compliance (sulfur cap).

Which standards and certifications apply?

Relevant standards and methods include ASTM D4294, ASTM D2622, ASTM D5453, ASTM D7039, ASTM D1266, ISO 8754, ISO 20846, IP 336, GB/T 11140 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and Radiation safety registration per local rules; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

X-ray based analyzers require radiation safety compliance, including registration or licensing according to local regulation and periodic leak checks of the tube housing. Calibrate with certified sulfur-in-oil standards matching the sample matrix, and verify daily with a check standard. Keep sample cells and films clean and replace films regularly, since pinholes or contamination bias results. For UV fluorescence, maintain the combustion furnace, inlet septa, oxygen and carrier gas supplies and dryer membranes. Control sample temperature and avoid evaporation of light samples before injection. Store low-sulfur samples in clean containers to prevent contamination. Record instrument drift and apply drift correction standards as specified. Follow solvent and waste handling rules for used cells, films and sample residues.

How do I select the right Benchtop ED-XRF Sulfur Analyzer for Petroleum Products?

Select the benchtop ED-XRF for routine sulfur checks above about 20 mg/kg where low operating cost, fast results and simple sample handling matter. It is not the first choice for ultra-low sulfur gasoline. In an RFQ specify sulfur range, matrices, sample changer, helium option and calibration standards.

How does it compare with other Sulfur Analyzers variants?

Choose by required detection limit and the methods in your specification. Low-cost energy-dispersive XRF (ASTM D4294) suits sulfur from about 16 mg/kg to 5 mass percent in crude oil, residual fuel and diesel. Wavelength-dispersive XRF (D2622) gives better resolution and lower detection limits for routine fuels. Monochromatic WDXRF (D7039) and UV fluorescence (D5453) are the usual choices for ultra-low sulfur gasoline and diesel at 10 mg/kg or below. The lamp method (D1266) is a classical check for light products. Also consider matrix effects from oxygenates and additives, sample throughput with autosamplers, helium or gas needs, radiation safety approvals for X-ray tubes, and availability of calibration standards in your concentration range. Service access and consumable cost matter as much as purchase price.

Full specifications

Full specifications
Test methodASTM D4294, ISO 8754, GB/T 17040
Measurement range16 mg/kg–5 mass %
DetectorSilicon drift or proportional counter
X-ray tubeLow-power, up to 50 W
Measurement time100–300 s
RepeatabilityPer method table, about 3–5% at 500 mg/kg
Sample cell volumeAbout 5–8 mL
Sample changerOptional 6 or 12 positions
Purge gasOptional helium
Power supply220 V, 50 Hz, 200 W

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Crude oil and heating oil screening
  • Marine fuel sulfur checks
  • Lubricant and additive sulfur analysis
  • Terminal incoming inspection

Key features

  • Low-power X-ray tube with shielded housing
  • Silicon drift or proportional counter detector
  • Disposable sample cells with thin film windows
  • Matrix compensation for carbon-to-hydrogen variation
  • Measurement time of 100–300 s per sample
  • Optional helium purge and multi-position sample changer

Frequently asked questions about Benchtop ED-XRF Sulfur Analyzer for Petroleum Products

ASTM D4294 is written for sulfur from about 16 mg/kg to 5 mass percent. Some instruments reach lower limits with longer counting or helium purge, but confirm documented performance with your matrix before relying on results near 10 mg/kg.

Related models

All Sulfur Analyzers

Request a quotation for the Benchtop ED-XRF Sulfur Analyzer for Petroleum Products

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote