High-Pressure Differential Scanning Calorimeter
高压差示扫描量热仪
- SKU DSC-0004
- DSC
- CE (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Pressure range | 0.1–15 MPa |
|---|---|
| Temperature range | 25–600 °C |
| Heating rate | 0.1–50 °C/min |
| Temperature accuracy | ±0.2 °C |
| Calorimetric accuracy | ±2% |
| Crucible type | Sealed stainless steel or gold-plated |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E967, ASTM E968, ASTM E1356
What is the High-Pressure Differential Scanning Calorimeter?
A high-pressure DSC measures heat flow in a sealed pressure cell, typically 1 to 15 MPa, so reactions, oxidation and phase behaviour can be studied at elevated pressure or with volatile samples retained. It is used for oils, lubricants, polymers and propellants.
What is the High-Pressure Differential Scanning Calorimeter?
A high-pressure DSC measures heat flow in a sealed pressure cell, typically 1 to 15 MPa, so reactions, oxidation and phase behaviour can be studied at elevated pressure or with volatile samples retained. It is used for oils, lubricants, polymers and propellants.
Overview
The measuring cell is enclosed in a pressure vessel with gas inlet and outlet, allowing measurements in nitrogen, oxygen, air or other gases at pressures from about 0.1 to 15 MPa. Sealed crucibles withstand the pressure and keep volatile samples inside, which is essential for oxidation induction time of lubricants and polymers, where oxygen is supplied at elevated partial pressure to accelerate aging. The technique also shifts boiling points and phase equilibria, so it is used to study vaporization, high-pressure crystallization, hydrates and reactions that evolve gas. Thermal response is slower than in an open cell because of the metal vessel mass, and purge flow and pressure control must be stable to keep baselines flat. Typical instruments cover room temperature to 600 degrees Celsius at pressures up to 10 or 15 MPa, with the pressure stated and logged for each run. Safety interlocks, pressure relief and correct crucible sealing are mandatory.
How does it work?
The sample and an empty reference pan sit on a sensor in a furnace whose temperature follows a program. In heat-flux instruments a single furnace heats both positions and thermocouples measure the small temperature difference that develops when the sample absorbs or releases heat; the difference is converted to heat flow using a calibration factor. Power-compensation instruments place the sample and reference in separate small furnaces and vary the heating power needed to keep both at the same temperature, recording power difference directly. Either way the result is a curve of heat flow against temperature or time, with the area under a peak giving transition enthalpy in joules per gram. A purge gas, usually nitrogen, prevents oxidation and removes volatiles, and cooling may use liquid nitrogen, a mechanical cooler or an air vortex. Temperature and enthalpy calibration use indium, zinc and other reference materials of known melting point and enthalpy.
What are the key features?
- Sealed pressure vessel for measurements from 0.1 to 15 MPa
- Gas supply with selectable oxygen, air, nitrogen or other gases
- Sealed crucibles that retain volatile and reactive samples
- Oxidation induction time measurement under elevated oxygen pressure
- Pressure and purge flow logging with each measurement
- Safety interlocks with pressure relief and burst protection
What are the technical specifications?
A typical high-pressure differential scanning calorimeter configuration offers pressure range of 0.1–15 MPa, temperature range of 25–600 °C, heating rate of 0.1–50 °C/min, temperature accuracy of ±0.2 °C, calorimetric accuracy of ±2% and crucible type of Sealed stainless steel or gold-plated. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The High-Pressure Differential Scanning Calorimeter is used in lubricant and oil oxidation stability, polymer oxidation induction time, propellant and energetic material studies, high-pressure phase behaviour and food oil and fat stability testing. Across the wider DSC class, laboratories also deploy it for polymer melting and crystallinity testing, pharmaceutical polymorph and purity screening and food starch and fat behaviour studies.
Which standards and certifications apply?
Relevant standards and methods include ASTM E967, ASTM E968, ASTM E1356, ASTM E1269, ISO 11357-1, ISO 11357-2 and USP <891>. Certification candidates for this model are CE, IEC 61010-1 safety and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Calibrate temperature and enthalpy before a measurement campaign, and verify at least one check standard afterwards. Choose the pan material and type for the temperature and chemistry involved: sealed or hermetic pans for volatile samples, high-pressure crucibles for pressure work and open pans for evolved gases. Keep sample mass similar between runs, because thermal lag grows with mass and shifts peaks to higher temperature. Use a defined purge gas flow, dry the cell if moisture condenses, and run an empty baseline periodically. Consumables such as pans, lids and seals are single-use, and crucibles used for pressure work must be inspected for deformation. Follow ASTM E967, E968 or ISO 11357 for calibration and reporting, and record pan type, gas, rate and mass with each curve.
How do I select the right High-Pressure Differential Scanning Calorimeter?
Select a high-pressure instrument when oxygen must be supplied under pressure, when samples are volatile or when pressure is a variable in the study. Routine melting and glass transition work does not need one. In an RFQ, state maximum pressure, gas types, crucible material, temperature range and the oxidation induction time method to be followed.
How does it compare with other DSC variants?
Start with the temperature range you need, because it decides the cooling system: studies above room temperature need only an air or mechanical cooler, while glass transitions well below zero require liquid nitrogen. Then choose the measuring principle. Heat-flux instruments are robust, accept larger samples and cover most materials work; power-compensation instruments give sharper peaks and better resolution for small transitions and fast heating rates; temperature-modulated instruments separate reversing from non-reversing events; high-pressure cells are needed for oxidation stability and reactions under pressure. Check calorimetric accuracy and repeatability, resolution of weak transitions, sample pan options including sealed and hermetic pans, purge gas control, autosampler availability and the calibration standards supplied. Specify the heating rates, sample mass and the standards the method must follow.
Full specifications
| Pressure range | 0.1–15 MPa |
|---|---|
| Temperature range | 25–600 °C |
| Heating rate | 0.1–50 °C/min |
| Temperature accuracy | ±0.2 °C |
| Calorimetric accuracy | ±2% |
| Crucible type | Sealed stainless steel or gold-plated |
| Sample mass | 1–20 mg |
| Gas supply | Oxygen, air, nitrogen, argon |
| Pressure measurement | Digital gauge, logged with data |
| Safety features | Interlocks, relief valve, burst disc |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Lubricant and oil oxidation stability
- Polymer oxidation induction time
- Propellant and energetic material studies
- High-pressure phase behaviour
- Food oil and fat stability testing
Key features
- Sealed pressure vessel for measurements from 0.1 to 15 MPa
- Gas supply with selectable oxygen, air, nitrogen or other gases
- Sealed crucibles that retain volatile and reactive samples
- Oxidation induction time measurement under elevated oxygen pressure
- Pressure and purge flow logging with each measurement
- Safety interlocks with pressure relief and burst protection
Frequently asked questions about High-Pressure Differential Scanning Calorimeter
OECD and ASTM oxidation induction time methods commonly use oxygen at 3.5 MPa, about 500 psi, and about 210 degrees Celsius for polyolefins. Higher pressures accelerate oxidation and shorten test time, but the value must always be reported with the pressure used.
They keep the sample in contact with the pressurised gas, prevent evaporation of volatiles and stop the specimen from being ejected if it decomposes rapidly. A correctly sealed crucible also protects the sensor and, in oxidation tests, keeps the sample in a defined oxygen atmosphere.
Yes. High-pressure DSC with oxygen requires pressure interlocks, a relief valve or burst disc, an oxygen-clean gas train free of hydrocarbon contamination and trained operators. Crucibles must be inspected for damage and not reused beyond the manufacturer's stated limit.
It measures heat flow as a function of temperature, so it detects melting and crystallization temperatures and enthalpies, glass transitions, specific heat capacity, curing, polymorphic transitions, oxidation induction time and decomposition onset when the sample is sealed.
Temperature and enthalpy are calibrated with reference materials of certified melting point and enthalpy, commonly indium at 156.6 degrees Celsius, plus zinc or tin for higher ranges. Specific heat capacity is calibrated separately with sapphire using ASTM E1269 or an equivalent procedure.
Below the glass transition the material is rigid with low heat capacity; above it, molecular mobility raises the heat capacity, so the curve steps upward. The midpoint of that step is reported as the glass transition temperature, and the result depends on heating rate and thermal history.
Related models