Skip to main content
Lab Manufacturers China
Flame Photometer Electrolyte Analyzer — product photo

Flame Photometer Electrolyte Analyzer

火焰光度法电解质分析仪

  • SKU EAN-0005
  • Electrolyte Analyzers
  • CE (to be confirmed)
  • ISO 13485 manufacturing (to be confirmed)
  • Gas appliance safety approval (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Measured analytesNa+, K+ (Li+ in some models)
Measuring range Na+80–200 mmol/L
Measuring range K+1.0–10.0 mmol/L
Sample volume10–50 µL diluted 1:50 to 1:200
Throughput60–100 samples/h
Gas supplyLPG, butane or natural gas at regulated pressure
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 15189, CLSI EP05, CLSI EP15

What is the Flame Photometer Electrolyte Analyzer?

A flame photometer measures sodium and potassium by the intensity of light emitted when a diluted sample is aspirated into a flame. It remains in use where the classical method is required, and it provides a low-cost alternative in laboratories that do not need ionised calcium or whole blood testing.

What is the Flame Photometer Electrolyte Analyzer?

A flame photometer measures sodium and potassium by the intensity of light emitted when a diluted sample is aspirated into a flame. It remains in use where the classical method is required, and it provides a low-cost alternative in laboratories that do not need ionised calcium or whole blood testing.

Overview

In flame photometry the sample is diluted with an internal lithium or caesium standard and aspirated into a propane, butane or natural gas flame. Sodium and potassium atoms are excited and emit light at characteristic wavelengths, 589 and 766 nanometers respectively, and a photodetector measures the emission intensity, which is proportional to concentration over the working range. The internal standard corrects for variations in aspiration rate and flame temperature. Results are available within seconds of aspiration, and calibration uses two or three aqueous standards. The method requires gas supply, ventilation of the flame compartment and annual gas safety checks; it reports total sodium and potassium, not ionised fractions, and cannot accept whole blood because the flame must be fed a diluted, protein-free solution.

How does it work?

An ion-selective electrode consists of a membrane that responds selectively to one ion, an internal filling solution and a reference electrode. When the sample contacts the membrane, a potential develops across it that follows the Nernst equation, changing by about 59 millivolts for a tenfold change in ion activity at 25 degrees Celsius. The analyzer measures this potential against the reference electrode, corrects for temperature, and calculates concentration from a two-point calibration with standard solutions of known activity. Direct measurement presents undiluted whole blood, plasma, serum or urine to the electrode, which is preferred for whole blood samples and for patients with abnormal protein or lipid content. Indirect measurement dilutes the sample with a reagent before the electrode, which suits diluted samples and automated high-throughput operation. Sodium, potassium, chloride, calcium, lithium and magnesium electrodes differ in membrane composition.

What are the key features?

  • Flame excitation of sodium and potassium with emission detection
  • Internal lithium or caesium standard for aspiration correction
  • Result within seconds of sample aspiration
  • Two or three point calibration with aqueous standards
  • Low reagent cost using diluted samples
  • Requires gas supply and flame compartment ventilation

What are the technical specifications?

A typical flame photometer electrolyte analyzer configuration offers measured analytes of Na+, K+ (Li+ in some models), measuring range Na+ of 80–200 mmol/L, measuring range K+ of 1.0–10.0 mmol/L, sample volume of 10–50 µL diluted 1:50 to 1:200, throughput of 60–100 samples/h and gas supply of LPG, butane or natural gas at regulated pressure. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Flame Photometer Electrolyte Analyzer is used in serum and urine sodium and potassium testing, plant and agricultural sample analysis, teaching laboratories demonstrating emission spectroscopy and low-cost routine electrolyte screening. Across the wider Electrolyte Analyzers class, laboratories also deploy it for critical care and emergency electrolyte testing, dialysis and renal unit monitoring and surgical and perioperative assessment.

Which standards and certifications apply?

Relevant standards and methods include ISO 15189, CLSI EP05, CLSI EP15, CLSI C29, IEC 61010-1 and IEC 61010-2-101. Certification candidates for this model are CE, ISO 13485 manufacturing and Gas appliance safety approval; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Install the analyzer on a level bench away from direct sunlight and heat sources, with a grounded power supply and enough space for reagent and waste containers. Verify performance at installation with aqueous standards and with split serum samples compared against another method, and check the slope of each electrode, which should fall within the expected Nernstian range. Daily tasks include calibration with the supplied standards, measurement of normal and abnormal control material, and inspection of the sample path for air bubbles, clots or protein deposits. Electrodes and tubing are replaced when slope, response time or precision deteriorates. Treat all blood samples as infectious, dispose of waste according to local rules, and record every maintenance action for accreditation.

How do I select the right Flame Photometer Electrolyte Analyzer?

Choose a flame photometer only when the classical method is required, when samples are serum or urine only, or when reagent cost and simplicity outweigh convenience. In the RFQ confirm the gas type and pressure, the exhaust requirement, the dilution protocol, and the calibration frequency. Plan for gas safety inspection and confirm that ionised calcium or whole blood testing is not needed.

How does it compare with other Electrolyte Analyzers variants?

Choose an electrolyte analyzer when the electrolyte request volume justifies a dedicated instrument, typically more than thirty tests per day, or when whole blood electrolytes must be available within minutes. Decide first between direct and indirect measurement, since they give different reference intervals and cannot be combined in one report. Then check whether you need whole blood capability, urine mode, or additional analytes such as ionised calcium, lithium or magnesium. Compare calibration frequency and electrode life, because these drive running cost, and the sample volume, which matters for pediatric and neonatal work. Confirm that the instrument reports temperature-corrected results, that it has a clot detection or blockage alarm, and that results can be transmitted to the laboratory information system.

Full specifications

Full specifications
Measured analytesNa+, K+ (Li+ in some models)
Measuring range Na+80–200 mmol/L
Measuring range K+1.0–10.0 mmol/L
Sample volume10–50 µL diluted 1:50 to 1:200
Throughput60–100 samples/h
Gas supplyLPG, butane or natural gas at regulated pressure
Internal standardLithium or caesium solution
Safety featuresFlame failure cut-off and exhaust connection
Power supplyAC 220–240 V, 50/60 Hz

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Serum and urine sodium and potassium testing
  • Plant and agricultural sample analysis
  • Teaching laboratories demonstrating emission spectroscopy
  • Low-cost routine electrolyte screening

Key features

  • Flame excitation of sodium and potassium with emission detection
  • Internal lithium or caesium standard for aspiration correction
  • Result within seconds of sample aspiration
  • Two or three point calibration with aqueous standards
  • Low reagent cost using diluted samples
  • Requires gas supply and flame compartment ventilation

Frequently asked questions about Flame Photometer Electrolyte Analyzer

No. Flame photometry reports total sodium and potassium, and ionised calcium requires an ion-selective electrode with a calcium membrane. Laboratories that need ionised calcium or blood gas parameters should choose an ISE or blood gas analyzer instead.

Request a quotation for the Flame Photometer Electrolyte Analyzer

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote