Flame Emission and Atomic Absorption Spectrometer
火焰发射与原子吸收两用光谱仪
- SKU FLA-0005
- Flame AAS
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety compliance (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Wavelength range | 185–900 nm |
|---|---|
| Modes | flame emission and atomic absorption |
| Emission elements | Na, K, Li, Ca, Ba, Sr |
| Detection limit (Na emission) | 0.002 mg/L |
| Flame types | air-acetylene |
| Monochromator | Czerny-Turner |
By Lab Manufacturers China Technical TeamLast reviewed Standards: Ph. Eur. 2.2.22 Atomic absorption spectrometry, USP <852> Atomic Absorption Spectroscopy, ISO 8288 Water quality — Flame atomic absorption determination of metals
What is the Flame Emission and Atomic Absorption Spectrometer?
This instrument combines flame emission and atomic absorption modes in one spectrometer. It is chosen where alkali and alkaline earth metals are measured frequently, since emission mode avoids hollow-cathode lamps for these elements.
What is the Flame Emission and Atomic Absorption Spectrometer?
This instrument combines flame emission and atomic absorption modes in one spectrometer. It is chosen where alkali and alkaline earth metals are measured frequently, since emission mode avoids hollow-cathode lamps for these elements.
Overview
Flame emission photometry measures the light emitted by thermally excited atoms in the flame, while atomic absorption measures light absorbed from a lamp. Elements with low excitation energies, such as sodium, potassium, lithium, calcium and barium, emit strongly in a flame and can be measured by emission with good sensitivity and no lamp cost. The instrument selects the mode through software, switching the detector from absorption to emission measurement. Emission mode avoids spectral interference from the lamp and is less affected by lamp drift, but it is more sensitive to flame temperature and composition, so gas control must be stable. This dual capability suits clinical, agricultural, food and water laboratories that measure electrolytes and nutrients alongside trace metals, and it reduces the number of lamps that must be stocked.
How does it work?
A peristaltic pump or the nebulizer vacuum draws sample into a nebulizer, where it forms an aerosol that mixes with fuel and oxidant gases in a burner head. Combustion produces a steady cloud of ground-state atoms. A hollow-cathode lamp specific to the element emits sharp resonance lines that pass through the flame to a monochromator and photodetector. Absorbance is proportional to concentration according to the Beer-Lambert law. The flame composition, burner slot geometry and observation height are chosen so that atomization is efficient and chemical interferences are limited. Deuterium or Zeeman background correction removes signals from smoke, droplets and molecular species. Calibration is performed with aqueous standards matched to the sample acid and matrix, and burner rotation or a flow spoiler extends the linear range for concentrated samples.
What are the key features?
- Switchable flame emission and atomic absorption modes
- No hollow-cathode lamp needed for alkali metals
- Suitable for sodium, potassium, lithium, calcium and barium
- Stable gas control for reproducible flame temperature
- Multi-lamp turret for conventional elements
- Mode selection integrated in method software
What are the technical specifications?
A typical flame emission and atomic absorption spectrometer configuration offers wavelength range of 185–900 nm, modes of flame emission and atomic absorption, emission elements of Na, K, Li, Ca, Ba, Sr, detection limit (Na emission) of 0.002 mg/L, flame types of air-acetylene and monochromator of Czerny-Turner. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Flame Emission and Atomic Absorption Spectrometer is used in serum and urine electrolyte measurement, soil extractable potassium and calcium, food and beverage sodium and potassium and water hardness and alkali metal analysis. Across the wider Flame AAS class, laboratories also deploy it for metals in drinking water and wastewater, soil, plant and fertilizer nutrient analysis and food and beverage mineral and toxic metal testing.
Which standards and certifications apply?
Relevant standards and methods include Ph. Eur. 2.2.22 Atomic absorption spectrometry, USP <852> Atomic Absorption Spectroscopy, ISO 8288 Water quality — Flame atomic absorption determination of metals, US EPA Method 7000B, ASTM D4691 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety compliance; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Flame atomic absorption requires a flame hood, secure gas cylinders, regulators and flashback arrestors, with acetylene kept away from oxidizing materials. Install the instrument on a stable bench away from strong drafts, because airflow disturbs the flame. Daily checks include gas pressure, burner cleanliness and calibration verification. The burner slot and nebulizer need periodic cleaning to remove salt deposits that cause noise and drift. Hollow-cathode lamps have limited life and may need warm-up and alignment. Sample solutions should be filtered and matched to standards in acid content, and method blanks track contamination. Keep service and lamp records for traceability.
How do I select the right Flame Emission and Atomic Absorption Spectrometer?
Select a dual-mode instrument when alkali and alkaline earth metals are a significant part of the workload and you want to avoid stocking and aligning many lamps. Confirm that emission sensitivity meets your method requirements and that gas control is stable enough for reproducible flame temperature. Specify the element list and whether background correction is still required for the absorption elements you measure.
How does it compare with other Flame AAS variants?
Select a flame system by the element list, concentration range and matrix. Single-beam optics suit most routine work at lower cost, while double-beam optics reduce drift from lamp and detector fluctuations for long unattended runs. An autosampler is essential for throughput and reproducibility in quality-control laboratories. If refractory elements such as aluminum, titanium or vanadium are required, choose a system with a nitrous oxide-acetylene burner and appropriate safety provisions. Alkali metals can be measured in flame emission mode, which improves sensitivity and avoids hollow-cathode lamp cost. Confirm burner materials suit your acid and dissolved-solids levels, and check that the gas control and interlocks meet local regulations.
Full specifications
| Wavelength range | 185–900 nm |
|---|---|
| Modes | flame emission and atomic absorption |
| Emission elements | Na, K, Li, Ca, Ba, Sr |
| Detection limit (Na emission) | 0.002 mg/L |
| Flame types | air-acetylene |
| Monochromator | Czerny-Turner |
| Spectral bandwidth | 0.1–2.0 nm selectable |
| Lamp turret | 4–6 hollow-cathode lamps |
| Gas control | automatic with pressure feedback |
| Interface | USB / RS-232 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Serum and urine electrolyte measurement
- Soil extractable potassium and calcium
- Food and beverage sodium and potassium
- Water hardness and alkali metal analysis
Key features
- Switchable flame emission and atomic absorption modes
- No hollow-cathode lamp needed for alkali metals
- Suitable for sodium, potassium, lithium, calcium and barium
- Stable gas control for reproducible flame temperature
- Multi-lamp turret for conventional elements
- Mode selection integrated in method software
Frequently asked questions about Flame Emission and Atomic Absorption Spectrometer
Flame emission suits elements with low excitation energy that emit strongly, including sodium, potassium, lithium, calcium, barium, strontium and rubidium. These are measured without a hollow-cathode lamp. Elements with high excitation energy are better measured by atomic absorption because their emission is weak.
Emission intensity depends strongly on flame temperature and the number of excited atoms, which follow a Boltzmann distribution. Small changes in fuel-oxidant ratio or nebulizer flow shift this population and the signal. Stable gas control, a warm-up period and frequent calibration therefore matter more in emission mode.
Emission measurements still see flame background and continuum emission, so a background measurement at an adjacent wavelength is subtracted. This is usually handled by the software. For high-sensitivity work, choice of wavelength and flame conditions controls background more effectively than correction alone.
Flame atomic absorption usually measures metals from about 0.001 to 100 milligrams per litre, depending on the element and flame conditions. Copper, zinc, iron and manganese are comfortably in this range. Below about 0.01 milligrams per litre, graphite furnace atomic absorption or ICP techniques offer lower detection limits with better reliability.
The flame produces heat, metal vapors and combustion products, and samples may contain acids and toxic elements. A flame hood removes these emissions and protects operators. It also stabilizes the environment around the burner so the flame is not disturbed by room air movement, which improves measurement reproducibility.
Use nitrous oxide-acetylene for refractory elements that form stable oxides in the cooler air-acetylene flame, such as aluminum, titanium, vanadium, silicon and molybdenum. The hotter flame improves atomization, but it also increases chemical interference risk, requires special burner heads and must be handled with additional safety precautions.
Related models




Nitrous Oxide-Acetylene Flame AAS for Refractory Elements
难熔元素一氧化二氮-乙炔火焰原子吸收光谱仪
- CE (to be confirmed)
- +2