Falling Number Apparatus
降落数值仪
- SKU GRA-0003
- Grain Analysis
- CE (to be confirmed)
- RoHS (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Measurement | Falling number, seconds |
|---|---|
| Measurement range | 60–1000 s |
| Sample mass | 7.0 g (14% moisture basis) |
| Water volume | 25 mL |
| Bath temperature | 100 °C, ±0.2 °C |
| Timing accuracy | ±0.1 s |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 712 (moisture), ISO 3093 (falling number), AACC Method 56-81B
What is the Falling Number Apparatus?
A laboratory instrument that measures alpha-amylase activity in grain by timing a stirrer through a heated flour paste. It is the right choice for assessing sprout damage in wheat and predicting baking quality.
What is the Falling Number Apparatus?
A laboratory instrument that measures alpha-amylase activity in grain by timing a stirrer through a heated flour paste. It is the right choice for assessing sprout damage in wheat and predicting baking quality.
Overview
The falling number apparatus consists of a boiling water bath, a tube with a stirrer, a dispenser and a timer. A ground grain sample is dispersed in water in the tube, which is placed in the boiling bath and stirred. As the starch gelatinises, the suspension thickens, and the time in seconds for the stirrer to fall a fixed distance through the paste is recorded as the falling number. High alpha-amylase activity from germinated or sprouted grain thins the paste and gives a low value, while sound grain gives a higher value. Only two parameters, sample mass and water volume, plus the grinding method and bath temperature, affect the result, so the method is robust but demands precision. Results guide wheat blending at the mill and are widely used in trade specifications. The apparatus can run several tubes in sequence, and automated models reduce operator timing errors.
How does it work?
Grain analysis uses several techniques. Near-infrared instruments measure the reflectance or transmission of light at selected wavelengths, and multivariate calibration models relate the spectra to protein, moisture, oil and starch content. Capacitance moisture meters measure the dielectric constant of a grain sample, which rises with water content, and are calibrated per species and temperature. The falling number apparatus gelatinises a ground grain suspension in a boiling water bath and times the fall of a stirrer through the paste, giving a viscosity-related index of sprout damage and alpha-amylase activity. Gluten content is determined by washing a dough ball mechanically to remove starch, or by near-infrared prediction, and gluten index measures its elasticity. Test weight, or hectolitre weight, is the mass of grain filling a standard volume, measured by a filling hopper and a calibrated container.
What are the key features?
- Boiling water bath with precise temperature control for repeatable results.
- Timing to 0.1 s gives an objective alpha-amylase index.
- Multiple tube positions allow several samples per run.
- Automatic dispenser and stirring reduce operator variability.
- Digital logging stores results against sample identifiers.
- Simple, robust chemistry needs only grain, water and correct grinding.
What are the technical specifications?
A typical falling number apparatus configuration offers measurement of Falling number, seconds, measurement range of 60–1000 s, sample mass of 7.0 g (14% moisture basis), water volume of 25 mL, bath temperature of 100 °C, ±0.2 °C and timing accuracy of ±0.1 s. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Falling Number Apparatus is used in wheat sprout damage assessment, flour mill blending decisions, grain trade specification testing and malting barley quality screening. Across the wider Grain Analysis class, laboratories also deploy it for grain elevator intake grading, flour and semolina milling control and feed ingredient quality control.
Which standards and certifications apply?
Relevant standards and methods include ISO 712 (moisture), ISO 3093 (falling number), AACC Method 56-81B, ICC Standard 105/2, ISO 21415-2 (gluten) and GB/T 5498 (test weight). Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Keep near-infrared instruments clean and free of dust, and run the daily check sample supplied with the calibration to detect drift. Store reference samples for the grain species in use and send periodic samples to a reference laboratory for biased calibration updates. For capacitance meters, control sample temperature and fill volume exactly, since both affect readings. The falling number apparatus needs a boiling water bath at a stable rolling boil, correct grinding and accurate timing to meet the standard. Grain samples should be representative, taken with a proper divider, and stored dry to avoid moisture change before analysis. Record instrument settings, sample mass and results so that data can be traced to the method used.
How do I select the right Falling Number Apparatus?
Select the falling number apparatus when sprout damage or alpha-amylase activity must be quantified for milling or trade. A rapid visco analyzer or a sedimentation test may substitute in some specifications. In an RFQ, specify the tube capacity, automation level, needed standards and grinding equipment.
How does it compare with other Grain Analysis variants?
Start from the parameters and the reference method required by the trading contract or the internal specification. For rapid multi-parameter grading at intake, near-infrared analyzers give protein, moisture, oil and starch in under a minute and suit high throughput, but their calibrations must cover the grain species, variety groups and moisture ranges encountered. For legally reported moisture, use a capacitance or oven method calibrated to the applicable standard. Select the falling number apparatus when sprout damage or alpha-amylase activity matters for milling and baking. Choose a gluten analyzer when dough strength is the key quality parameter for flour. Confirm sample presentation, calibration maintenance, reference laboratory support and the ability to update models as new varieties are grown.
Full specifications
| Measurement | Falling number, seconds |
|---|---|
| Measurement range | 60–1000 s |
| Sample mass | 7.0 g (14% moisture basis) |
| Water volume | 25 mL |
| Bath temperature | 100 °C, ±0.2 °C |
| Timing accuracy | ±0.1 s |
| Tube positions | 1–6, model dependent |
| Repeatability | Within 5% between replicates |
| Grinding | Hammer mill, 0.8 mm sieve |
| Operating temperature | 10–35 °C ambient |
| Power supply | 220 VAC, 50 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Wheat sprout damage assessment
- Flour mill blending decisions
- Grain trade specification testing
- Malting barley quality screening
Key features
- Boiling water bath with precise temperature control for repeatable results.
- Timing to 0.1 s gives an objective alpha-amylase index.
- Multiple tube positions allow several samples per run.
- Automatic dispenser and stirring reduce operator variability.
- Digital logging stores results against sample identifiers.
- Simple, robust chemistry needs only grain, water and correct grinding.
Frequently asked questions about Falling Number Apparatus
Values above about 300 seconds generally indicate sound, un-sprouted wheat suitable for most bread baking, while values below 250 seconds suggest sprout damage and declining baking quality. Trade specifications set their own thresholds.
Particle size controls how quickly starch gelatinises and how the paste thickens, so the standard specifies a hammer mill with a set sieve. Different grinding shifts the result even when the same sample is used.
The bath must maintain a rolling boil at 100 °C, which depends on atmospheric pressure and water quality. Instruments include heaters with precise control and often display temperature, and distilled or deionised water is recommended to limit scale.
The falling number is the time in seconds for a stirrer to fall through a heated flour and water paste. A low value indicates high alpha-amylase activity and sprout damage, which reduces bread quality, while a high value indicates sound, un-sprouted grain.
The dielectric constant depends on grain structure and seed coat as well as water content, so wheat, maize, rice and oilseeds respond differently. Instruments therefore hold separate calibrations per species and often per variety group, and readings should be checked against an oven method.
With a calibration built for the local varieties and reference laboratory data, near-infrared protein predictions are typically within about 0.3 to 0.5 percent absolute of the reference method. Accuracy falls for varieties not represented in the calibration set.