High-Temperature Differential Thermal Analyzer
高温差热分析仪
- SKU DTA-0002
- DTA
- CE (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | Ambient–1600 °C |
|---|---|
| Heating rate | 1–30 °C/min |
| Thermocouple type | Type S, Type B or tungsten-rhenium |
| Temperature accuracy | ±3 °C |
| Signal resolution | 0.5 °C |
| Sample mass | 50–200 mg |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E967, ASTM E794, ASTM D3418
What is the High-Temperature Differential Thermal Analyzer?
A high-temperature differential thermal analyzer reaches 1500 to 1600 degrees Celsius with platinum-rhodium or tungsten-rhenium thermocouples, making it suitable for cements, refractories, ceramics, glasses, slags and minerals whose reactions occur far above 700 degrees Celsius.
What is the High-Temperature Differential Thermal Analyzer?
A high-temperature differential thermal analyzer reaches 1500 to 1600 degrees Celsius with platinum-rhodium or tungsten-rhenium thermocouples, making it suitable for cements, refractories, ceramics, glasses, slags and minerals whose reactions occur far above 700 degrees Celsius.
Overview
Platinum-rhodium or tungsten-rhenium thermocouples and an alumina or platinum holder allow operation to 1500 or 1600 degrees Celsius in air, argon or nitrogen. The furnace uses silicon carbide or molybdenum disilicide elements, and above 1600 degrees Celsius a graphite furnace with a protective atmosphere is required. Peak positions are usually reported in air or inert gas because some reactions depend strongly on oxygen availability. Typical specifications include a temperature accuracy of about plus or minus 3 degrees Celsius, heating rates from 1 to 30 degrees Celsius per minute, and sample masses of 50 to 200 milligrams for weakly reacting inorganic materials. Applications include quartz inversion, carbonate decomposition, cement clinker formation, glass transition and crystallization, and the study of sintering additives and ceramic binders. Holding the sample in a shallow crucible and controlling packing are essential for reproducible peak shapes.
How does it work?
The sample and a thermally inert reference, usually calcined alumina, sit in the same furnace and are heated on the same program. A differential thermocouple or two matched thermocouples measure the temperature difference between them, recorded as microvolts against temperature or time, alongside the sample temperature. When the sample undergoes a transition that absorbs heat, its temperature lags the reference and produces an endothermic peak; exothermic reactions such as crystallization or oxidation give a peak in the opposite direction. The peak position identifies the transition temperature and the peak area is roughly proportional to the enthalpy change, although DTA is less quantitative than DSC because the heat path is not calibrated in absolute terms. Sensitivity depends on sample mass, packing density, thermocouple position and atmosphere. Typical samples are 10 to 100 milligrams, and inert, oxidising, reducing or vacuum atmospheres are selectable.
What are the key features?
- Furnace operation to 1500–1600 °C with silicon carbide or MoSi₂ elements
- Platinum-rhodium or tungsten-rhenium thermocouples matched to the range
- Alumina, platinum or graphite sample holders
- Atmosphere control with air, argon, nitrogen or vacuum options
- Sample masses of 50–200 mg for weak inorganic transitions
- Peak evaluation software with onset and area reporting
What are the technical specifications?
A typical high-temperature differential thermal analyzer configuration offers temperature range of Ambient–1600 °C, heating rate of 1–30 °C/min, thermocouple type of Type S, Type B or tungsten-rhenium, temperature accuracy of ±3 °C, signal resolution of 0.5 °C and sample mass of 50–200 mg. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The High-Temperature Differential Thermal Analyzer is used in cement clinker and gypsum studies, refractory and ceramic development, glass transition and crystallization, slag and fly ash characterisation and catalyst support phase transitions. Across the wider DTA class, laboratories also deploy it for clay and mineral phase identification, cement and gypsum dehydration studies and ceramic and refractory development.
Which standards and certifications apply?
Relevant standards and methods include ASTM E967, ASTM E794, ASTM D3418, ISO 11357-1, IEC 61010-1 and GB/T 13464. Certification candidates for this model are CE, IEC 61010-1 safety and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Package the sample uniformly and keep mass and packing the same as the reference to avoid baseline distortion, and dilute reactive samples with alumina when necessary. Select crucible and thermocouple materials that do not react with the sample at the working temperature, and use a platinum or alumina holder for oxidising conditions. Purge the furnace with the intended gas at a stable flow, and verify the thermocouple pair against melting point standards such as indium, tin, aluminium, silver or gold depending on range. Record the crucible type, sample mass, gas, flow rate and heating rate with every curve. Thermal analysis of minerals often shows overlapping peaks that require a consistent, documented heating rate and reference material to be compared between laboratories, so method details should be exchanged with results.
How do I select the right High-Temperature Differential Thermal Analyzer?
Choose the high-temperature configuration for inorganic materials whose reactions lie between 700 and 1600 degrees Celsius, typically cements and ceramics. If reactions occur above 1600 degrees Celsius, a graphite-furnace version is required. In an RFQ, state maximum temperature, thermocouple type, holder material and atmosphere.
How does it compare with other DTA variants?
The decisive parameter is the temperature range, because DTA is chosen mainly when transitions lie above the working range of a DSC: minerals, clays, cements, refractories and glasses often require 1500 to 1600 degrees Celsius, and some slags and alloys need graphite furnaces above 2000 degrees Celsius. Below 700 degrees Celsius a DSC is usually a better choice because it also gives enthalpies. After range, consider the thermocouple type and hence the maximum temperature, the crucible and thermocouple materials compatible with the sample, atmosphere and vacuum control, sensitivity to small transitions, and whether automation is required. Peak resolution depends on the heat path, so sample packing and mass must be matched between runs. For routine quality control, an automated instrument with a sample changer reduces operator time, and calibration with melting point and phase transition references should be part of the method.
Full specifications
| Temperature range | Ambient–1600 °C |
|---|---|
| Heating rate | 1–30 °C/min |
| Thermocouple type | Type S, Type B or tungsten-rhenium |
| Temperature accuracy | ±3 °C |
| Signal resolution | 0.5 °C |
| Sample mass | 50–200 mg |
| Sample holder | Alumina, platinum or graphite |
| Atmosphere | Air, argon, nitrogen, vacuum option |
| Furnace elements | Silicon carbide or MoSi₂ |
| Cooling | Natural or forced air, up to 30 °C/min |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Cement clinker and gypsum studies
- Refractory and ceramic development
- Glass transition and crystallization
- Slag and fly ash characterisation
- Catalyst support phase transitions
Key features
- Furnace operation to 1500–1600 °C with silicon carbide or MoSi₂ elements
- Platinum-rhodium or tungsten-rhenium thermocouples matched to the range
- Alumina, platinum or graphite sample holders
- Atmosphere control with air, argon, nitrogen or vacuum options
- Sample masses of 50–200 mg for weak inorganic transitions
- Peak evaluation software with onset and area reporting
Frequently asked questions about High-Temperature Differential Thermal Analyzer
Many high-temperature reactions such as oxidation, reduction and carbonate decomposition depend on the gas composition. Peak temperatures for the same material can differ substantially in air and in argon, so the atmosphere, flow rate and heating rate must be reported together with the curve.
Alumina suits most oxides and neutral conditions but can react with alkaline melts, platinum is inert in air but expensive and can alloy with some metals, and graphite is used in inert or reducing atmospheres. Chemical compatibility with the sample is the first criterion.
Weakly reacting inorganic materials such as refractories need 50 to 200 milligrams to produce detectable peaks, while strongly reacting samples such as carbonates can be run at 20 to 50 milligrams to avoid overflowing the crucible and distorting the baseline.
DTA records the temperature difference between sample and reference and is mainly qualitative, while DSC measures heat flow and can be calibrated to give transition enthalpies. DTA is usually preferred at very high temperatures where DSC cells are difficult to build.
Clay minerals show characteristic dehydration, dehydroxylation and recrystallization events at defined temperatures, giving a fingerprint pattern. Kaolinite, for example, loses hydroxyl groups near 500 degrees Celsius and recrystallizes near 980 degrees Celsius, which helps identify the clay type.
Temperature is calibrated with high-purity melting point standards spanning the working range, in contrast to DSC where enthalpies are also calibrated. Peak onset rather than peak maximum is commonly used as the transition temperature, and the same heating rate must be used for standards and samples.