Power-Compensation Differential Scanning Calorimeter
功率补偿型差示扫描量热仪
- SKU DSC-0002
- DSC
- CE (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | −170–730 °C |
|---|---|
| Heating rate | 0.1–200 °C/min |
| Cooling | Liquid nitrogen or mechanical, up to 100 °C/min |
| Temperature accuracy | ±0.1 °C |
| Temperature precision | ±0.01 °C |
| Enthalpy precision | ±0.5% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E967, ASTM E968, ASTM E1356
What is the Power-Compensation Differential Scanning Calorimeter?
A power-compensation DSC heats the sample and reference in separate small furnaces and records the power difference needed to keep them at identical temperature. Sharp peak shapes and fast response make it suited to small transitions, fast heating rates and precise heat capacity work.
What is the Power-Compensation Differential Scanning Calorimeter?
A power-compensation DSC heats the sample and reference in separate small furnaces and records the power difference needed to keep them at identical temperature. Sharp peak shapes and fast response make it suited to small transitions, fast heating rates and precise heat capacity work.
Overview
Two independent low-mass furnaces, each with its own platinum resistance thermometer and heater, are controlled so that sample and reference follow exactly the same temperature program. The electrical power needed to maintain equality is proportional to the heat flow from the sample, which removes the thermal lag of a disc sensor and yields narrow melting peaks and clean resolution of closely spaced transitions. Typical temperature range is minus 170 to 730 degrees Celsius with liquid nitrogen or mechanical cooling, with heating rates up to 200 degrees Celsius per minute for fast-scan work. Temperature accuracy of plus or minus 0.1 degrees Celsius and enthalpy precision near 0.5 percent are usual. Baseline stability and sensitivity to small heat capacity changes make the design useful for specific heat determination, weak polymorphic transitions and thin films, where sample mass is only a few milligrams. The low furnace mass also permits rapid controlled cooling.
How does it work?
The sample and an empty reference pan sit on a sensor in a furnace whose temperature follows a program. In heat-flux instruments a single furnace heats both positions and thermocouples measure the small temperature difference that develops when the sample absorbs or releases heat; the difference is converted to heat flow using a calibration factor. Power-compensation instruments place the sample and reference in separate small furnaces and vary the heating power needed to keep both at the same temperature, recording power difference directly. Either way the result is a curve of heat flow against temperature or time, with the area under a peak giving transition enthalpy in joules per gram. A purge gas, usually nitrogen, prevents oxidation and removes volatiles, and cooling may use liquid nitrogen, a mechanical cooler or an air vortex. Temperature and enthalpy calibration use indium, zinc and other reference materials of known melting point and enthalpy.
What are the key features?
- Separate low-mass sample and reference furnaces with individual heaters
- Temperature range from −170 to 730 °C with controlled cooling
- Heating rates up to 200 °C/min for fast-scan measurements
- Narrow peak shape that resolves closely spaced transitions
- Direct specific heat capacity determination from power difference
- Suitable for samples of only a few milligrams
What are the technical specifications?
A typical power-compensation differential scanning calorimeter configuration offers temperature range of −170–730 °C, heating rate of 0.1–200 °C/min, cooling of Liquid nitrogen or mechanical, up to 100 °C/min, temperature accuracy of ±0.1 °C, temperature precision of ±0.01 °C and enthalpy precision of ±0.5%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Power-Compensation Differential Scanning Calorimeter is used in pharmaceutical polymorph screening, specific heat capacity measurement, fast-scan kinetic studies, thin film and coating analysis and high-purity reference material certification. Across the wider DSC class, laboratories also deploy it for polymer melting and crystallinity testing, pharmaceutical polymorph and purity screening and food starch and fat behaviour studies.
Which standards and certifications apply?
Relevant standards and methods include ASTM E967, ASTM E968, ASTM E1356, ASTM E1269, ISO 11357-1, ISO 11357-2 and USP <891>. Certification candidates for this model are CE, IEC 61010-1 safety and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Calibrate temperature and enthalpy before a measurement campaign, and verify at least one check standard afterwards. Choose the pan material and type for the temperature and chemistry involved: sealed or hermetic pans for volatile samples, high-pressure crucibles for pressure work and open pans for evolved gases. Keep sample mass similar between runs, because thermal lag grows with mass and shifts peaks to higher temperature. Use a defined purge gas flow, dry the cell if moisture condenses, and run an empty baseline periodically. Consumables such as pans, lids and seals are single-use, and crucibles used for pressure work must be inspected for deformation. Follow ASTM E967, E968 or ISO 11357 for calibration and reporting, and record pan type, gas, rate and mass with each curve.
How do I select the right Power-Compensation Differential Scanning Calorimeter?
Select a power-compensation instrument when resolution of small or closely spaced transitions, precise heat capacity or fast heating rates are the priority, as in pharmaceutical polymorph work. For routine materials testing, a heat-flux instrument is usually sufficient and less costly. In an RFQ, specify scanning rates, sample mass, autosampler and calibration standards.
How does it compare with other DSC variants?
Start with the temperature range you need, because it decides the cooling system: studies above room temperature need only an air or mechanical cooler, while glass transitions well below zero require liquid nitrogen. Then choose the measuring principle. Heat-flux instruments are robust, accept larger samples and cover most materials work; power-compensation instruments give sharper peaks and better resolution for small transitions and fast heating rates; temperature-modulated instruments separate reversing from non-reversing events; high-pressure cells are needed for oxidation stability and reactions under pressure. Check calorimetric accuracy and repeatability, resolution of weak transitions, sample pan options including sealed and hermetic pans, purge gas control, autosampler availability and the calibration standards supplied. Specify the heating rates, sample mass and the standards the method must follow.
Full specifications
| Temperature range | −170–730 °C |
|---|---|
| Heating rate | 0.1–200 °C/min |
| Cooling | Liquid nitrogen or mechanical, up to 100 °C/min |
| Temperature accuracy | ±0.1 °C |
| Temperature precision | ±0.01 °C |
| Enthalpy precision | ±0.5% |
| Furnace design | Two independent low-mass furnaces |
| Sample mass | 0.5–10 mg typical |
| Purge gas | Nitrogen or argon, mass flow controlled |
| Automation | Optional 50-position autosampler |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Pharmaceutical polymorph screening
- Specific heat capacity measurement
- Fast-scan kinetic studies
- Thin film and coating analysis
- High-purity reference material certification
Key features
- Separate low-mass sample and reference furnaces with individual heaters
- Temperature range from −170 to 730 °C with controlled cooling
- Heating rates up to 200 °C/min for fast-scan measurements
- Narrow peak shape that resolves closely spaced transitions
- Direct specific heat capacity determination from power difference
- Suitable for samples of only a few milligrams
Frequently asked questions about Power-Compensation Differential Scanning Calorimeter
Because each pan sits in its own low-mass furnace with its own heater and thermometer, the control loop responds almost immediately to heat released or absorbed. The thermal lag of a disc sensor is eliminated, so melting peaks stay narrow even at moderate heating rates.
Yes. The power difference needed to hold equal temperatures is proportional to heat capacity, so a single run with a sapphire standard can give specific heat values without the multiple baselines required by some other designs, provided the calibration is verified regularly.
Two furnaces and two control loops make the instrument more complex and more expensive, and the small pans limit sample mass, which reduces sensitivity for very weak bulk transitions. Sample preparation and pan handling also require more care than with disc-type cells.
It measures heat flow as a function of temperature, so it detects melting and crystallization temperatures and enthalpies, glass transitions, specific heat capacity, curing, polymorphic transitions, oxidation induction time and decomposition onset when the sample is sealed.
Temperature and enthalpy are calibrated with reference materials of certified melting point and enthalpy, commonly indium at 156.6 degrees Celsius, plus zinc or tin for higher ranges. Specific heat capacity is calibrated separately with sapphire using ASTM E1269 or an equivalent procedure.
Below the glass transition the material is rigid with low heat capacity; above it, molecular mobility raises the heat capacity, so the curve steps upward. The midpoint of that step is reported as the glass transition temperature, and the result depends on heating rate and thermal history.