Basic Isocratic Flash Chromatography System
基础等度快速制备色谱系统
- SKU FLC-0001
- Flash Chromatography
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Flow rate range | 5–100 mL/min |
|---|---|
| Maximum operating pressure | 7 bar (100 psi) |
| Solvent channels | 1 (isocratic) |
| Detector | UV 254 nm fixed or 200–400 nm variable |
| Cartridge size range | 4–120 g silica |
| Sample loading capacity | up to 1 g crude per 10 g silica |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, ATEX 2014/34/EU, ISO 9001
What is the Basic Isocratic Flash Chromatography System?
A basic isocratic flash chromatography system pumps a single premixed solvent through a silica cartridge with a fixed-wavelength UV detector and simple fraction collection. It is the lowest-cost automated option for academic and teaching laboratories running routine normal-phase purifications of well-separated compounds.
What is the Basic Isocratic Flash Chromatography System?
A basic isocratic flash chromatography system pumps a single premixed solvent through a silica cartridge with a fixed-wavelength UV detector and simple fraction collection. It is the lowest-cost automated option for academic and teaching laboratories running routine normal-phase purifications of well-separated compounds.
Overview
This entry-level configuration uses a single-channel pump delivering 5–100 mL/min at up to 7 bar, a fixed-wavelength UV detector at 254 nm or a variable UV detector, and a basic rack-based fraction collector with 20–100 tubes. The operator prepares the mobile phase in advance, usually from a TLC solvent system giving a product Rf of 0.2–0.35, and runs it isocratically. There is no solvent blending, so solvents must be premixed and changed manually. The system suits compounds with large Rf differences, teaching of purification practice and high-volume simple separations. It does not perform automatic gradient programming, reversed-phase work or mass-triggered collection. Cartridges from 4 g to 120 g are supported through standard Luer or pressure-rated fittings. Compared with manual glass column work, it saves time and reduces silica waste because fractions are collected by UV detection rather than by eye.
How does it work?
Flash chromatography forces solvent through a packed cartridge at moderate pressure, typically 1–20 bar, instead of relying on gravity. Cartridges contain 4 g to over 1 kg of irregular or spherical silica of 20–60 µm, or bonded phases such as C18, amino and cyano. A binary pump blends two solvents, for example hexane and ethyl acetate for normal phase or water and acetonitrile for reversed phase, into linear or step gradients. Dissolved or dry-loaded samples are applied at the column head, and compounds elute according to polarity. A UV or diode-array detector monitors 200–400 nm, and an evaporative light scattering detector (ELSD) detects compounds lacking a chromophore. A fraction collector triggers by threshold or slope and deposits fractions into tubes or flasks, with software showing a real-time chromatogram. Elution can be compared to TLC Rf values to predict gradients.
What are the key features?
- Single-channel pump delivering 5–100 mL/min at up to 7 bar for routine cartridges
- Fixed-wavelength 254 nm or variable-wavelength UV detector with threshold fraction triggering
- Rack-based fraction collector accepting 16–25 mm tubes, 20–100 positions
- Standard Luer-lock cartridge interface supporting 4–120 g silica cartridges
- Simple touch-screen or button interface requiring minimal training
What are the technical specifications?
A typical basic isocratic flash chromatography system configuration offers flow rate range of 5–100 mL/min, maximum operating pressure of 7 bar (100 psi), solvent channels of 1 (isocratic), detector of UV 254 nm fixed or 200–400 nm variable, cartridge size range of 4–120 g silica and sample loading capacity of up to 1 g crude per 10 g silica. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Basic Isocratic Flash Chromatography System is used in teaching laboratory purification, routine normal-phase cleanup of simple reaction mixtures, academic synthesis groups and high-volume simple separations. Across the wider Flash Chromatography class, laboratories also deploy it for purification of synthetic reaction products, isolation of natural product fractions and purification of pharmaceutical intermediates.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, ATEX 2014/34/EU, ISO 9001, USP <621>, ICH Q3C, EN 60079-0, GB/T 15481 and ISO/IEC 17025. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install in a fume hood or ventilated enclosure because normal-phase solvents such as hexane, ethyl acetate and dichloromethane are flammable or hazardous, and keep ignition sources away. Use grounded tubing and containers to avoid static buildup. Flush lines with compatible solvent before changing between normal and reversed phase, using miscible intermediates such as isopropanol. Inspect pump seals, check valves and cartridge holders regularly, and do not exceed cartridge pressure ratings, often 10–20 bar. Calibrate detector baselines with blank gradients and replace UV lamps as recommended. Dispose of solvent waste and used silica following local regulations. Flash systems are not normally subject to GMP qualification, but record methods and audit results for traceability.
How do I select the right Basic Isocratic Flash Chromatography System?
Select the isocratic system for simple separations with clear TLC resolution and limited budget. If compounds run close together or require gradients, choose a gradient model. In an RFQ specify flow range, maximum cartridge size, UV wavelength options, fraction collector capacity and solvent compatibility.
How does it compare with other Flash Chromatography variants?
Choose by scale, detection and phase. Cartridge size determines loading: roughly 1–10% of silica mass for easy separations, so a 40 g cartridge handles 0.4–4 g of crude. Systems with UV only suit chromophoric compounds, while adding ELSD or a mass detector is required for sugars, lipids and many natural products. Reversed-phase capable systems need pumps and seals compatible with water and higher backpressure, up to 20 bar or more. Consider maximum flow rate, typically 100–300 mL/min for mid-scale and up to 1000 mL/min for kilogram-scale work, and the number of fractions and tube sizes for the collector. Solvent safety matters, so check ATEX-rated designs, closed fraction bays and waste management. Software features such as TLC-to-gradient conversion and method libraries speed routine purification.
Full specifications
| Flow rate range | 5–100 mL/min |
|---|---|
| Maximum operating pressure | 7 bar (100 psi) |
| Solvent channels | 1 (isocratic) |
| Detector | UV 254 nm fixed or 200–400 nm variable |
| Cartridge size range | 4–120 g silica |
| Sample loading capacity | up to 1 g crude per 10 g silica |
| Fraction collector capacity | 20–100 tubes |
| Tube size | 13–25 mm |
| Flow accuracy | ±5% |
| Dimensions (W×D×H) | approx. 400 × 450 × 450 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Teaching laboratory purification
- Routine normal-phase cleanup of simple reaction mixtures
- Academic synthesis groups
- High-volume simple separations
Key features
- Single-channel pump delivering 5–100 mL/min at up to 7 bar for routine cartridges
- Fixed-wavelength 254 nm or variable-wavelength UV detector with threshold fraction triggering
- Rack-based fraction collector accepting 16–25 mm tubes, 20–100 positions
- Standard Luer-lock cartridge interface supporting 4–120 g silica cartridges
- Simple touch-screen or button interface requiring minimal training
Frequently asked questions about Basic Isocratic Flash Chromatography System
Use TLC to find a solvent mixture that gives the target compound an Rf of 0.2–0.35 and at least 0.1 Rf difference from nearby spots. Run the same composition isocratically on the cartridge. Expect the compound to elute at about 3–5 column volumes.
Not directly. A UV detector misses compounds without chromophores, so you would collect fractions by time and check them by TLC with staining. For such compounds an ELSD-equipped system is recommended.
It reduces labor and the silica and solvent consumed per purification because fractions are collected accurately. Prepacked cartridges cost more than bulk silica but are consistent and save packing time. Overall savings depend on throughput of the laboratory.
Flash chromatography uses larger particles of 20–60 µm at pressures below about 20 bar to purify grams of material quickly at low cost. Preparative HPLC uses 5–10 µm particles at 100–400 bar to achieve higher resolution for difficult separations and higher purity, but with lower loading and greater cost.
As a guide, load 1–10% of the silica mass. Easy separations with large Rf differences tolerate up to 10%, while close-running compounds need 1% or less. A 40 g silica cartridge therefore takes roughly 0.4–4 g crude. Dry loading onto silica improves resolution for poorly soluble samples.
Use ELSD when target compounds lack a UV chromophore, such as sugars, lipids, saponins, terpenoids and many polymers. It responds to any non-volatile analyte in a volatile mobile phase, but is destructive and requires a splitter to divert a small portion of flow.