Fluorescence Lateral Flow Reader
荧光侧向层析读数仪
- SKU FLI-0003
- Fluorescence Immunoassay
- CE (to be confirmed)
- ISO 13485 manufacturing (to be confirmed)
- NMPA class II registration (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Reading principle | Fluorescence scanning of test and control lines |
|---|---|
| Excitation wavelength | Typically 470–530 nm |
| Emission wavelength | Typically 520–610 nm |
| Measuring range | Assay specific, 2 to 3 orders of magnitude |
| Precision | CV 5–15 % across the measuring range |
| Reading time | 10–15 min after sample application |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 15189, CLSI EP05, CLSI EP15
What is the Fluorescence Lateral Flow Reader?
A fluorescence lateral flow reader scans the test and control lines of a fluorescent immunochromatographic strip and converts the signal into a quantitative or semi-quantitative result. It suits point-of-care settings that need quantitative results from a strip format in ten to fifteen minutes.
What is the Fluorescence Lateral Flow Reader?
A fluorescence lateral flow reader scans the test and control lines of a fluorescent immunochromatographic strip and converts the signal into a quantitative or semi-quantitative result. It suits point-of-care settings that need quantitative results from a strip format in ten to fifteen minutes.
Overview
The strip contains fluorescently labelled antibody conjugates that migrate with the sample and are captured on test and control lines. The reader inserts the strip into a scanning position, excites the lines with an LED or laser and measures emission, calculating the ratio of test to control signal to compensate for variations in sample flow. Quantitative results require strict control of sample volume, migration time and reading time, which is why most systems pair the reader with a fixed-volume sampler or an integrated cartridge. Calibration curves are stored per reagent lot and identified from the cartridge or a code chip. Typical precision is a coefficient of variation of five to fifteen percent, and results are available in ten to fifteen minutes. Readers usually store several thousand results and can transmit them through USB or a network interface.
How does it work?
Fluorescence immunoassay uses an antibody labelled with a fluorophore such as fluorescein, a europium chelate or a quantum dot. After the immune complex forms, the label is excited by light of one wavelength and emits light at a longer wavelength, and a photodetector measures the emission after optical filtering. The difference between excitation and emission wavelengths allows the instrument to separate the signal from the excitation light, and background fluorescence from sample components is reduced by careful filter selection, by washing away unbound label and, in time-resolved systems, by delaying the measurement until short-lived background fluorescence has decayed. Since the fluorescent signal is measured against a low background, sensitivity is generally better than absorbance-based enzyme immunoassay. Quantification relies on a calibration curve stored in the analyzer, and strip-based tests use a reader that scans the test and control lines.
What are the key features?
- LED or laser scanning of test and control lines
- Test-to-control ratio calculation to compensate flow variation
- Lot-specific curves identified from cartridge code or chip
- Fixed-volume sampler or integrated cartridge for quantitation
- Results in ten to fifteen minutes
- Storage of several thousand results with export options
What are the technical specifications?
A typical fluorescence lateral flow reader configuration offers reading principle of Fluorescence scanning of test and control lines, excitation wavelength of Typically 470–530 nm, emission wavelength of Typically 520–610 nm, measuring range of Assay specific, 2 to 3 orders of magnitude, precision of CV 5–15 % across the measuring range and reading time of 10–15 min after sample application. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Fluorescence Lateral Flow Reader is used in point-of-care inflammation testing, cardiac marker screening in clinics, infectious disease antigen testing and field and community screening programmes. Across the wider Fluorescence Immunoassay class, laboratories also deploy it for point-of-care cardiac and inflammation testing, hormone and fertility testing in clinics and infectious disease antigen detection.
Which standards and certifications apply?
Relevant standards and methods include ISO 15189, CLSI EP05, CLSI EP15, CLSI EP17, CLSI POCT04, IEC 61010-1 and IEC 61010-2-101. Certification candidates for this model are CE, ISO 13485 manufacturing and NMPA class II registration; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Keep the instrument away from strong ambient light and place it on a stable bench, because stray light and vibration reduce fluorescence measurement precision. Verify performance at installation with the manufacturer's calibration material and with controls at two or three levels, and check the optical path and detector gain as specified. Daily tasks include control measurement and checking that the correct assay lot and curve are selected. Lateral flow readers require consistent sample application volume, controlled reading time after the reaction and careful handling of cartridges, since a delay of a few minutes can change the result near a cut-off. Store cartridges and reagents within their temperature range, use them before expiry, and treat all used devices as infectious waste.
How do I select the right Fluorescence Lateral Flow Reader?
Specify a lateral flow reader when the workflow requires strip-based testing with a quantitative result and the menu covers your clinical need. In the RFQ confirm the reading window after sample application, the calibration mechanism, the sampler accuracy, and whether the strip supplier and the reader supplier provide a validated combination, since mixing suppliers can invalidate the calibration.
How does it compare with other Fluorescence Immunoassay variants?
Choose the format from the clinical setting: a strip or cartridge reader suits single tests at the point of care, while a plate or automated platform suits batch testing in a laboratory. Then check the assay menu and its regulatory status, because fluorescence platforms are kit specific and an instrument without the tests you need has no value. Compare sensitivity at the clinical decision point, measuring range, sample volume and time to result, which ranges from ten minutes for lateral flow to several hours for batch formats. For quantitative work, confirm that the analyzer stores calibration curves per lot and applies them automatically by reading a cartridge barcode. Consider whether fluorescence interference from bilirubin, haemoglobin or medication is documented for each assay, and whether controls and calibration material are available locally.
Full specifications
| Reading principle | Fluorescence scanning of test and control lines |
|---|---|
| Excitation wavelength | Typically 470–530 nm |
| Emission wavelength | Typically 520–610 nm |
| Measuring range | Assay specific, 2 to 3 orders of magnitude |
| Precision | CV 5–15 % across the measuring range |
| Reading time | 10–15 min after sample application |
| Sample volume | 5–100 µL by fixed-volume sampler |
| Data storage | 5,000 results with operator identification |
| Interfaces | USB, LAN, optional Bluetooth printer |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Point-of-care inflammation testing
- Cardiac marker screening in clinics
- Infectious disease antigen testing
- Field and community screening programmes
Key features
- LED or laser scanning of test and control lines
- Test-to-control ratio calculation to compensate flow variation
- Lot-specific curves identified from cartridge code or chip
- Fixed-volume sampler or integrated cartridge for quantitation
- Results in ten to fifteen minutes
- Storage of several thousand results with export options
Frequently asked questions about Fluorescence Lateral Flow Reader
The fluorescent signal on the test line continues to change as the sample migrates and the reaction proceeds, so results depend strongly on when the strip is read. Reading too early or too late shifts the value near a cut-off, which is why fixed reading windows and operator training are essential.
Only if the reader supports the required wavelengths and the strip supplier provides a validated calibration for that reader. In practice most manufacturers validate specific strip and reader combinations, and using unmatched components makes the laboratory responsible for full revalidation.
Most assays accept serum, plasma, whole blood or a buffer-extracted swab sample, depending on the target analyte. Viscous or particulate samples can block migration, so the instructions for use specify accepted matrices and dilution steps that must be followed exactly.
Run the manufacturer's control strips or a control cartridge at the start of a testing session and after a change of reagent lot, and record the values with the run. Readers should also be checked for optical cleanliness and, where available, with an internal reference strip.
Fluorescent labels emit against a low background, so assays can detect lower concentrations than absorbance-based methods and often cover a wider measuring range. The trade-off is greater sensitivity to interference from sample fluorescence and to variations in optical alignment, which is why controls matter.
Time-resolved systems use lanthanide chelates with a long fluorescence lifetime, and the instrument waits a short interval after excitation before measuring. Short-lived background fluorescence from serum components decays during that delay, so the measured signal contains less interference and allows lower detection limits.
Yes, quantitative readers scan the intensity of the test line and compare it with a stored calibration curve, provided the sample volume and reading time are controlled. Precision is lower than laboratory platforms, so quantitative strip tests are usually validated against a laboratory method before clinical use.